{"id":43537,"date":"2020-10-14T14:30:00","date_gmt":"2020-10-14T14:30:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/analysis-of-synthetic-cannabinoids-and-metabolites-adding-new-compounds-to-an-existing-lc-msms-method\/"},"modified":"2025-12-10T21:57:16","modified_gmt":"2025-12-10T21:57:16","slug":"analysis-of-synthetic-cannabinoids-and-metabolites-adding-new-compounds-to-an-existing-lc-msms-method","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/zh-hans\/application-notes\/ffan2614\/analysis-of-synthetic-cannabinoids-and-metabolites-adding-new-compounds-to-an-existing-lc-msms-method","title":{"rendered":"Analysis of Synthetic Cannabinoids and Metabolites: Adding New Compounds to an Existing LC-MS\/MS Method"},"content":{"rendered":"\n<h2 class=\"wp-block-heading\">Abstract<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">The analysis of synthetic cannabinoids and their metabolites can be a difficult and challenging task. Keeping up with the ever-growing list of synthetic cannabinoids that illicit drugmakers produce further complicates the analysis. As shown here, the retention and selectivity of the Raptor Biphenyl column allow new drugs to be added to an existing method, providing labs with an important vehicle for improving efficiency and productivity.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Introduction<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">The analysis of synthetic cannabinoids and their metabolites has become a routine procedure in many forensic toxicology laboratories as new drugs appear on the market. When developing methods for these compounds, optimization of analysis time, resolution between metabolites, method robustness, and the ability to add emerging compounds are of ultimate importance as they influence method effectiveness and longevity. Because the Raptor Biphenyl column combines the speed of superficially porous particles (SPP) with the resolution of highly selective USLC technology, we chose it when developing a simple dilute-and-shoot method that included both synthetic cannabinoids and their metabolites in urine. Our original work produced a fast method that separated 29 target compounds in less than seven minutes.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">When analyzing synthetic cannabinoids, chromatographic resolution of some target compounds that cannot be determined by MS alone is essential. For example, synthetic cannabinoids JWH-018 and JWH-073 and their metabolites must be separated chromatographically due to the presence of multiple positional isomers among the monohydroxylated metabolites. These isomers form because each parent compound has multiple sites available for hydroxylation (Figure 1). The more target compounds there are in a method, the more difficult it can be to obtain adequate resolution where needed. Here, we expand on our original method and show proof of concept for how new drugs can be added, while maintaining complete resolution of isobars and separation from matrix interferences in diluted human urine. In the ever-changing landscape of illegal drugs, the ability to add emerging drugs to existing methods allows for much more effective use of laboratory resources as analysts and instruments can focus on routine sample analysis rather than new method development.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 1:<\/strong>\u00a0Analysis of synthetic cannabinoids is complicated by the multiple hydroxylation sites for JWH-018.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n\n<figure class=\"wp-block-image\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FFAN2614-01.jpg\" alt=\"a diagram of a question mark\" title=\"-\"><\/figure>\n\n<\/div><\/div><\/div>\n\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h2 class=\"wp-block-heading\">Experimental<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Samples were prepared in human urine and diluted 3x in a 0.2 \u03bcm PVDF Thomson SINGLE StEP filter vial with water:methanol (50:50) prior to analysis. A Raptor Biphenyl column was used as the analytical column as established in the original method. Evaluations were performed on a Waters ACQUITY UPLC I-Class equipped with a Xevo TQ-S and a Shimadzu Nexera UHPLC equipped with a SCIEX API 4500 MS\/MS. Both systems utilized electrospray ionization in positive ion mode using scheduled multiple reaction monitoring (MRM). Instrument conditions were as follows and analyte transitions for the original target analytes and additional emerging drugs of interest are provided in the figures.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td>Analytical column:<\/td><td>Raptor Biphenyl (2.7 \u00b5m, 50 mm x 3.0 mm;\u00a0cat.# 9309A5E)<\/td><\/tr><tr><td>Guard column:<\/td><td>Raptor Biphenyl EXP guard column (2.7 \u00b5m, 5 mm x 3.0 mm;\u00a0cat.