文章

The Advantage of 2.1 mm ID Columns for LC-MS/MS Analysis of Drugs of Abuse

03 Oct 2025

feature CFAR4190

Introduction

The biphenyl phase lends superior selectivity over a C18 column for drugs of abuse (DOA) panels [1], but choosing the right column dimension for analysis is key for obtaining robust and accurate data. Each column dimension can be advantageous in different scenarios, but generally clinical labs are all working towards the same goals: high throughput, low sample volume, good sensitivity, and low cost. In this article, the advantages of narrow-bore columns will be discussed and demonstrated for drugs of abuse.  

Efficiency

Column efficiency is the ability of a column to produce narrow chromatographic peaks and is defined by plate number (N) following Equation 1:

Equation 1:
Where Rt is the retention time of the analyte and W is peak width.

Where Rt is the retention time of the analyte and W is peak width at baseline.

There can be several contributing factors for peak width and retention time that will ultimately determine the efficiency of the column. Band-broadening processes are factors that contribute to peak dispersion. These include Eddy diffusion, longitudinal mass transfer, and mobile phase and stationary phase mass transfer, which all collectively make up the van Deemter equation [2]. Eddy diffusion occurs when analytes take different flow paths through the column, which is exacerbated when using larger ID columns, longer column dimensions, larger particle sizes, and columns with poor packing efficiency. When a sample is injected on a column, the analyte is most concentrated in the middle, and as this band moves through the column, the analytes get dispersed outward. This is referred to as longitudinal mass transfer and can be reduced by running at higher flow rates. Mass transfer of the mobile phase and stationary phase is defined as analyte partitioning in and out of the stationary phase. This contributing factor to band broadening can be reduced by using smaller particle sizes and elevating the column oven temperature to increase the rate of diffusion. In general, using narrow-bore columns will result in less band broadening, increased peak efficiency, and improved analyte sensitivity. Particle type can also aid in increasing efficiency. Superficially porous particles decrease the analyte flow path and allow for increased efficiency when compared to fully porous particles.

To demonstrate the superior sensitivity of narrow-bore columns, buprenorphine was investigated on Raptor Biphenyl 50 mm columns. In the following example (Figure 1A), 1 µL of 50 ppb buprenorphine was injected and analyzed by the outlined method using a 2.1 mm internal diameter column. Next, the same method was used on a 4.6 mm internal diameter column, adjusting the flow rate from 0.6 mL/min to 0.9 mL/min and injecting the same sample (Figure 1B). When the same amount of sample is injected on both columns, the larger-bore column produces approximately half of the peak height/sensitivity as the smaller-bore column. The sensitivity of analytes can be increased by injecting more on the larger ID column. In this example, when injecting twice as much onto the 4.6 mm ID column (2 µL), approximately the same sensitivity can be achieved (Figure 1C). The downside to increasing injection volume is the increased introduction of matrix on column that can adversely affect chromatography, decrease column lifetime, and enhance matrix interferences.

Figure 1: Buprenorphine was analyzed and injected onto these 50 mm Biphenyl columns:

A: 1 µL on 2.1 mm ID column
B: 1 µL on a 4.6 mm ID column
C: 2 µL on a 4.6 mm ID column

Buprenorphine on Raptor Biphenyl (Figure 1 for Biphenyl Advantage Article)

LC_CF0794

Peaks

PeaksPrecursorProduct Ion 1Product Ion 2
1.Buprenorphine468.3 55.1 414.2

Conditions

ColumnSee notes
Temp.:45 °C
Standard/Sample
Diluent:50:50 Water:methanol
Conc.:50 ng/mL
Mobile Phase
A:Water, 0.1% formic acid
B:Methanol, 0.1% formic acid
Time (min)%A%B
0.009010
6.002575
7.000100
8.000100
8.019010
9.009010
Flow:0.6-0.9 mL/min
DetectorShimadzu 8060
Ion Source:Electrospray
Ion Mode:ESI+
InstrumentShimadzu Nexera X2
Sample PreparationStandards were aliquoted into 2 mL, screw-thread vials (cat.# 21143) and capped with short-cap, screw-vial closures (cat.# 24498).
NotesFigure 1A
Column: Raptor Biphenyl 50 x 2.1 mm, 2.7 µm (cat.# 9309A52)
Guard: Raptor Biphenyl EXP guard column cartridge 5 mm, 2.1 mm ID, 2.7 µm (cat.# 9309A0252)
Inj. Vol.: 1 µL
Flow (mL/min): 0.6

