文章

Simplify Alcohol Analyses with Application-Designed Chromatography Columns

21 Jul 2021

On the surface, alcohol analyses appear relatively straightforward. They involve few target analytes, are not extremely sensitive, and have quick sample preparation. However, they are not as simple as they seem. Ethanol analyses present distinct chromatography challenges when assaying in biological samples. Compounds found in biological matrices can interfere with the identification and quantitation of ethanol and its associated analytes.

Restek has developed tailored columns to address the common analyte interferences that can hinder productive chromatography. Dual capillary column sets for GC are available for common blood alcohol analyses, and high-throughput analytical LC column and guard cartridges are available for urinary alcohol assays. Starting method development with tailored columns designed specifically to lessen interferences can help labs develop methods faster and keep instruments running.

Blood Alcohol Content (BAC)

Gas chromatography coupled with headspace sampling and a flame ionization detector (HS-GC-FID) is the most commonly used technique for the analysis of blood alcohol in biological specimens from both living and dead persons. HS-GC-FID is remarkedly simple, reliable, and linear over the analytical range needed for the quantitative detection of alcohol. The troublesome part of the chromatographic analysis is ensuring analyte specificity from endogenous compounds. This is extremely important when analyzing postmortem samples due to the risk that fermentation or putrefaction may produce volatile chemical byproducts that can interfere with the analysis. For this reason, it is always recommended to use a dual column method. By splitting the sample after the injection port and analyzing it simultaneously on two different stationary phases, we can both quantify and confirm the analyte’s identity in a single run (Figure 1). The target analytes will elute in the calibrated retention time window on both columns, thus confirming a positive identification. To simplify column selection, Restek has paired two distinct sets of column phases: Rtx-BAC1 and Rtx-BAC2 columns and Rtx-BAC Plus 1 and Rtx-BAC Plus 2 columns for in vivo and postmortem analysis.

Figure 1: HS-GC-FID dual column setup is common for BAC testing.
figure article CFAR3698 01

Rtx-BAC1 and Rtx-BAC2 columns are designed for routine in vivo blood alcohol analyses (Figure 2). This column pair has been in use for decades and has proven to provide stable, defensible results for routine BAC testing. When using n-propanol as the internal standard, this is the industry standard.

Figure 2: The industry standard for in vivo BAC analysis is the Rtx-BAC1 and Rtx-BAC2 column set.

BAC Resolution Control Standard with n-Propanol on Rtx-BAC1 Column

BAC Resolution Control Standard n-Propanol on Rtx-BAC1

GC_CF1172

Peaks

PeakstR (min)Conc.
(µg/mL)
1.Methanol0.886200
2.Acetaldehyde0.993200
3.Ethanol1.123200
4.2-Propanol1.381200
5.Acetone1.709200
6.n-Propanol (IS)1.873200

Conditions

ColumnRtx-BAC1, 30 m, 0.32 mm ID, 1.8 µm (cat.# 18003)
Standard/SampleBAC resolution control standard n-P (cat.# 36010)
Diluent:Water
Conc.:200 µg/mL
Injectionheadspace-loop split (split ratio 50:1)
Liner:Topaz 1.0 mm ID straight inlet liner (cat.# 23333)
Inj. Port Temp.:220 °C
Headspace-Loop
Instrument:Tekmar HT3
Inj. Time:0.5 min
Transfer Line Temp.:200 °C
Valve Oven Temp.:200 °C
Standby flow rate:100 mL/min
Sample Temp.:60 °C
Platen temp equil. time:60 min
Sample Equil. Time:5 min
Vial Pressure:30 psi
Pressurize Time:2 min
Pressure Equilibration Time:0.2 min
Loop Pressure:20 psi
Loop Fill Time:1 min
Oven
Oven Temp.:40 °C (hold 4 min)
Carrier GasHe, constant flow
Linear Velocity:80 cm/sec
DetectorFID @ 240 °C
Make-up Gas Flow Rate:45 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:400 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890A GC

BAC Resolution Control Standard with n-Propanol on Rtx-BAC2 Column

BAC Resolution Control Standard n-Propanol on Rtx-BAC2

GC_CF1174

Peaks

PeakstR (min)Conc.
(µg/mL)
1.Acetaldehyde1.011200
2.Methanol1.199200
3.Ethanol1.623200
4.Acetone1.758200
5.2-Propanol1.929200
6.n-Propanol (IS)3.165200