# 9309A0253)<\/td><\/tr><tr><td>Mobile phase A:<\/td><td>0.1% Formic acid in water<\/td><\/tr><tr><td>Mobile phase B:<\/td><td>0.1% Formic acid in acetonitrile<\/td><\/tr><tr><td>Gradient<\/td><td>Time (min) &nbsp;&nbsp;&nbsp;&nbsp;%B<\/td><\/tr><tr><td>&nbsp;<\/td><td>0.00&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;25<\/td><\/tr><tr><td>&nbsp;<\/td><td>1.00&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;25<\/td><\/tr><tr><td>&nbsp;<\/td><td>5.00&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;95<\/td><\/tr><tr><td>&nbsp;<\/td><td>5.50&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;95<\/td><\/tr><tr><td>&nbsp;<\/td><td>5.51&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;25<\/td><\/tr><tr><td>&nbsp;<\/td><td>7.00&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;25<\/td><\/tr><tr><td>Flow rate:<\/td><td>0.6 mL\/min<\/td><\/tr><tr><td>Injection volume:<\/td><td>2 \u00b5L<\/td><\/tr><tr><td>Column temp.:<\/td><td>30 \u00b0C<\/td><\/tr><tr><td>Ion mode:<\/td><td>Positive ESI<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h2 class=\"wp-block-heading\">Results and Discussion<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Today, laboratories are faced with the difficult task of keeping up with the ever-growing list of synthetic cannabinoids that illicit drugmakers produce to avoid legal classification and detection. Previously, we developed a method for the comprehensive screening of 17 synthetic cannabinoids, 12 metabolites, and five internal standards prepared at 5 ng\/mL in human urine and diluted. Complete resolution of isobars and good separation from major matrix interferences were achieved with a cycle time of 5 minutes and a total analysis time of 7 minutes due to the unique selectivity and retention of the Raptor Biphenyl column (Figure 2).<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 2:<\/strong>\u00a0Original Method Developed for the Combined Analysis of Synthetic Cannabinoids and Metabolites in Diluted Human Urine<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/lc_cf0613.png\" alt=\"Combined Analysis of Synthetic Cannabinoids &amp; Metabolites in Urine by LC-MS\/MS on Raptor Biphenyl\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> LC_CF0613<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12 peak-50\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Precursor Ion<\/th><th style=\"text-align: center;width: 75px\">Quantifier Product Ion <\/th><th style=\"text-align: center;width: 75px\">Qualifier Product Ion<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\">Pravadoline<\/td><td class=\"oth\">2.15<\/td><td class=\"oth\">379.29<\/td><td class=\"oth\">135.04<\/td><td class=\"oth\">114.16<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\">AM2233<\/td><td class=\"oth\">2.44<\/td><td class=\"oth\">459.25<\/td><td class=\"oth\">112.2<\/td><td class=\"oth\">98.15<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\">JWH-200-d5<\/td><td class=\"oth\">2.47<\/td><td class=\"oth\">390.34<\/td><td class=\"oth\">155.07<\/td><td class=\"oth\">NA<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for JWH-200\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/103610-04-4\/JWH-200\" target=\"_blank\" rel=\"noopener\">JWH-200<\/a><\/td><td class=\"oth\">2.48<\/td><td class=\"oth\">385.28<\/td><td class=\"oth\">155.07<\/td><td class=\"oth\">114.16<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\">WIN 55, 212<\/td><td class=\"oth\">3.34<\/td><td class=\"oth\">427.29<\/td><td class=\"oth\">155.07<\/td><td class=\"oth\">127.14<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\">JWH-073 N-butanoic acid<\/td><td class=\"oth\">3.39<\/td><td class=\"oth\">358.27<\/td><td class=\"oth\">155.08<\/td><td class=\"oth\">127.11<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\">JWH-073 4-hydroxybutyl<\/td><td class=\"oth\">3.4<\/td><td class=\"oth\">344.24<\/td><td class=\"oth\">155.09<\/td><td class=\"oth\">127.09<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\">JWH-018 N-pentanoic acid<\/td><td class=\"oth\">3.49<\/td><td class=\"oth\">372.18<\/td><td class=\"oth\">155.08<\/td><td class=\"oth\">127.14<\/td><\/tr>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\">JWH-018 5-hydroxypentyl-d5<\/td><td class=\"oth\">3.54<\/td><td class=\"oth\">363.5<\/td><td class=\"oth\">155.08<\/td><td class=\"oth\">NA<\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\">JWH-018 5-hydroxypentyl<\/td><td class=\"oth\">3.55<\/td><td class=\"oth\">358.27<\/td><td class=\"oth\">155.08<\/td><td class=\"oth\">127.11<\/td><\/tr>\n<tr><td class=\"num\">11.<\/td><td class=\"cmpd\">JWH-073 6-hydroxyindole<\/td><td class=\"oth\">3.77<\/td><td class=\"oth\">344.24<\/td><td class=\"oth\">155.09<\/td><td class=\"oth\">127.09<\/td><\/tr>\n<tr><td class=\"num\">12.<\/td><td class=\"cmpd\">JWH-073 5-hydroxyindole-d7<\/td><td class=\"oth\">3.81<\/td><td class=\"oth\">351.21<\/td><td class=\"oth\">155.07<\/td><td class=\"oth\">NA<\/td><\/tr>\n<tr><td class=\"num\">13.<\/td><td class=\"cmpd\">JWH-073 5-hydroxyindole<\/td><td class=\"oth\">3.83<\/td><td class=\"oth\">344.24<\/td><td class=\"oth\">155.09<\/td><td class=\"oth\">127.09<\/td><\/tr>\n<tr><td class=\"num\">14.<\/td><td class=\"cmpd\">JWH-073 7-hydroxyindole<\/td><td class=\"oth\">3.92<\/td><td class=\"oth\">344.24<\/td><td class=\"oth\">155.09<\/td><td class=\"oth\">127.09<\/td><\/tr>\n<tr><td class=\"num\">15.