应用笔记

Analysis of Volatile Organic Compounds and Odorants in Drinking Water via HS-SPME Arrow-GC-MS

08 Oct 2025

feature EVAN4283

Abstract

Volatile organic compounds are commonly tested in drinking water to ensure safety for consumption. Additionally, odorants, like methylisoborneol (MIB) and geosmin, are of interest due to their off-putting aromas. It is of interest to combine these methods, which are typically run separately, into one method to achieve higher sample throughput. Solid-Phase Microextraction (SPME) using the SPME Arrow is of great interest due to its ability to combine sample preparation and sample introduction into one device. The SPME Arrow’s ability to be automated using a rail-type autosampler lends itself to creating an even greater advantage in minimizing time spent preparing samples. The present method utilizes a simplified sample preparation workflow combined with Headspace-Solid Phase Microextraction-Gas Chromatography-Mass Spectrometry (HS-SPME-GC-MS) to determine VOCs and odorants in drinking water. This method shows a span of calibration ranges achieving r2 values of 0.923-0.999 (average: 0.998); % residual standard errors (%RSEs) of 2-27% (average: 6%); analytical precisions of 2-15% RSDs (average: 4% RSD); and limit of quantification (LOQ) values of 0.0029-3.033 ng/mL (ppb, average: 0.223 ng/mL).

Introduction

Volatile organic compounds (VOCs) are commonly tested for in water due to their known toxicity and impact on human health. While odor compounds, like 2-methylisoborneol (MIB) and geosmin, do not pose a health risk, they are commonly tested for due to their off-putting, earthy, and musty taste and odor [1]. The analysis of VOCs and odor compounds is typically separated into two different analyses. By combining these methods, higher throughput and laboratory efficiency can be achieved. An approach to determine the presence of these compounds in water is Headspace-Solid Phase Microextraction-Gas Chromatography-Mass Spectrometry (HS-SPME-GC-MS) is presented here. 

Experimental

Preparation of VOCs and Odor Solutions

Restek’s Drinking Water Odor Standard (cat# 30608) and a custom VOC standard (see Restek’s custom reference standards) were used for method development. Additionally, three internal standards (ISTDs) were selected: fluorobenzene (cat# 30030); 1,4-dioxane-d8 (cat# 30614); and 4-bromofluorobenzene (cat# 30026). Individual working solutions were prepared in methanol for each calibration point for both the VOCs and Odors, and Method Detection Limits (MDL) were determined. The final sample size for this method was 5 mL in a 20 mL HS vial. After identifying the concentrations for the calibration and MDL, the working solution concentrations were calculated for making a 10 µL addition to reach the desired final concentrations. See Tables I & II for working solution concentrations and their corresponding calibration concentrations.

Table I: VOC Calibration and Working Solutions Concentrations

VOCs
LevelCalibration Concentration (ppb, ng/mL)Working Solution Concentration (ppb, ng/mL)
10.2100
20.5250
31500
421000
552500
6105000
72512,500
87537,500
9200100,000

Table II: Odorants Calibration and Working Solution Concentrations

Odorants
LevelCalibration Concentration (ppb, ng/mL)Working Solution Concentration (ppb, ng/mL)
10.0010.5
20.0021
30.0042
40.0084
50.0168
60.03216
70.06432
80.12864
90.256128

A working solution of the three selected ISTDs was prepared in methanol. The working solution was prepared at 10 µg/mL (ppm). Once calibration and MDL samples were prepared, the samples were then spiked with the 50 µL of the working solution to reach a final concentration of 100 ng/mL (ppb) in the final sample.

MDLs for each compound were determined similar to method detection limits per the procedure as dictated by the U.S. Environment Protection Agency (EPA) [2]. In following this procedure, seven replicate injections were made at low-level spikes (VOCs at 0.002 and 2 ng/mL, Odors at 0.008 ng/mL). See Table III for MDL spike concentrations. The standard deviation of these results was then multiplied by the student’s t-value of 3.143 to determine the MDL. The limit of quantification (LOQ) for each compound was determined as 10 times (10x) the standard deviation of the MDL samples. 

Table III: MDL Spike Levels

#CompoundMDL Concentration (ng/mL, ppb)
1Chloroethene0.2
21,1-Dichloroethene0.2
3Dichloromethane2
4trans-1,2-Dichloroethene0.2
5MTBE0.02
6cis-1,2-Dichloroethene0.02
7Chloroform0.02
8Carbon Tetrachloride0.02
91,1,1-Trichloroethane0.02
10Benzene0.02
111,2-Dichloroethane0.02
12Trichloroethene0.02
131,2-Dichloropropane0.02
14Bromodichloromethane2
151,4-Dioxane0.02
16cis-1,3-Dichloropropene0.02
17Toluene0.02
18Tetrachloroethene0.02
19trans-1,3-Dichloropropene0.02
20Dibromochloromethane0.02
21m,p-Xylene0.02
22o-Xylene0.02
23Bromoform0.02
241,2-Dichlorobenzene0.02
252-Methylisoborneol0.008
26Geosmin0.008

HS-SPME Arrow-GC-MS Parameters

All results were obtained using HS-SPME Arrow-GC-MS. HS-SPME Arrow, GC, and MS parameters can be found in Table IV, V, and VI respectively.