Figure 1B
Column: Raptor Biphenyl 50 x 4.6 mm, 2.7 µm (cat.# 9309A55)
Guard: Raptor Biphenyl EXP guard column cartridge 5 mm, 4.6 mm ID, 2.7 µm (cat.# 9309A0250)
Inj. Vol.: 1 µL
Flow (mL/min): 0.9

Figure 1C
Column: Raptor Biphenyl 50 x 4.6 mm, 2.7 µm (cat.# 9309A55)
Guard: Raptor Biphenyl EXP guard column cartridge 5 mm, 4.6 mm ID, 2.7 µm (cat.# 9309A0250)
Inj. Vol.: 2 µL
Flow (mL/min): 0.9


Sensitivity can become even more challenging when applied to matrix. In the next example buprenorphine is spiked at 2 ng/mL into urine, hydrolyzed, diluted 30-fold, and centrifuged.  

Figure 2: Buprenorphine was spiked into urine at 2 ng/mL and analyzed and injected onto these 50 mm Biphenyl columns: 

A: 5 µL on 2.1 mm ID column
B: 5 µL on a 4.6 mm ID column
C: 10 µL on a 4.6 mm ID column

Buprenorphine on Raptor Biphenyl (Figure 2 for Biphenyl Advantage Article)

LC_CF0797

Peaks

PeaksPrecursorProduct Ion 1Product Ion 2Product Ion 3
1.Buprenorphine468.3 55.1 414.2396.4

Conditions

ColumnSee notes
Temp.:45 °C
Standard/Sample
Diluent:90:10 Water:methanol containing 0.1% formic acid
Conc.:2 ng/mL in urine, diluted 30-fold
Mobile Phase
A:Water, 0.1% formic acid
B:Methanol, 0.1% formic acid
Time (min)%A%B
0.009010
6.002575
7.000100
8.000100
8.019010
9.009010
Flow:0.6-0.9 mL/min
DetectorShimadzu 8060
Ion Source:Electrospray
Ion Mode:ESI+
InstrumentShimadzu Nexera X2
Sample PreparationBuprenorphine was spiked at 2 ng/mL into urine, hydrolyzed, diluted 30-fold, and centrifuged. The sample was aliquoted into 2 mL, screw-threat vials (cat.# 21143) and capped with short-cap, screw-vial closures (cat.# 24498).
NotesFigure 2A
Column: Raptor Biphenyl 50 x 2.1 mm, 2.7 µm (cat.# 9309A52)
Guard: Raptor Biphenyl EXP guard column cartridge 5 mm, 2.1 mm ID, 2.7 µm (cat.# 9309A0252)
Inj. Vol.: 5 µL
Flow (mL/min): 0.6

Figure 2B
Column: Raptor Biphenyl 50 x 4.6 mm, 2.7 µm (cat.# 9309A55)
Guard: Raptor Biphenyl EXP guard column cartridge 5 mm, 4.6 mm ID, 2.7 µm (cat.# 9309A0250)
Inj. Vol.: 5 µL
Flow (mL/min): 0.9

Figure 2C
Column: Raptor Biphenyl 50 x 4.6 mm, 2.7 µm (cat.# 9309A55)
Guard: Raptor Biphenyl EXP guard column cartridge 5 mm, 4.6 mm ID, 2.7 µm (cat.# 9309A0250)
Inj. Vol.: 10 µL
Flow (mL/min): 0.9

Figure 2 outlines the importance of sensitivity when analyzing matrix. The injection volume must be increased to 5 µL to be able to detect 2 ng/mL in urine on a 2.1 mm ID column. When analyzing the same amount (5 µL) on a 4.6 mm ID column, buprenorphine could not be detected. To achieve approximately the same amount of sensitivity, the injection volume had to be increased to 10 µL. The increase in sample volume also means an increase in matrix being injected onto column. To best conserve sample and maximize sensitivity, narrow-bore columns are favored. 

To produce precise and rugged quantitative methods, a resolution of 1.5 or greater should be achieved. In the next example, nine groups of drugs of abuse isobars were analyzed on a Raptor Biphenyl 50 x 2.1 mm column and their resolutions calculated using Equation 2.

