{"id":43518,"date":"2020-10-15T00:00:00","date_gmt":"2020-10-15T00:00:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/fast-analysis-of-polycyclic-aromatic-hydrocarbons-pahs-by-hplc-uv\/"},"modified":"2025-12-08T14:23:33","modified_gmt":"2025-12-08T14:23:33","slug":"fast-analysis-of-polycyclic-aromatic-hydrocarbons-pahs-by-hplc-uv","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/zh-hans\/articles\/evss2396\/fast-analysis-of-polycyclic-aromatic-hydrocarbons-pahs-by-hplc-uv","title":{"rendered":"Fast Analysis of Polycyclic Aromatic Hydrocarbons (PAHs) by HPLC-UV"},"content":{"rendered":"\n<ul class=\"wp-block-list\">\n<li>Quick ~15-minute analysis of PAHs by HPLC provides fast results without the expense of UHPLC equipment.<\/li>\n\n\n\n<li>Baseline resolution improves accuracy for commonly analyzed PAHs of global concern.<\/li>\n\n\n\n<li>Restek has method-specific PAH certified reference materials, as well as the only standard formulated to contain all major PAH interferences for priority PAHs.<\/li>\n<\/ul>\n\n\n\n<p class=\"wp-block-paragraph\">Polycyclic aromatic hydrocarbons (PAHs) are a group of nonpolar organic molecules containing two or more benzene rings. PAHs are released from the incomplete combustion of organic fuels, and they generally occur as complex mixtures in our food and environment. Although not all polycyclic aromatic hydrocarbons are known to have detrimental health effects, several PAHs have been demonstrated to be toxic, carcinogenic, or mutagenic. To protect human health, many organizations have mandated exposure limits for PAHs. Air, water, soil, and food samples are commonly tested globally; however, the list of target compounds varies widely around the world. Many HPLC methods recommend a C18 column, but analysis times can be long (~30 minutes) and target analytes are not always completely resolved.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Restek\u2019s Pinnacle II PAH column was developed as an alternative to a C18 column specifically for the analysis of PAHs by HPLC. The stationary phase incorporates a proprietary bonding and unique selectivity that baseline resolves key PAH isomers, such as those in the European and U.S. PAH lists shown in the chromatograms below. These separations were accomplished in short analysis times (~15 minutes) on a conventional HPLC system (&lt;400 bar pressure). Analyzing PAHs by HPLC with a Pinnacle II PAH column allows labs to report PAH results quickly and accurately without investing in UHPLC equipment.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/lc_ff0529.png\" alt=\"EU 15+1 PAHs on Pinnacle II PAH (UV\/Vis)\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> LC_FF0529<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12 peak-50\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzo[c]fluorene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/205-12-9\/Benzo[c]fluorene\" target=\"_blank\" rel=\"noopener\">Benzo[c]fluorene<\/a><\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Cyclopenta[cd]pyrene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/27208-37-3\/Cyclopenta[cd]pyrene\" target=\"_blank\" rel=\"noopener\">Cyclopenta[cd]pyrene<\/a><\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzo[a]anthracene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/56-55-3\/Benzo[a]anthracene\" target=\"_blank\" rel=\"noopener\">Benzo[a]anthracene<\/a><\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Chrysene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/218-01-9\/Chrysene\" target=\"_blank\" rel=\"noopener\">Chrysene<\/a><\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 5-Methylchrysene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/3697-24-3\/5-Methylchrysene\" target=\"_blank\" rel=\"noopener\">5-Methylchrysene<\/a><\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzo(j)fluoranthene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/205-82-3\/Benzo(j)fluoranthene\" target=\"_blank\" rel=\"noopener\">Benzo(j)fluoranthene<\/a><\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzo[b]fluoranthene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/205-99-2\/Benzo[b]fluoranthene\" target=\"_blank\" rel=\"noopener\">Benzo[b]fluoranthene<\/a><\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzo[k]fluoranthene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/207-08-9\/Benzo[k]fluoranthene\" target=\"_blank\" rel=\"noopener\">Benzo[k]fluoranthene<\/a><\/td><\/tr>\n<\/tbody><\/table>\n<table class=\"peaks col-lg-6 col-12 peak-50\"><thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><\/tr><\/thead><tbody>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzo[a]pyrene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/50-32-8\/Benzo[a]pyrene\" target=\"_blank\" rel=\"noopener\">Benzo[a]pyrene<\/a><\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dibenzo[a,l]pyrene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/191-30-0\/Dibenzo[a,l]pyrene\" target=\"_blank\" rel=\"noopener\">Dibenzo[a,l]pyrene<\/a><\/td><\/tr>\n<tr><td class=\"num\">11.