<\/td><td class=\"cmpd\">JWH-018 6-hydroxyindole<\/td><td class=\"oth\">3.94<\/td><td class=\"oth\">358.27<\/td><td class=\"oth\">155.08<\/td><td class=\"oth\">127.11<\/td><\/tr>\n<tr><td class=\"num\">16.<\/td><td class=\"cmpd\">JWH-018 5-hydroxyindole<\/td><td class=\"oth\">3.99<\/td><td class=\"oth\">358.27<\/td><td class=\"oth\">155.08<\/td><td class=\"oth\">127.11<\/td><\/tr>\n<tr><td class=\"num\">17.<\/td><td class=\"cmpd\">JWH-018 7-hydroxyindole<\/td><td class=\"oth\">4.08<\/td><td class=\"oth\">358.27<\/td><td class=\"oth\">155.08<\/td><td class=\"oth\">127.11<\/td><\/tr>\n<\/tbody><\/table>\n<table class=\"peaks col-lg-6 col-12 peak-50\"><thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Precursor Ion<\/th><th style=\"text-align: center;width: 75px\">Quantifier Product Ion <\/th><th style=\"text-align: center;width: 75px\">Qualifier Product Ion<\/th><\/tr><\/thead><tbody>\n<tr><td class=\"num\">18.<\/td><td class=\"cmpd\">RCS-4<\/td><td class=\"oth\">4.15<\/td><td class=\"oth\">322.27<\/td><td class=\"oth\">135.12<\/td><td class=\"oth\">77.09<\/td><\/tr>\n<tr><td class=\"num\">19.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for XLR-11\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1364933-54-9\/XLR-11\" target=\"_blank\" rel=\"noopener\">XLR-11<\/a><\/td><td class=\"oth\">4.21<\/td><td class=\"oth\">330.25<\/td><td class=\"oth\">232.17<\/td><td class=\"oth\">125.1<\/td><\/tr>\n<tr><td class=\"num\">20.<\/td><td class=\"cmpd\">JWH-015-d7<\/td><td class=\"oth\">4.27<\/td><td class=\"oth\">335.28<\/td><td class=\"oth\">155.07<\/td><td class=\"oth\">NA<\/td><\/tr>\n<tr><td class=\"num\">21.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for JWH-250\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/864445-43-2\/JWH-250\" target=\"_blank\" rel=\"noopener\">JWH-250<\/a><\/td><td class=\"oth\">4.27<\/td><td class=\"oth\">336.28<\/td><td class=\"oth\">121.12<\/td><td class=\"oth\">91.07<\/td><\/tr>\n<tr><td class=\"num\">22.<\/td><td class=\"cmpd\">JWH-015<\/td><td class=\"oth\">4.29<\/td><td class=\"oth\">328.26<\/td><td class=\"oth\">155.07<\/td><td class=\"oth\">127.13<\/td><\/tr>\n<tr><td class=\"num\">23.<\/td><td class=\"cmpd\">AM2201<\/td><td class=\"oth\">4.30<\/td><td class=\"oth\">360.26<\/td><td class=\"oth\">155.07<\/td><td class=\"oth\">127.14<\/td><\/tr>\n<tr><td class=\"num\">24.<\/td><td class=\"cmpd\">JWH-203<\/td><td class=\"oth\">4.39<\/td><td class=\"oth\">340.23<\/td><td class=\"oth\">188.18<\/td><td class=\"oth\">125.09<\/td><\/tr>\n<tr><td class=\"num\">25.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for JWH-073\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/208897-48-8\/JWH-073\" target=\"_blank\" rel=\"noopener\">JWH-073<\/a><\/td><td class=\"oth\">4.42<\/td><td class=\"oth\">328.26<\/td><td class=\"oth\">155.07<\/td><td class=\"oth\">127.13<\/td><\/tr>\n<tr><td class=\"num\">26.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for UR-144\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1199943-44-6\/UR-144\" target=\"_blank\" rel=\"noopener\">UR-144<\/a><\/td><td class=\"oth\">4.44<\/td><td class=\"oth\">312.32<\/td><td class=\"oth\">214.17<\/td><td class=\"oth\">125.1<\/td><\/tr>\n<tr><td class=\"num\">27.<\/td><td class=\"cmpd\">JWH-073 4-hydroxyindole<\/td><td class=\"oth\">4.53<\/td><td class=\"oth\">344.24<\/td><td class=\"oth\">155.09<\/td><td class=\"oth\">127.09<\/td><\/tr>\n<tr><td class=\"num\">28.<\/td><td class=\"cmpd\">JWH-018-d9<\/td><td class=\"oth\">4.55<\/td><td class=\"oth\">351.34<\/td><td class=\"oth\">155.07<\/td><td class=\"oth\">NA<\/td><\/tr>\n<tr><td class=\"num\">29.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for JWH-018\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/209414-07-3\/JWH-018\" target=\"_blank\" rel=\"noopener\">JWH-018<\/a><\/td><td class=\"oth\">4.57<\/td><td class=\"oth\">342.27<\/td><td class=\"oth\">155.08<\/td><td class=\"oth\">127.11<\/td><\/tr>\n<tr><td class=\"num\">30.<\/td><td class=\"cmpd\">JWH-081<\/td><td class=\"oth\">4.64<\/td><td class=\"oth\">372.28<\/td><td class=\"oth\">185.12<\/td><td class=\"oth\">157.09<\/td><\/tr>\n<tr><td class=\"num\">31.<\/td><td class=\"cmpd\">JWH-018 4-hydroxyindole<\/td><td class=\"oth\">4.66<\/td><td class=\"oth\">358.27<\/td><td class=\"oth\">155.08<\/td><td class=\"oth\">127.11<\/td><\/tr>\n<tr><td class=\"num\">32.<\/td><td class=\"cmpd\">JWH-122<\/td><td class=\"oth\">4.69<\/td><td class=\"oth\">356.29<\/td><td class=\"oth\">169.12<\/td><td class=\"oth\">141.11<\/td><\/tr>\n<tr><td class=\"num\">33.<\/td><td class=\"cmpd\">JWH-019<\/td><td class=\"oth\">4.70<\/td><td class=\"oth\">356.29<\/td><td class=\"oth\">155.07<\/td><td class=\"oth\">127.1<\/td><\/tr>\n<tr><td class=\"num\">34.<\/td><td class=\"cmpd\">JWH-210<\/td><td class=\"oth\">4.84<\/td><td class=\"oth\">370.31<\/td><td class=\"oth\">183.12<\/td><td class=\"oth\">153.26<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Raptor Biphenyl  (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/9309A5E\" rel=\"noopener\">cat.