文章

TEST figure testing block

12 Nov 2025

Figure Heading 1
Rum on Rtx-1301

GC_FF1390

Peaks

PeakstR (min)
1.Acetaldehyde5.410
2.Methanol5.634
3.Ethanol7.026
4.1-Propanol9.492
5.Ethyl acetate10.321
6.Isobutanol11.387
7.Acetal13.138
8.3-Methyl-1-butanol14.334
9.2-Methyl-1-butanol14.420
10.Propylene glycol15.749

Conditions

ColumnRtx-1301, 60 m, 0.25 mm ID, 1.0 µm (cat.# 16056)
Standard/SampleRum (neat)
Diluent:n/a
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz, precision inlet liner, 4.0 mm x 6.3 x 78.5, with wool (cat.# 23305)
Inj. Temp.:250 °C
Split Vent Flow Rate:144.4 mL/min
Oven
Oven Temp.:35 °C (hold 3 min) to 220 °C at 8 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
Linear Velocity:25.457 cm/sec @ 35 °C
DetectorFID @ 250 °C
Make-up Gas Flow Rate:25 mL/min
Make-up Gas Type:N2
Hydrogen flow:35 mL/min
Air flow:400 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890A GC

heading 2

Semivolatiles Analysis on Rxi-SVOCms

GC_EV1604

Peaks

PeakstR (min)
1.(IS) 1,4-Dioxane-d81.87
2.<a class="cmpd_link" title="View compound information for N-Nitrosodimethylamine” title=”View compound information for N-Nitrosodimethylamine” href=”https://ez.restek.com/compound/view/en/62-75-9/N-Nitrosodimethylamine”>N-Nitrosodimethylamine2.00
3.Pyridine2.03
4.(SS) 2-Fluorophenol2.67
5.(SS) Phenol-d63.29
6.Phenol3.30
7.Aniline3.36
8.Bis(2-chloroethyl) ether3.40
9.2-Chlorophenol3.46
10.1,3-Dichlorobenzene3.59
11.(IS) 1,4-Dichlorobenzene-D43.63
12.1,4-Dichlorobenzene3.65
13.Benzyl alcohol3.72
14.1,2-Dichlorobenzene3.78
15.2-Methylphenol3.80
16.Bis(2-Chloroisopropyl)ether3.84
17.4-Methylphenol3.93
18.3-Methylphenol3.93
19.<a class="cmpd_link" title="View compound information for N-Nitrosodi-N-propylamine” title=”View compound information for N-Nitrosodi-N-propylamine” href=”https://ez.restek.com/compound/view/en/621-64-7/N-Nitrosodi-N-propylamine”>N-Nitrosodi-N-propylamine3.95
20.Hexachloroethane4.07
21.(SS) Nitrobenzene-D54.10
22.Nitrobenzene4.11
23.Isophorone4.32
24.2-Nitrophenol4.40
25.2,4-Dimethylphenol4.42
26.Benzoic acid4.46
27.Bis(2-chloroethoxy)methane4.51
28.2,4-Dichlorophenol4.61
29.1,2,4-Trichlorobenzene4.70
30.(IS) Naphthalene-D84.76
31.Naphthalene4.78
PeakstR (min)
32.4-Chloroaniline4.82
33.Hexachlorobutadiene4.89
34.4-Chloro-3-methylphenol5.26
35.2-Methylnaphthalene5.43
36.1-Methylnaphthalene5.53
37.Hexachlorocyclopentadiene5.59
38.2,4,6-Trichlorophenol5.70
39.2,4,5-Trichlorophenol5.73
40.(SS) 2-Fluorobiphenyl5.79
41.2-Chloronaphthalene5.91
42.2-Nitroaniline6.00
43.1,4-Dinitrobenzene6.13
44.Dimethyl phthalate6.18
45.1,3-Dinitrobenzene6.20
46.2,6-Dinitrotoluene6.24
47.1,2-Dinitrobenzene6.29
48.Acenaphthylene6.31
49.3-Nitroaniline6.40
50.(IS) Acenaphthene-D106.45
51.Acenaphthene6.48
52.2,4-Dinitrophenol6.50
53.4-Nitrophenol6.55
54.2,4-Dinitrotoluene6.63
55.Dibenzofuran6.65
56.2,3,5,6-Tetrachlorophenol6.73
57.2,3,4,6-Tetrachlorophenol6.77
58.Diethyl phthalate6.88
59.4-Chlorophenyl phenyl ether6.99
60.Fluorene6.99
61.4-Nitroaniline7.00
62.4,6-Dinitro-2-methylphenol7.03
PeakstR (min)
63.<a class="cmpd_link" title="View compound information for N-Nitrosodiphenylamine” title=”View compound information for N-Nitrosodiphenylamine” href=”https://ez.restek.com/compound/view/en/86-30-6/N-Nitrosodiphenylamine”>N-Nitrosodiphenylamine7.10
64.<a class="cmpd_link" title="View compound information for N,N-Diphenylhydrazine” title=”View compound information for N,N-Diphenylhydrazine” href=”https://ez.restek.com/compound/view/en/530-50-7/N,N-Diphenylhydrazine”>N,N-Diphenylhydrazine7.15
65.(SS) 2,4,6-Tribromophenol7.23
66.4-Bromophenyl phenyl ether7.47
67.Hexachlorobenzene7.53
68.Pentachlorophenol7.72
69.(IS) Phenanthrene-D107.92
70.Phenanthrene7.94
71.Anthracene7.99
72.Carbazole8.15
73.di-n-Butyl phthalate8.49
74.Fluoranthene9.12
75.Benzidine9.24
76.(SS) Pyrene-D109.32
77.Pyrene9.34
78.(SS) p-Terphenyl-d149.49
79.3,3′-Dimethylbenzidine9.98
80.Butyl benzyl phthalate10.00
81.Bis(2-ethylhexyl) adipate10.09
82.3,3′-Dichlorobenzidine10.62
83.Benz[a]anthracene10.66
84.(IS) Chrysene-D1210.67
85.Chrysene10.71
86.Bis(2-ethylhexyl) phthalate10.71
87.Di-n-octyl phthalate11.68
88.Benzo[b]fluoranthene12.30
89.Benzo[k]fluoranthene12.34
90.Benzo[a]pyrene12.89
91.(IS) Perylene-D1213.00
92.Indeno[1,2,3-cd]pyrene15.32
93.Dibenz[a,h]anthracene15.40
94.Benzo[ghi]perylene15.95
All compounds are 2 ng on column.