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dibenz[a,h]anthracene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/53-70-3\/Dibenz[a,h]anthracene\" target=\"_blank\" rel=\"noopener\">Dibenz[a,h]anthracene<\/a><\/td><\/tr>\n<tr><td class=\"num\">12.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzo[ghi]perylene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/191-24-2\/Benzo[ghi]perylene\" target=\"_blank\" rel=\"noopener\">Benzo[ghi]perylene<\/a><\/td><\/tr>\n<tr><td class=\"num\">13.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Indeno[1,2,3-cd]pyrene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/193-39-5\/Indeno[1,2,3-cd]pyrene\" target=\"_blank\" rel=\"noopener\">Indeno[1,2,3-cd]pyrene<\/a><\/td><\/tr>\n<tr><td class=\"num\">14.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dibenzo[a,e]pyrene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/192-65-4\/Dibenzo[a,e]pyrene\" target=\"_blank\" rel=\"noopener\">Dibenzo[a,e]pyrene<\/a><\/td><\/tr>\n<tr><td class=\"num\">15.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dibenzo(a,i)pyrene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/189-55-9\/Dibenzo(a,i)pyrene\" target=\"_blank\" rel=\"noopener\">Dibenzo(a,i)pyrene<\/a><\/td><\/tr>\n<tr><td class=\"num\">16.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dibenzo(a,h)pyrene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/189-64-0\/Dibenzo(a,h)pyrene\" target=\"_blank\" rel=\"noopener\">Dibenzo(a,h)pyrene<\/a><\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Pinnacle II PAH  (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/921946E\" rel=\"noopener\">cat.# 921946E<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Dimensions:<\/th><td>150 mm x 3.0 mm ID<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Particle Size:<\/th><td>4 \u00b5m<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Pore Size:<\/th><td>110 \u00c5<\/td><\/tr><tr><td><\/td><tr><th class=\"sub conditions_header\" scope=\"row\">Temp.:<\/th><td>30 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td>EU 15+1 PAH standard (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/32470\" rel=\"noopener\">cat.# 32470<\/a>)<\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>Acetonitrile<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>10 \u00b5g\/mL (ppm)<\/td><\/tr><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>5 \u00b5L <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Mobile Phase<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">A:<\/th><td>Water <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">B:<\/th><td>Acetonitrile:tetrahydrofuran (95:5) <\/td><\/tr><tr><td><\/td><td><table class=\"cgram_ramp\"><thead><tr><th>Time (min)<\/th><th>Flow (mL\/min)<\/th><th>%A<\/th><th>%B<\/th><\/tr><\/thead><tbody><tr><td>0.00<\/td><td>1.5<\/td><td>30<\/td><td>70<\/td><\/tr><tr><td>3.00<\/td><td>1.5<\/td><td>25<\/td><td>75<\/td><\/tr><tr><td>10.00<\/td><td>1.5<\/td><td>0<\/td><td>100<\/td><\/tr><tr><td>14.50<\/td><td>1.5<\/td><td>0<\/td><td>100<\/td><\/tr><tr><td>14.51<\/td><td>1.5<\/td><td>30<\/td><td>70<\/td><\/tr><tr><td>16.5<\/td><td>1.5<\/td><td>30<\/td><td>70<\/td><\/tr><\/tbody><\/table><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Max Pressure:<\/th><td>275 bar<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>UV\/Vis @ 292, 4 nm<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Cell Temp:<\/th><td>40 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Flow Cell Size:<\/th><td>10 \u00b5L <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Shimadzu Prominence<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>32470<\/li><li>921946E<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>15+1<\/li><li>EU 15+1<\/li><li>PAH<\/li><li>UV\/Vis<\/li><li>european union<\/li><li>pinnacle<\/li><li>pinnacle II<\/li><li>polycyclic aromatic hydrocarbons<\/li><li>polynuclear aromatic hydrocarbons<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/lc_ff0529.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/lc_ev0537.png\" alt=\"EPA 8310 PAHs on Pinnacle II PAH by LC-UV\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> LC_EV0537<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12 peak-50\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 50px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 50px\">Conc.<br \/>(\u00b5g\/mL)<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Naphthalene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/91-20-3\/Naphthalene\" target=\"_blank\" rel=\"noopener\">Naphthalene<\/a><\/td><td class=\"oth\">2.68<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acenaphthylene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/208-96-8\/Acenaphthylene\" target=\"_blank\" rel=\"noopener\">Acenaphthylene<\/a><\/td><td class=\"oth\">3.09<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1-Methylnaphthalene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/90-12-0\/1-Methylnaphthalene\" target=\"_blank\" rel=\"noopener\">1-Methylnaphthalene<\/a><\/td><td