# 9309A5E<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Dimensions:<\/th><td>50 mm x 3.0 mm ID<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Particle Size:<\/th><td>2.7 \u00b5m<\/td><\/tr><tr><td><\/td><tr><th class=\"sub conditions_header\" scope=\"row\">Guard Column:<\/th><td>Raptor Biphenyl EXP guard column cartridge 5.0 mm, 3.0 mm ID, 2.7 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/9309A0253\" rel=\"noopener\">cat.# 9309A0253<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Temp.:<\/th><td>30 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>5 ng\/mL standard was prepared in urine and diluted 3x with 50:50 methanol:water<\/td><\/tr><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>2 \u00b5L <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Mobile Phase<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">A:<\/th><td>0.1% Formic acid in water <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">B:<\/th><td>0.1% Formic acid in acetonitrile <\/td><\/tr><tr><td><\/td><td><table class=\"cgram_ramp\"><thead><tr><th>Time (min)<\/th><th>Flow (mL\/min)<\/th><th>%A<\/th><th>%B<\/th><\/tr><\/thead><tbody><tr><td>0.00<\/td><td>0.6<\/td><td>75<\/td><td>25<\/td><\/tr><tr><td>1.00<\/td><td>0.6<\/td><td>75<\/td><td>25<\/td><\/tr><tr><td>5.00<\/td><td>0.6<\/td><td>5<\/td><td>95<\/td><\/tr><tr><td>5.50<\/td><td>0.6<\/td><td>5<\/td><td>95<\/td><\/tr><tr><td>5.51<\/td><td>0.6<\/td><td>75<\/td><td>25<\/td><\/tr><tr><td>7.00<\/td><td>0.6<\/td><td>75<\/td><td>25<\/td><\/tr><\/tbody><\/table><\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>MS\/MS<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ion Mode:<\/th><td>ESI+ <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Mode:<\/th><td>MRM <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>UHPLC<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>9309A0253<\/li><li>9309A5E<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/lc_cf0613.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\"><\/p>\n\n\n\n<p class=\"wp-block-paragraph\">To determine whether new drugs could be added to the original method, five supplementary synthetic cannabinoids and salvinorin A (a psychotropic terpenoid) were analyzed under conditions identical to the established method. These emerging drugs were prepared at 50 ng\/mL in human urine and diluted. As shown in Figure 3, the Raptor Biphenyl column provided excellent retention, which allowed separation of the target compounds from the early-eluting matrix interferences. Additionally, all six emerging drugs eluted within the gradient and were well resolved from each other. An overlay of the chromatograms for both the original and emerging drug lists demonstrates that the new drug compounds could easily be added to the existing method without the need for adjustments to the mobile phase, gradient, or analytical column (Figure 4).<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 3:<\/strong>\u00a0Analysis of Emerging Drugs in Diluted Human Urine<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/lc_cf0668.png\" alt=\"Emerging Drugs Poster Figure 3: Analysis of Emerging Drugs in Diluted Human Urine\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> LC_CF0668<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Precursor Ion<\/th><th style=\"text-align: center;width: 75px\">Product Ion<\/th><th style=\"text-align: center;width: 75px\">Product Ion<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for AB-FUBINACA\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1185282-01-2\/AB-FUBINACA\" target=\"_blank\" rel=\"noopener\">AB-FUBINACA<\/a><\/td><td class=\"oth\">3.15<\/td><td class=\"oth\">369.0<\/td><td class=\"oth\">253.0<\/td><td class=\"oth\">109.1<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for AB-PINACA\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1445752-09-9\/AB-PINACA\" target=\"_blank\" rel=\"noopener\">AB-PINACA<\/a><\/td><td class=\"oth\">3.23<\/td><td class=\"oth\">331.2<\/td><td class=\"oth\">215.0<\/td><td class=\"oth\">286.1<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\">Salvinorin A<\/td><td class=\"oth\">3.39<\/td><td class=\"oth\">433.1<\/td><td class=\"oth\">373.0<\/td><td class=\"oth\">91.1<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\">5F-PB-22<\/td><td class=\"oth\">4.04<\/td><td class=\"oth\">377.2<\/td><td class=\"oth\">232.1<\/td><td class=\"oth\">143.9<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\">PB-22<\/td><td class=\"oth\">4.30<\/td><td class=\"oth\">359.2<\/td><td class=\"oth\">214.0<\/td><td class=\"oth\">144.0<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\">APINACA (AKB-48)<\/td><td class=\"oth\">4.76<\/td><td class=\"oth\">366.3<\/td><td class=\"oth\">135.1<\/td><td class=\"oth\">93.1<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Raptor Biphenyl  (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/9309A5E\" rel=\"noopener\">cat.# 9309A5E<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Dimensions:<\/th><td>50 mm x 3.0 mm ID<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Particle Size:<\/th><td>2.7 \u00b5m<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Pore Size:<\/th><td>90 \u00c5<\/td><\/tr><tr><td><\/td><tr><th class=\"sub conditions_header\" scope=\"row\">Guard Column:<\/th><td>Raptor Biphenyl EXP guard column cartridge 5 mm, 3 mm ID, 2.7 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/9309A0253\" rel=\"noopener\">cat.