FAQ

What size SPE cartridge do I need?

30 Jul 2025

Cartridge size should be selected based on sample volume while the bed weight of the packing is determined by the compound load (analytes of interest and interferences). Choose the SPE cartridge that is best for your application based on the following table.

Sample VolumeCompound LoadCartridge SizeBed Weight
1-10 mL2-6 mg1 mL100-300 mg
10-100 mL6-1000 mg3 mL300-500 mg
100 mL – 1 L>1000 mg6 mL>500 mg

If the sample volume exceeds the capacity of the SPE cartridge being used, then a larger, empty SPE cartridge can be used along with a connector to hold the additional volume. Alternatively, the Resprep Sample Delivery System can be used.

作者

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    查阅所有文章
GNOT5142