class=\"oth\">3.39<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Methylnaphthalene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/91-57-6\/2-Methylnaphthalene\" target=\"_blank\" rel=\"noopener\">2-Methylnaphthalene<\/a><\/td><td class=\"oth\">3.56<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acenaphthene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/83-32-9\/Acenaphthene\" target=\"_blank\" rel=\"noopener\">Acenaphthene<\/a><\/td><td class=\"oth\">3.74<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Fluorene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/86-73-7\/Fluorene\" target=\"_blank\" rel=\"noopener\">Fluorene<\/a><\/td><td class=\"oth\">3.91<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Phenanthrene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/85-01-8\/Phenanthrene\" target=\"_blank\" rel=\"noopener\">Phenanthrene<\/a><\/td><td class=\"oth\">4.40<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Anthracene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/120-12-7\/Anthracene\" target=\"_blank\" rel=\"noopener\">Anthracene<\/a><\/td><td class=\"oth\">4.93<\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Fluoranthene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/206-44-0\/Fluoranthene\" target=\"_blank\" rel=\"noopener\">Fluoranthene<\/a><\/td><td class=\"oth\">5.49<\/td><td class=\"oth\">10<\/td><\/tr>\n<\/tbody><\/table>\n<table class=\"peaks col-lg-6 col-12 peak-50\"><thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 50px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 50px\">Conc.<br \/>(\u00b5g\/mL)<\/th><\/tr><\/thead><tbody>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Pyrene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/129-00-0\/Pyrene\" target=\"_blank\" rel=\"noopener\">Pyrene<\/a><\/td><td class=\"oth\">5.91<\/td><td class=\"oth\">10<\/td><\/tr>\n<tr><td class=\"num\">11.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzo[a]anthracene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/56-55-3\/Benzo[a]anthracene\" target=\"_blank\" rel=\"noopener\">Benzo[a]anthracene<\/a><\/td><td class=\"oth\">7.39<\/td><td class=\"oth\">10<\/td><\/tr>\n<tr><td class=\"num\">12.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Chrysene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/218-01-9\/Chrysene\" target=\"_blank\" rel=\"noopener\">Chrysene<\/a><\/td><td class=\"oth\">7.77<\/td><td class=\"oth\">10<\/td><\/tr>\n<tr><td class=\"num\">13.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzo[b]fluoranthene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/205-99-2\/Benzo[b]fluoranthene\" target=\"_blank\" rel=\"noopener\">Benzo[b]fluoranthene<\/a><\/td><td class=\"oth\">8.95<\/td><td class=\"oth\">10<\/td><\/tr>\n<tr><td class=\"num\">14.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzo[k]fluoranthene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/207-08-9\/Benzo[k]fluoranthene\" target=\"_blank\" rel=\"noopener\">Benzo[k]fluoranthene<\/a><\/td><td class=\"oth\">9.52<\/td><td class=\"oth\">5<\/td><\/tr>\n<tr><td class=\"num\">15.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzo[a]pyrene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/50-32-8\/Benzo[a]pyrene\" target=\"_blank\" rel=\"noopener\">Benzo[a]pyrene<\/a><\/td><td class=\"oth\">9.99<\/td><td class=\"oth\">10<\/td><\/tr>\n<tr><td class=\"num\">16.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dibenzo[a,h]anthracene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/53-70-3\/Dibenzo[a,h]anthracene\" target=\"_blank\" rel=\"noopener\">Dibenzo[a,h]anthracene<\/a><\/td><td class=\"oth\">10.75<\/td><td class=\"oth\">10<\/td><\/tr>\n<tr><td class=\"num\">17.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzo[ghi]perylene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/191-24-2\/Benzo[ghi]perylene\" target=\"_blank\" rel=\"noopener\">Benzo[ghi]perylene<\/a><\/td><td class=\"oth\">11.18<\/td><td class=\"oth\">10<\/td><\/tr>\n<tr><td class=\"num\">18.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Indeno[1,2,3-cd]pyrene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/193-39-5\/Indeno[1,2,3-cd]pyrene\" target=\"_blank\" rel=\"noopener\">Indeno[1,2,3-cd]pyrene<\/a><\/td><td class=\"oth\">11.60<\/td><td class=\"oth\">10<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Pinnacle II PAH  (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/921946E\" rel=\"noopener\">cat.# 921946E<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Dimensions:<\/th><td>150 mm x 3.0 mm ID<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Particle Size:<\/th><td>4 \u00b5m<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Pore Size:<\/th><td>110 \u00c5<\/td><\/tr><tr><td><\/td><tr><th class=\"sub conditions_header\" scope=\"row\">Temp.:<\/th><td>30 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td>EPA Method 8310 quality control check (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/31843\" rel=\"noopener\">cat.