Conditions

ColumnRxi-SVOCms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 16623)
Standard/Sample
8270 MegaMix standard (cat.# 31850)
8270 Benzidines mix (cat.# 31852)
Benzoic acid (cat.# 31879)
Revised SV internal standard mix (cat.# 31886)
Revised B/N surrogate mix (cat.# 31888)
Acid surrogate mix (cat.# 31063)
Diluent:Dichloromethane
Conc.:20 µg/mL
Injection
Inj. Vol.:1 µL split (split ratio 10:1)
Liner:Topaz 4.0 mm ID single taper inlet liner with wool (cat.# 23303)
Inj. Temp.:250 °C
Split Vent Flow Rate:12 mL/min
Oven
Oven Temp.:60 °C (hold 0.5 min) to 285 °C at 25 °C/min to 305 °C at 3 °C/min to 330 °C at 20 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1.2 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
11.5535-5005.9
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Source Type:Inert
Drawout Plate:6 mm ID
Source Temp.:330 °C
Quad Temp.:180 °C
Electron Energy:70 eV
Tune Type:DFTPP
Ionization Mode:EI
InstrumentAgilent 7890A GC & 5975C MSD
Sample PreparationSamples were aliquoted into amber 2 mL, 9 mm short-cap, screw-thread vials (cat.# 21143) containing glass Big Mouth inserts (cat.# 21782) and sealed with 2.0 mL, 9 mm short-cap, screw-vial closures (cat.# 23842).
Figure Heading 2
Rtx-1301 Water, Acetaldehyde, and Methanol Separation

GC_FF1389

Peaks

PeakstR (min)Conc.
(µg/mL)
1.Water4.312
2.Acetaldehyde4.483100
3.Methanol4.699100
4.Ethanol5.871100

Conditions

ColumnRtx-1301, 60 m, 0.25 mm ID, 1.00 µm (cat.# 16056)
Standard/SampleCustom acetaldehyde, methanol, and ethanol standard in water
Diluent:Water
Conc.:100 µg/mL
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz Precision inlet liner with wool, 4.0 mm x 6.3 x 78.5 (cat.# 23305)
Inj. Temp.:250 °C
Split Vent Flow Rate:141.4 mL/min
Oven
Oven Temp.:35 °C (hold 2.45 min) to 220 °C at 7 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
Linear Velocity:30.261 cm/sec @ 35 °C
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
1010-4006.84
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Source Temp.:230 °C
Quad Temp.:150 °C
Electron Energy:70 eV
Solvent Delay Time:0.0 min
Tune Type:PFTBA
Ionization Mode:EI
InstrumentAgilent 7890A GC & 5975C MSD
NotesThe EZGC method translator was used to establish the MS conditions.

Author

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    View all posts