Conditions

ColumnRtx-BAC2, 30 m, 0.32 mm ID, 1.2 µm (cat.# 18002)
Standard/SampleBAC resolution control standard n-P (cat.# 36010)
Diluent:Water
Conc.:200 µg/mL
Injectionheadspace-loop split (split ratio 50:1)
Liner:Topaz 1.0 mm ID straight inlet liner (cat.# 23333)
Inj. Port Temp.:220 °C
Headspace-Loop
Instrument:Tekmar HT3
Inj. Time:0.5 min
Transfer Line Temp.:200 °C
Valve Oven Temp.:200 °C
Standby flow rate:100 mL/min
Sample Temp.:60 °C
Platen temp equil. time:1 min
Sample Equil. Time:5 min
Vial Pressure:30 psi
Pressurize Time:2 min
Pressure Equilibration Time:0.2 min
Loop Pressure:20 psi
Loop Fill Time:1 min
Oven
Oven Temp.:40 °C (hold 4 min)
Carrier GasHe, constant flow
Linear Velocity:80 cm/sec
DetectorFID @ 240 °C
Make-up Gas Flow Rate:45 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:400 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890A GC

For labs looking to expand their HS-GC-FID platform beyond routine in vivo BAC testing, the Rtx-BAC Plus 1 and Rtx-BAC Plus 2 column set is a more suitable option. When analyzing postmortem blood alcohol, t-butanol is often used as the internal standard instead of n-propanol. N-propanol can be created by putrefaction, which can alter quantitation. When the internal standard, which exists to stabilize quantitation, is erroneously high, the results will lose accuracy. The Rtx-BAC Plus 1 and Rtx-BAC Plus 2 columns were designed to fully resolve the alternate internal standard, t-butanol, making it amenable to a fast, two-minute postmortem BAC analysis (Figure 3). With this column set, GC-FID workflows can also be expanded to include inhalantsglycols and GHB.

Figure 3: The Rtx-BAC Plus 1 and Rtx-BAC Plus 2 column set was designed to resolve the alternative internal standard, t-butanol, for postmortem BAC analysis.

BAC Resolution Control Standard with t-Butanol on Rtx-BAC Plus 1 Column

BAC Resolution Control Standard t-Butanol on Rtx-BAC Plus1

GC_CF1178

Peaks

PeakstR (min)Conc.
(µg/mL)
1.Methanol0.859200
2.Acetaldehyde0.910200
3.Ethanol1.085200
4.2-Propanol1.328200
5.Acetone1.444200
6.<a class="cmpd_link" title="View compound information for t-Butanol (IS)” title=”View compound information for t-Butanol (IS)” href=”https://ez.restek.com/compound/view/en/75-65-0/t-Butanol”>t-Butanol (IS)1.574200

Conditions

ColumnRtx-BAC Plus 1, 30 m, 0.32 mm ID, 1.8 µm (cat.# 18004)
Standard/SampleBAC resolution control standard t-B (cat.# 36011)
Diluent:Water
Conc.:200 µg/mL
Injectionheadspace-loop split (split ratio 50:1)
Liner:Topaz 1.0 mm ID straight inlet liner (cat.# 23333)
Inj. Port Temp.:220 °C
Headspace-Loop
Instrument:Tekmar HT3
Inj. Time:0.5 min
Transfer Line Temp.:200 °C
Valve Oven Temp.:200 °C
Standby flow rate:100 mL/min
Sample Temp.:60 °C
Platen temp equil. time:1 min
Sample Equil. Time:5 min
Vial Pressure:30 psi
Pressurize Time:2 min
Pressure Equilibration Time:0.2 min
Loop Pressure:20 psi
Loop Fill Time:1 min
Oven
Oven Temp.:40 °C (hold 4 min)
Carrier GasHe, constant flow
Linear Velocity:80 cm/sec
DetectorFID @ 240 °C
Make-up Gas Flow Rate:45 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:400 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890A GC

BAC Resolution Control Standard with t-Butanol on Rtx-BAC Plus 2 Column

BAC Resolution Control Standard t-Butanol on Rtx-BAC Plus 2

GC_CF1176

Peaks

PeakstR (min)Conc.
(µg/mL)
1.Acetaldehyde0.771200
2.Methanol0.818200
3.Ethanol0.979200
4.Acetone1.060200
5.2-Propanol1.116200
6.<a class="cmpd_link" title="View compound information for t-Butanol (IS)” title=”View compound information for t-Butanol (IS)” href=”https://ez.restek.com/compound/view/en/75-65-0/t-Butanol”>t-Butanol (IS)1.229200