Figure 3: Chromatogram and Conditions for the Analysis of Isobar Panel
Drugs of Abuse on Raptor Biphenyl (Figure 3 for Biphenyl Advantage Article)

LC_CF0800

Peaks

PeakstR (min)PrecursorProduct Ion 1Product Ion 2
1.Morphine0.85286.2152.1165.0
2.Oxymorphone0.92302.1227.2198.2
3.Hydromorphone1.08286.2184.9156.9
4.Methamphetamine1.51150.391.1119.0
5.Noroxycodone1.63302.1227.0197.9
6.Naloxone1.64328.3310.1212.3
7.Phentermine1.65150.291.1133.1
8.Norhydrocodone1.73286.1199.0128.2
9.Codeine1.76300.2152.0165.1
10.6-Acetylmorphine1.82328.2165.0211.0
11.Hydrocodone2.03300.1199.0128.0
PeakstR (min)PrecursorProduct Ion 1Product Ion 2
12.O-Desmethylvenlafaxine2.17264.158.1107.0
13.Lamotrigine2.26256.0211.1145.0
14.Hydroxybupropion2.52256.0130.2166.0
15.Tramadol2.74264.258.077.1
16.7-Aminoclonazepam3.09286.1121.2250.1
17.Mirtazapine3.33266.1195.172.1
18.Citalopram3.93325.2109.1262.0
19.Nortriptyline4.65264.191.1115.2
20.Alpha-hydroxyalprazolam5.42325.1297216.2
21.CBD6.74315.0193.0123.0
22.9-THC7.03315.2193.0123.2

Conditions

ColumnRaptor Biphenyl (cat.# 9309A52)
Dimensions:50 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Guard Column:Raptor Biphenyl EXP guard column cartridge 5 mm, 2.1 mm ID, 2.7 µm (cat.# 9309A0252)
Temp.:45 °C
Standard/Sample
Diluent:50:50 Water:methanol
Conc.:50 ng/mL
Inj. Vol.:1 µL
Mobile Phase
A:Water, 0.1% formic acid
B:Methanol, 0.1% formic acid
Time (min)Flow (mL/min)%A%B
0.000.69010
6.000.62575
7.000.60100
8.000.60100
8.010.69010
9.000.69010
DetectorShimadzu 8060
Ion Source:Electrospray
Ion Mode:ESI+
InstrumentShimadzu Nexera X2
Sample PreparationThe sample was aliquoted into 2 mL, screw-threat vials (cat.# 21143) and capped with short-cap, screw-vial closures (cat.# 24498).
Equation 2:
Where Rt is the retention time for the compounds that are being compared and W is the width of peak.

Where Rt is the retention time for the compounds that are being compared and W is the width of the peak at baseline.

Table I: Compound Name; Shared Molecular Weight; Analyte Retention Time (Min); Peak Width; and Calculated Resolution between Isobar Groups Using Equation 2

Isobar GroupNameMolecular Weight (g/mol)Retention Time (min)Peak WidthResolution
1Methamphetamine149.231.510.0911.5
Phentermine1.650.096
2Oxymorphone301.340.920.0947.5
Noroxycodone1.630.094
3Citalopram324.393.930.10813.7
Alpha-hydroxyalprazolam5.420.109
4Naloxone327.271.640.0992.0
6-acetylmorphine1.820.082
5Morphine285.340.850.1022.3
Hydromorphone1.080.097
Norhydrocodone1.730.0857.1
7-aminoclonazepam3.090.09914.9
6Lamotrigine256.092.260.0932.7
Hydroxybupropion255.742.520.095
7Codeine299.361.760.0883.1
Hydrocodone2.030.089
8O-desmethylvenlafaxine263.372.170.0916.3
Tramadol2.740.091
Mirtazapine3.330.0976.3
Nortriptyline4.650.11012.7
9CBD314.476.740.1022.8
9-THC7.030.104

In this example, all pairs of isobars return a resolution value of 1.5 or greater and are retained well on a 2.1 mm ID column. Several contributing factors are at play to achieve acceptable resolution, including the superior selectivity of the biphenyl phase, the pairing to superficially porous particles, and the narrow-bore column mitigating band-broadening contributions.

The Green Advantage

The use of narrow-bore column dimensions oftentimes means greener solutions, which in this case means being more environmentally friendly while also reducing costs. Smaller ID columns can be advantageous over larger diameter columns when considering the consumption of solvent, as typically the larger the column internal diameter the higher the flow rate. Higher flow rates also hinder ionization efficiency for MS detection and can reduce sensitivity. In the following example, DOA isobars are compared using two different internal diameter columns, 2.1 and 4.6 mm.