Table IV: HS-SPME Arrow Parameters

CTC PAL Parameters
Sampling ModeHeadspace
SPME Device1.10 mm SPME Arrow (cat.# 28903-1)
SPME Phase120 µm CWR/PDMS
Vial Penetration Depth45 mm
Injector Penetration Depth50 mm
Incubation Temperature/Time30 °C/120 s
Extraction Temperature/Time30 °C/120 s
Desorption Temperature/Time280 °C/1 min
Conditioning Temperature/Time280 °C
Pre Conditioning1 min
Post Conditioning0 min

Table V: GC Parameters

Agilent 7890B/5977B GC-AAgilent 7890B/5977B GC-MS Parameters
ColumnRtx-VMS– 30 m x 0.25 mm x 1.40 µm (cat# 19915)
Injection ModeSplit (15:1)
Injection VolumeSee CTC PAL Parameters
LinerTopaz 1.8 mm ID Straight/SPME Inlet Liner (cat# 23280)
Injection Temperature280 °C
Purgle Flow3 mL/min
Oven35 °C (hold 3 min) to 60 °C at 8 °C/min (hold 0 min) to 280 °C (hold 3 min) at 30 °C/min; Total time = 16.46 min
Carrier GasHe, constant flow
Flow Rate2.0 mL/min
DetectorHES – MS
ModeSIM
Transfer Line Temperature260 °C
Source Temperature 325 °C
Quad Temperature200 °C
Acquisition Range29 – 300 amu
Rate5.1 scans/sec

Table VI: MS SIM Parameters

5977B MS SIM Parameters
Group #Start TimeMass, Dwell Time
Group 10 min(62, 30) (64, 15)
Group 21.8 min(61, 15) (96, 30)
Group 32.3 min(49,30) (57, 30) (61, 30) (73, 30) (84, 30) (96, 30)
Group 43.5 min(61, 30) (96, 30)
Group 54.3 min    (47, 30) (61, 30) (82,30) (83, 30) (97, 30) (117,30)
Group 64.9 min(52 ,30) (78, 30)
Group 75.3 min(49, 30) (62, 30) (70, 30)
Group 86.2 min(63, 30) (76, 30) (83, 30) (129, 30)
Group 96.75 min(57, 15) (64, 15) (88, 15) (96, 15)
Group 107.1 min(65, 30) (75, 30) (91, 30) (110, 30)
Group 117.6 min(75, 30) (79, 30) (110, 30) (129, 30) (166, 30)
Group 128.4 min(91, 30) (93, 30) (106, 30) (173, 30)
Group 139.1 min(95, 30) (174, 30)
Group 149.6 min(111, 30) (146, 30)
Group 1510.5 min(95, 30) (107, 30) (135, 30)

Results and Discussion

The separation of VOCs and odor compounds is shown in Figure 1. Method performance can be seen in Table VII. The calibration for the VOCs spanned from 0.2-200 ng/mL, while the odors calibration ranged from 0.001-0.256 ng/mL. The calibrations were assessed for each targeted compound, leading to different calibration ranges depending on the compound of interest. These differences are attributed to the SPME Arrows phase capacity and the extraction rate of each compound by the SPME Arrow phase.

The calibration r2 values ranged from 0.923-0.999 (average 0.998). A better assessment of each calibration is evaluating the % Residual Standard Error (%RSE), which tests each calibration level to the respective calibration curve. The %RSE across all compounds ranged from 2-27% (average: 6%). The two outliers being dichloromethane and 1,4-dioxane with %RSE values of 18% and 27%. All other compounds fell between 2-10% RSE. Method precision was evaluated from the seven replicate injections, showing %RSDs (Relative Standard Deviation) ranging from 2-15% (average: 4%). This falls within the typical method acceptance criteria of +/- 30%. Finally, the LOQs ranged from 0.0029-3.033 ng/mL (ppb, average: 0.223 ng/mL).