# 9309A0253<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Temp.:<\/th><td>30 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>  50 ng\/mL in human urine and diluted 3x in a 0.2 \u03bcm PVDF Thomson SINGLE StEP filter vial with 50:50 water:methanol<\/td><\/tr><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>2 \u00b5L <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Mobile Phase<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">A:<\/th><td>0.1% Formic acid in water <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">B:<\/th><td>0.1% Formic acid in acetonitrile <\/td><\/tr><tr><td><\/td><td><table class=\"cgram_ramp\"><thead><tr><th>Time (min)<\/th><th>Flow (mL\/min)<\/th><th>%A<\/th><th>%B<\/th><\/tr><\/thead><tbody><tr><td>0.00<\/td><td>0.6<\/td><td>75<\/td><td>25<\/td><\/tr><tr><td>1.00<\/td><td>0.6<\/td><td>75<\/td><td>25<\/td><\/tr><tr><td>5.00<\/td><td>0.6<\/td><td>5<\/td><td>95<\/td><\/tr><tr><td>5.50<\/td><td>0.6<\/td><td>5<\/td><td>95<\/td><\/tr><tr><td>5.51<\/td><td>0.6<\/td><td>75<\/td><td>25<\/td><\/tr><tr><td>7.00<\/td><td>0.6<\/td><td>75<\/td><td>25<\/td><\/tr><\/tbody><\/table><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Max Pressure:<\/th><td>400 bar<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>MS\/MS<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ion Mode:<\/th><td>ESI+ <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Mode:<\/th><td>Scheduled MRM <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>UHPLC<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>9309A0253<\/li><li>9309A5E<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>Drugs<\/li><li>Emerging<\/li><li>cannabinoids<\/li><li>emerging drugs<\/li><li>metabolites<\/li><li>psychoactive<\/li><li>psychoactive drugs<\/li><li>spice<\/li><li>synthetic<\/li><li>synthetic cannabinoids<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/lc_cf0668.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 4:<\/strong>\u00a0Overlay of Emerging Drugs Chromatogram and Synthetic Cannabinoids and Metabolites Chromatogram in Diluted Human Urine<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/lc_cf0669.png\" alt=\"Emerging Drugs Poster Figure 4: Overlay of Emerging Drugs Chromatogram and Synthetic Cannabinoids and Metabolites Chromatogram\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> LC_CF0669<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Precursor Ion<\/th><th style=\"text-align: center;width: 75px\">Product Ion<\/th><th style=\"text-align: center;width: 75px\">Product Ion<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for AB-FUBINACA\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1185282-01-2\/AB-FUBINACA\" target=\"_blank\" rel=\"noopener\">AB-FUBINACA<\/a><\/td><td class=\"oth\">3.15<\/td><td class=\"oth\">369.0<\/td><td class=\"oth\">253.0<\/td><td class=\"oth\">109.1<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for AB-PINACA\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1445752-09-9\/AB-PINACA\" target=\"_blank\" rel=\"noopener\">AB-PINACA<\/a><\/td><td class=\"oth\">3.23<\/td><td class=\"oth\">331.2<\/td><td class=\"oth\">215.0<\/td><td class=\"oth\">286.1<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\">Salvinorin A<\/td><td class=\"oth\">3.39<\/td><td class=\"oth\">433.1<\/td><td class=\"oth\">373.0<\/td><td class=\"oth\">91.1<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\">5F-PB-22<\/td><td class=\"oth\">4.04<\/td><td class=\"oth\">377.2<\/td><td class=\"oth\">232.1<\/td><td class=\"oth\">143.9<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\">PB-22<\/td><td class=\"oth\">4.30<\/td><td class=\"oth\">359.2<\/td><td class=\"oth\">214.0<\/td><td class=\"oth\">144.0<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\">APINACA (AKB-48)<\/td><td class=\"oth\">4.76<\/td><td class=\"oth\">366.3<\/td><td class=\"oth\">135.1<\/td><td class=\"oth\">93.1<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Raptor Biphenyl  (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/9309A5E\" rel=\"noopener\">cat.# 9309A5E<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Dimensions:<\/th><td>50 mm x 3.0 mm ID<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Particle Size:<\/th><td>2.7 \u00b5m<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Pore Size:<\/th><td>90 \u00c5<\/td><\/tr><tr><td><\/td><tr><th class=\"sub conditions_header\" scope=\"row\">Guard Column:<\/th><td>Raptor Biphenyl EXP guard column cartridge 5 mm, 3 mm ID, 2.7 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/9309A0253\" rel=\"noopener\">cat.