# 31843<\/a>)<\/td><\/tr>\n<td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>10 \u00b5L <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Mobile Phase<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">A:<\/th><td>Water <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">B:<\/th><td>Acetonitrile <\/td><\/tr><tr><td><\/td><td><table class=\"cgram_ramp\"><thead><tr><th>Time (min)<\/th><th>Flow (mL\/min)<\/th><th>%A<\/th><th>%B<\/th><\/tr><\/thead><tbody><tr><td>0.00<\/td><td>1.2<\/td><td>50<\/td><td>50<\/td><\/tr><tr><td>10.0<\/td><td>1.2<\/td><td>0<\/td><td>100<\/td><\/tr><tr><td>12.5<\/td><td>1.2<\/td><td>0<\/td><td>100<\/td><\/tr><tr><td>12.51<\/td><td>1.2<\/td><td>50<\/td><td>50<\/td><\/tr><tr><td>15.0<\/td><td>1.2<\/td><td>50<\/td><td>50<\/td><\/tr><\/tbody><\/table><\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>UV\/Vis @ 280 nm<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>31843<\/li><li>921946E<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>610<\/li><li>8310<\/li><li>PAH<\/li><li>Polycyclic<\/li><li>aromatic<\/li><li>hydrocarbons<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/lc_ev0537.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<p class=\"wp-block-paragraph\"><\/p>\n","protected":false},"excerpt":{"rendered":"<p>Pinnacle II PAH columns were developed by Restek as an alternative to C18 columns specifically for the analysis of PAHs by HPLC. These reliable columns provide baseline resolution of key PAHs in a fast ~15 minute analysis without the expense of UHPLC equipment.<\/p>\n","protected":false},"author":11,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[7],"tags":[],"industries-application":[2162,2163,2161,2167,2169,2668,2166],"post-badge":[],"resource-type":[2623],"product-library":[2391,2373],"resource-technique":[2299],"ppma_author":[414],"class_list":["post-43518","post","type-post","status-publish","format-standard","hentry","category-articles","industries-application-air-monitoring","industries-application-drinking-water-wastewater-testing","industries-application-environmental","industries-application-food-beverage","industries-application-food-safety","industries-application-semivolatiles","industries-application-soil-sediment-biota-testing","resource-type-featured-application","product-library-lc-columns","product-library-liquid-chromatography-products","resource-technique-liquid-chromatography"],"acf":[],"taxonomy_info":{"category":[{"value":7,"label":"Articles"}],"industries-application":[{"value":2162,"label":"Air Monitoring"},{"value":2163,"label":"Drinking Water &amp; Wastewater Testing"},{"value":2161,"label":"Environmental"},{"value":2167,"label":"Food &amp; Beverage"},{"value":2169,"label":"Food Safety"},{"value":2668,"label":"Semivolatiles"},{"value":2166,"label":"Soil, Sediment, Biota Testing"}],"resource-type":[{"value":2623,"label":"Featured Application"}],"product-library":[{"value":2391,"label":"LC Columns"},{"value":2373,"label":"Liquid Chromatography Products"}],"resource-technique":[{"value":2299,"label":"Liquid Chromatography"}]},"featured_image_src_large":false,"author_info":{"display_name":"Restek Corporation","author_link":"https:\/\/discover.restek.com\/zh-hans\/author\/restek-corporation\/"},"comment_info":0,"category_info":[{"term_id":7,"name":"Articles","slug":"articles","term_group":0,"term_taxonomy_id":7,"taxonomy":"category","description":"","parent":0,"count":510,"filter":"raw","cat_ID":7,"category_count":510,"category_description":"","cat_name":"Articles","category_nicename":"articles","category_parent":0}],"tag_info":false,"authors":[{"term_id":414,"user_id":11,"is_guest":0,"slug":"restek-corporation","display_name":"Restek Corporation","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/Restek_Favicon_300x300.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/Restek_Favicon_300x300.jpg"},"author_category":"1","first_name":"Restek","last_name":"Corporation","user_url":"https:\/\/www.linkedin.com\/company\/restek-corporation\/posts\/?feedView=all","job_title":"Chromatography Products and Solutions","description":"Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions."}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts\/43518","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/users\/11"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/comments?post=43518"}],"version-history":[{"count":6,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts\/43518\/revisions"}],"predecessor-version":[{"id":72015,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts\/43518\/revisions\/72015"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/media?parent=43518"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/categories?post=43518"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/tags?post=43518"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/industries-application?post=43518"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/post-badge?post=43518"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/resource-type?post=43518"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/product-library?post=43518"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/resource-technique?post=43518"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/ppma_author?post=43518"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}