Conditions

ColumnRtx-BAC Plus 2, 30 m, 0.32 mm ID, 0.6 µm (cat.# 18006)
Standard/SampleBAC resolution control standard t-B (cat.# 36011)
Diluent:Water
Conc.:200 µg/mL
Injectionheadspace-loop split (split ratio 50:1)
Liner:Topaz 1.0 mm ID straight inlet liner (cat.# 23333)
Inj. Port Temp.:220 °C
Headspace-Loop
Instrument:Tekmar HT3
Inj. Time:0.5 min
Transfer Line Temp.:200 °C
Valve Oven Temp.:200 °C
Standby flow rate:100 mL/min
Sample Temp.:60 °C
Platen temp equil. time:1 min
Sample Equil. Time:5 min
Vial Pressure:30 psi
Pressurize Time:2 min
Pressure Equilibration Time:0.2 min
Loop Pressure:20 psi
Loop Fill Time:1 min
Oven
Oven Temp.:40 °C (hold 4 min)
Carrier GasHe, constant flow
Linear Velocity:80 cm/sec
DetectorFID @ 240 °C
Make-up Gas Flow Rate:45 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:400 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890A GC

Urinary Alcohol Metabolites

When ethanol metabolites are being monitored in urine, which is commonly performed in treatment programs, high-throughput LC-MS/MS methods are employed. This technique is highly sensitive and fast. Since mass spectrometry is being used, analyte identification is not as significant a concern compared to when GC-FID is used. However, matrix interferences are still problematic during chromatographic assays. The presence of ethanol is determined using two urinary metabolites—ethyl glucuronide and ethyl sulfate. These analytes are highly water-soluble and, therefore, can be difficult to retain using traditional reversed-phase columns. Because of the inherent variability of urine samples and the common presence of early eluting matrix components, not retaining the target analytes beyond the zone of matrix interferences can cause ionization suppression, resulting in low sensitivity and problematic quantitation. Restek has developed a novel stationary phase to retain and separate the polar analytes EtG and EtS from common urine matrix interferences. As part of our Raptor column line, the Raptor EtG/EtS LC column features an SPP particle for improved cycle times and stability. It effectively separates EtG and EtS from potentially interfering matrix, ensuring sensitivity and accurate quantitation, in a fast, four-minute analysis.

Figure 4: The Raptor EtG/EtS column has the retention needed to separate ethanol metabolites from common matrix interference.
EtG/EtS on Raptor EtG/EtS: Updated Method Conditions (Mobile Phase and Gradient)

LC_CF0739

Peaks

PeakstR (min)Conc.
(ng/mL)
Precursor IonProduct IonProduct Ion
1.Ethyl-β-D-glucuronide-d5 (EtG-d5)1.21200226.285.0
2.Ethyl-β-D-glucuronide (EtG)1.23500221.275.185.1
3.Ethyl sulfate-d5 (EtS-d5)2.3650130.198.0
4.Ethyl sulfate (EtS)2.38500125.197.180.0

Conditions

ColumnRaptor EtG/EtS (cat.# 9325A12)
Dimensions:100 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Guard Column:UltraShield UHPLC precolumn filter, 0.2 μm frit (cat.# 25809)
Temp.:35 °C
Standard/Sample
Ethyl-β-D-glucuronide-d5 (EtG-d5) (cat.# 34102)
Ethyl-β-D-glucuronide (EtG) (cat.# 34101)
Ethyl sulfate-d5 sodium salt (EtS-d5) (cat.# 34104)
Ethyl sulfate sodium salt (EtS) (cat.# 34103)
Diluent:0.01% Formic acid in water
Conc.:500 ng/mL (40x acetonitrile precipitation sample prep)
Inj. Vol.:10 µL
Mobile Phase
A:0.01% Formic acid in water
B:0.1% Formic acid in acetonitrile
Time (min)Flow (mL/min)%A%B
0.000.5955
3.000.56535
3.010.5955
4.500.5955
DetectorMS/MS
Ion Source:Electrospray
Ion Mode:ESI-
Mode:MRM
InstrumentHPLC
Sample PreparationA 500 ng/mL standard was prepared in urine. A 50 μL aliquot was mixed with 10 μL of internal standard (20 μg/mL EtG-d5 and 5 μg/mL EtS-d5 in water) and 150 µL of acetonitrile by vortexing at 3000 rpm for 10 seconds and centrifuged at 4300 rpm for 10 minutes at 10 °C. After centrifugation, 100 µL of the supernatant was diluted with 900 µL (40x dilution) of 0.01% formic acid in water. The sample was then vortexed at 3000 rpm for 10 seconds and injected for LC-MS/MS analysis.

作者

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    View Articles
CFAR3698