博客

Cryogenic Cooling for Air Analysis Part 2 – Combining TO-15A and Ethylene Oxide

30 Apr 2020

In my previous blog on cryogenic cooling I touched briefly on ethylene oxide (EtO) and why it may be of interest in ambient air analysis. While OSHA has a time weighted average (TWA) limit for EtO at 1ppm for an 8 hour exposure, recent work by the US EPA has shown that even low doses of ethylene oxide can increase cancer risks over a person’s lifetime (https://cfpub.epa.gov/ncea/iris/iris_documents/documents/toxreviews/1025tr.pdf), which explains the interest in testing for EtO at sub ppb levels. Since this brings EtO testing to similar levels as TO-15A, why not combine the two?

While my previous blog covered the need to separate EtO from the air peaks introduced into the instrument, there are several other potential interferences that have to be managed. Acetaldehyde has an almost identical structure and mass spectrum. It is very common in nature and produced in a wide variety of industries, and it is possible to be present in both lab blanks and samples. Methanol also shares several ions with EtO, and as a common solvent for volatile standards (e.g., 8260 and TO-15A internal standards) it’s likely present in most air labs.

blog cryogenic cooling for air analysis part 2 combining to 15a and ethylene oxide 01

Fig. 1 – Comparison of EtO, Acetaldehyde, and methanol mass spectrum

Fortunately the cryo cooling helps with these separations as well. In addition, I also found that using selected ion monitoring (SIM) produced a cleaner baseline and better signal to noise ratio, allowing for detection of EtO down to 0.05ppb or lower (Fig.2 lower trace).

blog cryogenic cooling for air analysis part 2 combining to 15a and ethylene oxide 02

Fig. 2 – Comparison of Scan (top) and SIM (bottom) signals for EtO (RT ~8.72) with acetaldehyde (RT ~7.50) and MeOH (RT ~9.00) interferences. EtO at 0.05ppb

Once the troublesome EtO/acetaldehyde/methanol separations are solved with cryo cooling, I was able to use EZGC to get a working oven program to separate the TO-15A compound list. Without the need for extra sensitivity on the TO-15A compounds I found it helpful to use the combined SIM/Scan capabilities of the Agilent 5977A mass spec, using the SIM data for EtO and the scan data for the TO-15A list. This meant I didn’t have to optimize the SIM parameters for nearly 80 compounds, keeping the method much simpler.

blog cryogenic cooling for air analysis part 2 combining to 15a and ethylene oxide 03
blog cryogenic cooling for air analysis part 2 combining to 15a and ethylene oxide 04
blog cryogenic cooling for air analysis part 2 combining to 15a and ethylene oxide 05

Fig. 3 – Combined TO-15A and EtO chromatogram with EIC for compounds 1-7 (top), SIM for EtO ( compound 8, middle), and TIC (compounds 9-79, bottom). TO-15A compounds at 0.2ppb, EtO at 0.05ppb.