# 9309A0253<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Temp.:<\/th><td>30 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td><\/td><\/tr><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>2 \u00b5L <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Mobile Phase<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">A:<\/th><td>0.1% Formic acid in water <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">B:<\/th><td>0.1% Formic acid in acetonitrile <\/td><\/tr><tr><td><\/td><td><table class=\"cgram_ramp\"><thead><tr><th>Time (min)<\/th><th>Flow (mL\/min)<\/th><th>%A<\/th><th>%B<\/th><\/tr><\/thead><tbody><tr><td>0.00<\/td><td>0.6<\/td><td>75<\/td><td>25<\/td><\/tr><tr><td>1.00<\/td><td>0.6<\/td><td>75<\/td><td>25<\/td><\/tr><tr><td>5.00<\/td><td>0.6<\/td><td>5<\/td><td>95<\/td><\/tr><tr><td>5.50<\/td><td>0.6<\/td><td>5<\/td><td>95<\/td><\/tr><tr><td>5.51<\/td><td>0.6<\/td><td>75<\/td><td>25<\/td><\/tr><tr><td>7.00<\/td><td>0.6<\/td><td>75<\/td><td>25<\/td><\/tr><\/tbody><\/table><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Max Pressure:<\/th><td>400 bar<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>MS\/MS<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ion Mode:<\/th><td>ESI+ <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Mode:<\/th><td>Scheduled MRM <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>UHPLC<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>9309A0253<\/li><li>9309A5E<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>Drugs<\/li><li>Emerging<\/li><li>cannabinoids<\/li><li>emerging drugs<\/li><li>metabolites<\/li><li>psychoactive<\/li><li>psychoactive drugs<\/li><li>spice<\/li><li>synthetic<\/li><li>synthetic cannabinoids<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/lc_cf0669.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\"><\/p>\n\n\n\n<p class=\"wp-block-paragraph\">The methodology used here allows for the simultaneous analysis of emerging drugs and key existing synthetic cannabinoids, including JWH-018 and JWH-073 and their metabolites. Notably, these results were achieved with faster analysis times than were possible using a 5 \u00b5m fully porous particle (FPP) column (Figures 5 and 6). The Raptor Biphenyl column contains superficially porous particles (SPP) that significantly reduce analysis time. Using a Raptor Biphenyl SPP column provided better resolution of more compounds (including isomers and emerging compounds) in less time than traditional FPP columns, which allows more samples to be analyzed per shift.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 5:<\/strong>\u00a0Analysis of Synthetic Cannabinoids on an Ultra Biphenyl FPP Column<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/lc_cf0526.png\" alt=\"Synthetic Cannabinoids on Ultra Biphenyl\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> LC_CF0526<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">MRM1<\/th><th style=\"text-align: center;width: 75px\">MRM2<\/th><th style=\"text-align: center;width: 75px\">MRM3<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\">WIN 48098<\/td><td class=\"oth\">3.85<\/td><td class=\"oth\">379.2\/135.2<\/td><td class=\"oth\">379.2\/114.3<\/td><td class=\"oth\">379.2\/107.3<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for JWH-200\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/103610-04-4\/JWH-200\" target=\"_blank\" rel=\"noopener\">JWH-200<\/a><\/td><td class=\"oth\">4.34<\/td><td class=\"oth\">385.3\/114.0<\/td><td class=\"oth\">385.3\/127.0<\/td><td class=\"oth\">385.3\/155.0<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\">WIN 55212-2<\/td><td class=\"oth\">5.81<\/td><td class=\"oth\">427.2\/155.1<\/td><td class=\"oth\">427.2\/127.1<\/td><td class=\"oth\">427.2\/100.1<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\">AM-694<\/td><td class=\"oth\">7.01<\/td><td class=\"oth\">436.1\/309.1<\/td><td class=\"oth\">436.1\/231.2<\/td><td class=\"oth\">&#8212;<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\">JWH-015<\/td><td class=\"oth\">7.23<\/td><td class=\"oth\">328.3\/155.1<\/td><td class=\"oth\">328.3\/200.1<\/td><td class=\"oth\">328.3\/200.1<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for JWH-250\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/864445-43-2\/JWH-250\" target=\"_blank\" rel=\"noopener\">JWH-250<\/a><\/td><td class=\"oth\">7.25<\/td><td class=\"oth\">336.3\/121.1<\/td><td class=\"oth\">336.3\/91.0<\/td><td class=\"oth\">336.3\/144.0<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for JWH-073\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/208897-48-8\/JWH-073\" target=\"_blank\" rel=\"noopener\">JWH-073<\/a><\/td><td class=\"oth\">7.48<\/td><td class=\"oth\">328.2\/127.1<\/td><td class=\"oth\">328.2\/155.2<\/td><td class=\"oth\">328.2\/199.9<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for JWH-018\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/209414-07-3\/JWH-018\" target=\"_blank\" rel=\"noopener\">JWH-018<\/a><\/td><td class=\"oth\">7.75<\/td><td class=\"oth\">342.3\/127.1<\/td><td class=\"oth\">342.3\/155.1<\/td><td class=\"oth\">342.3\/144.9<\/td><\/tr>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\">JWH-081<\/td><td class=\"oth\">7.88<\/td><td class=\"oth\">372.2\/185.2<\/td><td class=\"oth\">372.2\/157.1<\/td><td class=\"oth\">372.2\/144.1<\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\">JWH-019<\/td><td class=\"oth\">8.01<\/td><td class=\"oth\">356.3\/127.2<\/td><td class=\"oth\">356.3\/155.0<\/td><td class=\"oth\">356.3\/144.2<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Ultra Biphenyl  (<a target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/9109552?utm_source=chromatograms&amp;utm_medium=link&amp;utm_campaign=LC_CF0526\" rel=\"noopener\">cat.