Figure 4: 22 Isobar Drugs of Abuse Compounds Analyzed on a Raptor Biphenyl Column Using:

A: 0.6 mL/min flow rate
B: 0.9 mL/min flow rate

Drugs of Abuse on Raptor Biphenyl (Figure 4 for Biphenyl Advantage Article)

LC_CF0801

Peaks

PeaksPrecursorProduct Ion 1Product Ion 2
1.Morphine286.2152.1165.0
2.Oxymorphone302.1227.2198.2
3.Hydromorphone286.2184.9156.9
4.Methamphetamine150.391.1119.0
5.Noroxycodone302.1227.0197.9
6.Naloxone328.3310.1212.3
7.Phentermine150.291.1133.1
8.Norhydrocodone286.1199.0128.2
9.Codeine300.2152.0165.1
10.6-Acetylmorphine328.2165.0211.0
11.Hydrocodone300.1199.0128.0
PeaksPrecursorProduct Ion 1Product Ion 2
12.O-Desmethylvenlafaxine264.158.1107.0
13.Lamotrigine256.0211.1145.0
14.Hydroxybupropion256.0130.2166.0
15.Tramadol264.258.077.1
16.7-Aminoclonazepam286.1121.2250.1
17.Mirtazapine266.1195.172.1
18.Citalopram325.2109.1262.0
19.Nortriptyline264.191.1115.2
20.Alpha-hydroxyalprazolam325.1297216.2
21.CBD315.0193.0123.0
22.9-THC315.2193.0123.2

Conditions

ColumnSee notes
Temp.:45 °C
Standard/Sample
Diluent:50:50 Water:methanol
Conc.:50 ng/mL
Mobile Phase
A:Water, 0.1% formic acid
B:Methanol, 0.1% formic acid
Time (min)%A%B
0.009010
6.002575
7.000100
8.000100
8.019010
9.009010
Flow:0.6-0.9 mL/min
DetectorShimadzu 8060
Ion Source:Electrospray
Ion Mode:ESI+
InstrumentShimadzu Nexera X2
Sample PreparationStandards were aliquoted into 2 mL, screw-threat vials (cat.# 21143) and capped with short-cap, screw-vial closures (cat.# 24498).
NotesFigure 4A
Column: Raptor Biphenyl 50 x 2.1 mm, 2.7 µm (cat.# 9309A52)
Guard: Raptor Biphenyl EXP guard column cartridge 5 mm, 2.1 mm ID, 2.7 µm (cat.# 9309A0252)
Inj. Vol.: 1 µL
Flow (mL/min): 0.6

Figure 4B
Column: Raptor Biphenyl 50 x 4.6 mm, 2.7 µm (cat.# 9309A55)
Guard: Raptor Biphenyl EXP guard column cartridge 5 mm, 4.6 mm ID, 2.7 µm (cat.# 9309A0250)
Inj. Vol.: 2 µL
Flow (mL/min): 0.9

While the run time does not increase in this example, which can sometimes be the case when transferring to a larger-bore column, the flow rate did need to be adjusted from 0.6 mL/min to 0.9 mL/min. This means that the consumption of solvent has increased by one third per minute, costing more money than the smaller bore column, and generating more waste. Since acceptable resolution is already achieved on the 2.1 mm ID column, there is minimal advantage to moving to larger column IDs.  In the long run, smaller columns will ultimately result in less solvent consumption and more cost-efficient solutions.

Robustness in matrix

To demonstrate the narrow-bore column’s robustness in matrix over time, 1000 injections were performed on a 2.1 mm ID column in urine matrix. The goal of these experiments was to demonstrate column robustness in matrix showing four different metrics: sensitivity, instrument pressure, resolution of a critical pair, and performance of an early-eluting compound. To demonstrate sensitivity, buprenorphine was spiked into urine at 2 ng/mL and diluted 30-fold. 

Figure 5: Lifetime Sensitivity Study of 1000 Injections of Buprenorphine in Urine.
Buprenorphine on Raptor Biphenyl (Figure 5 Robustness Test for Biphenyl Advantage Article)

LC_CF0802

Figure 5 shows the results of the sensitivity study over lifetime for buprenorphine. Buprenorphine was spiked into urine at 2 ng/mL and diluted in solvent 30-fold. Injection 1 in the lifetime study returned 1464 counts for peak area and 487 counts for peak height. Injection 1000 returned 1402 counts for peak area and 430 counts for height. The percent difference over the lifetime of the column was calculated to be 4% for peak area between injection 1 and 1000, demonstrating that sensitivity of the analyte has not deteriorated over time. 