#NameRet Time
1Propylene4.17
2Dichlorodifluoromethane4.43
31,2-Dichlorotetrafluoroethane5.45
4Chloromethane5.62
5n-Butane6.52
6Vinyl chloride6.54
71,3-Butadiene6.87
8Ethylene Oxide8.72
9Bromomethane8.75
10Chloroethane9.64
11Vinyl bromide10.71
12Trichlorofluoromethane11.21
13n-Pentane11.85
14Ethanol13.29
15Acrolein13.74
161,1-Dichloroethene13.94
171,1,2-Trichlorotrifluoroethane14.3
18Carbon disulfide14.49
19Acetone14.55
20Acetonitrile15.83
21Isopropyl alcohol15.92
22Methylene chloride16.5
23trans-1,2-Dichloroethene17.6
24Tertiary butanol17.67
25Methyl tert-butyl ether (MTBE)17.73
26Hexane18.8
271,1-Dichloroethane19.35
28Vinyl acetate19.65
29cis-1,2-Dichloroethene21.49
302-Butanone (MEK)21.62
31Ethyl acetate21.9
32Bromochloromethane22.29
33Tetrahydrofuran22.35
34Chloroform22.74
351,1,1-Trichloroethane23
36Cyclohexane23.12
37Carbon tetrachloride23.35
38Benzene23.8
391,2-Dichloroethane23.96
40Isooctane24.09
41Heptane24.45
421,4-Difluorobenzene24.66
43Trichloroethylene24.98
441,1,2-Trichloroethane24.98
451,2-Dichloropropane25.36
46Methyl methacrylate25.49
471,4-Dioxane25.49
48Bromodichloromethane25.75
49cis-1,3-Dichloropropene26.28
504-Methyl-2-2pentanone (MIBK)26.46
51Toluene26.64
52trans-1,3-Dichloropropene26.91
53Tetrachloroethene27.18
542-Hexanone (MBK)27.32
55Dibromochloromethane27.49
561,2-Dibromoethane27.6
57Chlorobenzene-d528.02
58Chlorobenzene28.04
59Ethylbenzene28.11
60n-Nonane28.2
61m- & p-Xylene28.22
62o-Xylene28.55
63Styrene28.56
64Bromoform28.74
65Cumene28.83
664-Bromofluorobenzene28.99
671,1,2,2-Tetrachloroethane29.08
68n-Propyl benzene29.16
694-Ethyltoluene29.24
702-Chlorotoluene29.25
711,3,5-Trimethylbenzene29.28
721,2,4-Trimethylbenzene29.57
731,3-Dichlorobenzene29.82
741,4-Dichlorobenzene29.89
75Benzyl chloride29.97
761,2-Dichlorobenzene30.17
771,2,4-Trichlorobenzene31.36
78Hexachlorobutadiene31.39
79Naphthalene31.6

Table 1 – RT for TO-15A and EtO.

GCAgilent 7890B
Injection typeOn-column
Column624Sil MS 60m x 0.25mm x 1.4um
Carrier gasHe , constant flow
Flow rate2mL/min
Oven temp0°C (hold 5 min) to 60°C at 3.5°C/min (hold 0 min) to 260°C at 24°C/min (hold 5 min)
DetectorMS (Agilent 5977A)
Acquisition modeSIM/Scan
Scan parameters 
Scan range (amu)29-226
Scan rate (scans/sec)3.7
SIM parameters 
SIM ions15, 29, 43, 44, 56
Dwell time50
Transfer line250°C
Analyzer typeQuadruple
Source typeExtractor
Source temp230°C
Quad temp150°C
Electron energy70eV
Solvent delay time0 min
Tune typeBFB
Ionization modeEI
PreconcentratorMarkes Unity 1+ CIA
Trap 1 settings 
Cooling temp5°C
Desorb temp300°C
Desorb flow6 mL/min
Desorb time180 sec
Internal Standard 
Purge flow50 mL/min
Purge time60 sec
Volume50mL
ISTD  flow50mL/min
Sample 
Volume400mL
Purge flow50mL/min
Purge time60 sec
Sample flow100mL/min

Table 2 – GC/MS and preconcentrator settings

While many labs may be reluctant to use cryogenic cooling due to costs and safety issues, it can be a powerful tool to separate out very volatile compounds. Here it was critical in the separation of EtO from methanol and acetaldehyde. In addition, the ability to acquire both SIM and scan MS data allowed for the increased signal to noise ratio for EtO in SIM mode, while maintaining the simplicity full scan for the TO-15A compounds. Together, cyro cooling and SIM/Scan can allow for the relatively simple addition of EtO down to 50ppt to TO-15A analysis.

作者

  • Jason Hoisington

    Jason Hoisington received his bachelor’s degree in general science with a focus on chemistry from the University of Alaska, Fairbanks. He worked for SGS Environmental for seven years in environmental soil and water testing, developing methods for the analysis of volatiles and semivolatile organics to include pesticides and polychlorinated biphenyls (PCBs). In 2012, Jason moved on to lab and application support for Dow Chemical Company, providing advanced analytical troubleshooting and method development. In 2019, Jason joined Restek and has focused on air applications.

    查阅所有文章
GNBL4636