# 9109552<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Dimensions:<\/th><td>50 mm x 2.1 mm ID<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Particle Size:<\/th><td>5 \u00b5m<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Pore Size:<\/th><td>100 \u00c5<\/td><\/tr><tr><td><\/td><tr><th class=\"sub conditions_header\" scope=\"row\">Temp.:<\/th><td>40 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>Methanol<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>50 ng\/mL<\/td><\/tr><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>5 \u00b5L <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Mobile Phase<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">A:<\/th><td>Water + 0.1% formic acid <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">B:<\/th><td>Acetonitrile + 0.1% formic acid <\/td><\/tr><tr><td><\/td><td><table class=\"cgram_ramp\"><thead><tr><th>Time (min)<\/th><th>Flow (mL\/min)<\/th><th>%A<\/th><th>%B<\/th><\/tr><\/thead><tbody><tr><td>0.00<\/td><td>0.5<\/td><td>95<\/td><td>5<\/td><\/tr><tr><td>10<\/td><td>0.5<\/td><td>5<\/td><td>95<\/td><\/tr><tr><td>10.1<\/td><td>0.5<\/td><td>95<\/td><td>5<\/td><\/tr><tr><td>12<\/td><td>stop<\/td><td><\/td><td><\/td><\/tr><\/tbody><\/table><\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>AB SCIEX API 4000 MS\/MS<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Model #:<\/th><td>API 4000 <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ion Source:<\/th><td>TurboIonSpray<sup>&reg;<\/sup> <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ion Mode:<\/th><td>ESI+ <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ion Spray Voltage:<\/th><td>3000 kV<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Curtain Gas:<\/th><td>40  psi (275.8 kPa)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Gas 1:<\/th><td>40  psi (275.8 kPa)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Gas 2:<\/th><td>40  psi (275.8 kPa)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Source Temp.:<\/th><td>600 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Mode:<\/th><td>MRM <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Dwell Time:<\/th><td>10 ms<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Applied Biosystems\/MDS Sciex LC-MS\/MS System<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Acknowledgement<\/th><td>Special thanks to Paul Kennedy and Cayman Chemical for standards.<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/lc_cf0526.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 6:<\/strong>\u00a0Analysis of Synthetic Cannabinoid Metabolites on an Ultra Biphenyl FPP Column<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/lc_cf0531.png\" alt=\"Synthetic Cannabinoid Metabolites on Ultra Biphenyl (50 ng\/mL Extracted Urine Standard)\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> LC_CF0531<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12 peak-50\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">MRM1<\/th><th style=\"text-align: center;width: 75px\">MRM2<\/th><th style=\"text-align: center;width: 75px\">MRM3<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\">JWH-073 4-hydroxybutyl<\/td><td class=\"oth\">2.05<\/td><td class=\"oth\">344.1\/155.1<\/td><td class=\"oth\">344.1\/127.2<\/td><td class=\"oth\">344.1\/144.0<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\">JWH-073 N-butanoic acid<\/td><td class=\"oth\">2.13<\/td><td class=\"oth\">358.1\/155.1<\/td><td class=\"oth\">358.1\/127.3<\/td><td class=\"oth\">358.1\/144.1<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\">JWH-018 N-pentanoic acid-d4 (IS)<\/td><td class=\"oth\">2.54<\/td><td class=\"oth\">376.1\/155.2<\/td><td class=\"oth\">376.1\/230.4<\/td><td class=\"oth\">376.1\/248.3<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\">JWH-018 N-pentanoic acid<\/td><td class=\"oth\">2.58<\/td><td class=\"oth\">372.1\/155.2<\/td><td class=\"oth\">372.1\/127.1<\/td><td class=\"oth\">372.1\/144.1<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\">JWH-018 5-hydroxypentyl<\/td><td class=\"oth\">2.58<\/td><td class=\"oth\">358.1\/155.2<\/td><td class=\"oth\">358.1\/127.1<\/td><td class=\"oth\">358.1\/230.3<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\">JWH-073 6-hydroxyindole<\/td><td class=\"oth\">3.52<\/td><td class=\"oth\">344.1\/155.2<\/td><td class=\"oth\">344.1\/127.1<\/td><td class=\"oth\">344.1\/145.1<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\">JWH-073 5-hydroxyindole<\/td><td class=\"oth\">3.68<\/td><td class=\"oth\">344.1\/155.2<\/td><td class=\"oth\">344.1\/127.1<\/td><td class=\"oth\">344.1\/160.0<\/td><\/tr>\n<\/tbody><\/table>\n<table class=\"peaks col-lg-6 col-12 peak-50\"><thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">MRM1<\/th><th style=\"text-align: center;width: 75px\">MRM2<\/th><th style=\"text-align: center;width: 75px\">MRM3<\/th><\/tr><\/thead><tbody>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\">JWH-073 7-hydroxyindole<\/td><td class=\"oth\">3.95<\/td><td class=\"oth\">344.2\/155.1<\/td><td class=\"oth\">344.2\/127.1<\/td><td class=\"oth\">344.2\/216.3<\/td><\/tr>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\">JWH-018 6-hydroxyindole<\/td><td class=\"oth\">4.00<\/td><td class=\"oth\">358.1\/155.1<\/td><td class=\"oth\">358.1\/127.2<\/td><td class=\"oth\">358.1\/145.2<\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\">JWH-018 5-hydroxyindole<\/td><td class=\"oth\">4.12<\/td><td class=\"oth\">358.1\/155.1<\/td><td class=\"oth\">358.1\/127.2<\/td><td class=\"oth\">358.1\/160.2<\/td><\/tr>\n<tr><td class=\"num\">11.