To demonstrate resolution stability over a 1000 injection lifetime study, phentermine and methamphetamine were spiked into urine at 2 ng/mL and diluted 30-fold. The retention time and peak width were extracted for injection 1 and 1000 and their resolution calculated. Resolution between the critical pairs was able to be maintained over the lifetime study and can be seen in Figure 6.

Figure 6: Lifetime Resolution Study of 1000 Injections of a Critical Pair, Phentermine and Methamphetamine, in Urine
Methamphetamine and Phentermine on Raptor Biphenyl (Figure 6 Robustness Test for Biphenyl Advantage Article)

LC_CF0803

An early-eluting compound, morphine, was spiked into urine and shown over the lifetime study to demonstrate robustness in matrix and can be seen in Figure 7. While there is a decrease in sensitivity across 1000 injections, the column still demonstrates comparable retention to the first injection.

Figure 7: Lifetime Study of 1000 Injections of an Early-Eluting Compound, Morphine, in Urine
Morphine on Raptor Biphenyl (Figure 7 for Biphenyl Advantage Article)

LC_CF0804

Finally, pressure profiles were collected from runs 1 and 1000 to compare. The pressure profiles were nearly identical, as seen in Figure 8, and demonstrate that debris is not being collected, and that the narrow-bore column is able to analyze matrix samples without increasing pressure or clogging. 

Figure 8: Pressure Profiles for Runs 1 and 1000 in Urine Matrix on Raptor Biphenyl 50 x 2.1 mm, 2.7 µm Column
figure article CFAR4190 08b

Conclusions

When choosing a column dimension for analysis, it’s important to keep in mind the benefits of narrow-bore columns. Small ID columns can be beneficial in the reduction of solvent and sample consumption for greener solutions and cost savings. The 2.1 mm ID columns demonstrate superior sensitivity over the 4.6 mm ID columns due to column efficiency, less analyte dilution, and reduced band broadening, while still achieving acceptable resolution for critical pairs. Even in matrix, the narrow-bore columns are reproducible and robust after repeat injections without increasing backpressure or clogging.

References

  1. S. Lupo, The big pain: Development of pain-free methods for analyzing 231 multiclass drugs and metabolites by LC-MS/MS, Application note, CFAR2309-UNV, Restek Corporation, 2015. https://www.restek.com/articles/the-big-pain-development-of-pain-free-methods-for-analyzing-231-multiclass-drugs-and-metabolites-by-lc-msms
  2. S. Lupo, LC-MS sensitivity: Practical strategies to boost your signal and lower your noise, LC GC N. Am., 36 (2018) (9) 652-660. https://www.chromatographyonline.com/view/lc-ms-sensitivity-practical-strategies-boost-your-signal-and-lower-your-noise

Related Products


Raptor Biphenyl, 2.7 µm, 50 x 2.1 mm HPLC Column
Raptor Inert Biphenyl, 2.7 µm, 50 x 2.1 mm HPLC Column
Raptor Biphenyl, 2.7 µm, 50 x 4.6 mm HPLC Column
Raptor Biphenyl EXP Guard Column Cartridge, 2.7 µm, 5 x 2.1 mm, 3-pk.
Raptor Biphenyl EXP Guard Column Cartridge, 2.7 µm, 5 x 4.6 mm, 3-pk.
EXP Direct Connect Holder for EXP Guard Cartridges, Includes Fitting & Ferrules

Authors

  • Jamie York is a principal scientist in the Applications Lab at Restek Corporation. She leads the development of innovative analytical methods tailored to the food, clinical, environmental, and cannabis markets. Jamie earned her PhD in chemistry from The University of Texas at Arlington, where she gained extensive expertise in a range of analytical techniques, including gas chromatography–vacuum ultraviolet (GC–VUV); gas chromatography–mass spectrometry (GC–MS); matrix-assisted laser desorption/ionization (MALDI); and liquid chromatography– mass spectrometry (LC–MS/MS); with a research emphasis on food and environmental analysis. Today, her work focuses on complex method development and advanced sample preparation strategies to support the evolving needs of the scientific community.  

    View all posts
  • Samantha Herbick

    Samantha Herbick is an applications scientist within the LC Solutions Department at Restek. Her primary focus is on the development of applications in the toxicology and life science markets. She attended Duquesne University where she earned a bachelor’s degree in Biochemistry and a master's degree in forensic science and law. Prior to joining Restek, Samantha worked as a scientist and method developer in a forensic toxicology lab. In this role, she performed analysis on toxicology casework and worked on the development and validation of new and existing assays using LC-MS/MS, GC-MS, and GC-FID.

    View all posts
CFAR4190