Analysis of VOCs and Odors in Drinking Water via HS-SPME Arrow-GC-MS

Figure 1: Analysis of VOCs and Odors in Drinking Water via HS-SPME-Arrow-GC-MS
Analysis of VOCs and Odors in Drinking Water via HS-SPME Arrow-GC-MS

GC_EV1517

Table VII: Method Performance

Peak #CompoundRet. Time (min)Calibration Range (ng/mL, ppb)Calibration Fit & Weightingr2Calibration (%RSE)LOQMethod Precision (%RSD)
1Chlorethene11.200.2-25Quad., 1/x20.9994%0.0784%
21,1-Dichloroethene12.070.2-25Quad., 1/x20.9919%0.1246%
3Dichloromethane12.632-200Quad., 1/x20.99918%3.03315%
4trans-1,2-Dichloroethene12.800.2-25Quad., 1/x20.9992%0.0352%
5MTBE12.940.2-10Quad., 1/x20.9993%0.0593%
6cis-1,2-Dichloroethene14.150.2-25Quad., 1/x20.9994%0.0563%
7Chloroform14.500.2-25Quad., 1/x20.9985%0.0633%
8Carbon Tetrachloride14.610.2-25Quad., 1/x20.9994%0.0583%
91,1,1-Trichloroethane14.700.2-25Quad., 1/x20.9978%0.063%
10Benzene15.180.2-75Quad., 1/x20.9985%0.0042%
111,2-Dichloroethane15.460.2-25Quad., 1/x20.9994%0.0094%
12Trichloroethene15.960.2-75Quad., 1/x20.9994%0.0975%
131,2-Dichloropropane16.580.2-75Quad., 1/x20.9969%0.052%
14Bromodichloromethane16.680.2-75Quad., 1/x20.9994%0.052%
151,4-Dioxane26.762-200Linear, 1/x20.92327%0.8953%
16cis-1,3-Dichloropentane17.270.2-200Quad., 1/x20.9975%0.0412%
17Toluene17.450.2-25Quad., 1/x20.9993%0.1236%
18Tetrachloroethene17.730.2-25Quad., 1/x20.9992%0.0422%
19trans-1,3-Dichloropropene17.790.2-200Quad., 1/x20.9975%0.0583%
20Dibromochloromethane38.010.2-25Quad., 1/x20.9992%0.0472%
21m,p-Xylene38.630.4-20Quad., 1/x20.9995%0.56414%
22o-Xylene38.890.2-10Quad., 1/x20.9994%0.1397%
23Bromoform38.940.2-75Quad., 1/x20.9984%0.0563%
241,2-Dichlorobenzene39.890.2-25Quad., 1/x20.9992%0.0453%
252-Methylisoborneol311.070.004-0.256Quad., 1/x20.99710%0.0059%
26Geosmin312.170.004-0.256Quad., 1/x20.99610%0.0035%

ISTD Reference – Fluorobenzene1, 1,4-Dioxane-d82, p-Bromofluorobenzene3

Conclusion

HS-SPME Arrow-GC-MS offers a simple approach for the determination of VOCs and Odorants in drinking water. This method shows the SPME Arrow’s capabilities of quantifying a range of different compounds at various detection limits. Notably, the odorants were able to be detected at low part per trillion levels proving this is a viable sampling and testing method for the determination of their presence in drinking water. Future considerations for method optimization would be to select a more appropriate ISTD for the MIB and Geosmin. 

References

  1. Cobb County-Marietta Water Authority (CCMWA), Geosmin and methyl-Isoborneol (MIB), online article, 2023. https://www.ccmwa.org/education/geosmin-and-methyl-isoborneol-mib
  2. United States Environmental Protection Agency (U.S. EPA), Definition and procedure for the determination of the method detection limit, revision 2, December 2016. https://www.epa.gov/sites/default/files/2016-12/documents/mdl-procedure_rev2_12-13-2016.pdf

相关产品


Drinking Water Odor Standard – Geosmin and 2-Methylisoborneol, 100 µg/mL in P&T Methanol, 1 mL/ampul
Fluorobenzene, 2000 µg/mL, P&T Methanol, 1 mL/ampul
1,4-Dioxane-d8 Standard, 2000 µg/mL, P&T Methanol, 1 mL/ampul
4-Bromofluorobenzene (BFB) Standard, 2000 µg/mL, P&T Methanol, 1 mL/ampul
PAL Smart SPME Arrow 1.10 mm: Carbon-WR/PDMS, Phase Thickness 120 µm, Phase Length 20 mm, Light Blue, ea.
Rtx-VMS GC Capillary Column, 30 m, 0.25 mm ID, 1.40 µm
Topaz, Straight/SPME Inlet Liner, 1.8 mm x 6.5 x 78.5, for Agilent GCs, Premium Deactivation, 5-pk.
PAL Smart SPME Arrow 1.10 mm: Carbon-WR/PDMS, Phase Thickness 120 µm, Phase Length 20 mm, Light Blue, 3-pk.

作者

  • Colton Myers

    Colton Myers is a Product Manager for GC columns at Restek Corporation with over 10 years of experience in research and development. He specializes in driving new GC column technologies from concept to market, combining deep technical expertise with a strong focus on innovation and commercialization. Colton holds a BS in chemistry from Juniata College and a Master of Management in Corporate Innovation and Entrepreneurship from Penn State University.

    View Articles
EVAN4283