<\/td><td class=\"cmpd\">JWH-018 7-hydroxyindole<\/td><td class=\"oth\">4.34<\/td><td class=\"oth\">358.1\/155.1<\/td><td class=\"oth\">358.1\/127.1<\/td><td class=\"oth\">358.1\/230.3<\/td><\/tr>\n<tr><td class=\"num\">12.<\/td><td class=\"cmpd\">JWH-073 4-hydroxyindole-d7 (IS)<\/td><td class=\"oth\">5.10<\/td><td class=\"oth\">351.3\/127.0<\/td><td class=\"oth\">351.3\/223.0<\/td><td class=\"oth\">351.3\/155.0<\/td><\/tr>\n<tr><td class=\"num\">13.<\/td><td class=\"cmpd\">JWH-073 4-hydroxyindole<\/td><td class=\"oth\">5.14<\/td><td class=\"oth\">344.1\/155.2<\/td><td class=\"oth\">344.1\/127.2<\/td><td class=\"oth\">344.1\/160.0<\/td><\/tr>\n<tr><td class=\"num\">14.<\/td><td class=\"cmpd\">JWH-018 4-hydroxyindole<\/td><td class=\"oth\">5.44<\/td><td class=\"oth\">358.1\/155.2<\/td><td class=\"oth\">358.1\/230.2<\/td><td class=\"oth\">358.1\/127.1<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Ultra Biphenyl  (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/9109552\" rel=\"noopener\">cat.# 9109552<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Dimensions:<\/th><td>50 mm x 2.1 mm ID<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Particle Size:<\/th><td>5 \u00b5m<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Pore Size:<\/th><td>100 \u00c5<\/td><\/tr><tr><td><\/td><tr><th class=\"sub conditions_header\" scope=\"row\">Temp.:<\/th><td>25 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>50:50 Mobile phase<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>50 ng\/mL extracted spiked sample<\/td><\/tr><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>10 \u00b5L <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Mobile Phase<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">A:<\/th><td>Water + 0.05% acetic acid (pH approx. 3.4) <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">B:<\/th><td>Acetonitrile + 0.05% acetic acid <\/td><\/tr><tr><td><\/td><td><table class=\"cgram_ramp\"><thead><tr><th>Time (min)<\/th><th>Flow (mL\/min)<\/th><th>%A<\/th><th>%B<\/th><\/tr><\/thead><tbody><tr><td>0.00<\/td><td>0.5<\/td><td>55<\/td><td>45<\/td><\/tr><tr><td>2.00<\/td><td>0.5<\/td><td>55<\/td><td>45<\/td><\/tr><tr><td>6.00<\/td><td>0.5<\/td><td>15<\/td><td>85<\/td><\/tr><tr><td>6.10<\/td><td>0.5<\/td><td>5<\/td><td>95<\/td><\/tr><tr><td>7.00<\/td><td>0.5<\/td><td>5<\/td><td>95<\/td><\/tr><tr><td>7.10<\/td><td>0.5<\/td><td>55<\/td><td>45<\/td><\/tr><tr><td>8.50<\/td><td>stop<\/td><td><\/td><td><\/td><\/tr><\/tbody><\/table><\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>API 4000<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Model #:<\/th><td>API 4000 <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ion Source:<\/th><td>TurboIonSpray<sup>&reg;<\/sup> <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ion Mode:<\/th><td>ESI+ <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ion Spray Voltage:<\/th><td>3 kV<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Curtain Gas:<\/th><td>40  psi (275.8 kPa)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Gas 1:<\/th><td>40  psi (275.8 kPa)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Gas 2:<\/th><td>40  psi (275.8 kPa)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Interface Temp.:<\/th><td>600 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Mode:<\/th><td>MRM <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Dwell Time:<\/th><td>30 ms<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>API LC-MS\/MS<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Sample Preparation<\/th><td>1) Spike 1 mL blank urine sample with analytes and internal standards.<br \/><br \/>2) <i>Hydrolyze sample:<\/i><br \/>   &#8211; Add 1 mL solution of beta-glucuronidase from keyhole limpet (Sigma-Aldrich cat.# G8132). Solution is prepared at a concentration of 5000 Fishman units\/mL in 100 mM ammonium acetate buffer (pH = 5.0).<br \/>   &#8211; Incubate at 60 \u00b0C for 3 hours.<br \/><br \/><i>3) Extract sample on 6 mL, 500 mg C18 high-load end-capped Resprep SPE cartridge cat.# 24052<\/i>:<br \/>   &#8211; Add 1 mL 5 mM ammonium acetate + 0.1% acetic acid (pH = 4.2) to sample.<br \/>   &#8211; Condition cartridge with 3x 1 mL acetonitrile.<br \/>   &#8211; Condition cartridge with 3x 1 mL 5 mM ammonium acetate + 0.1% acetic acid.<br \/>   &#8211; Apply sample and allow to pass through under gravity.<br \/>   &#8211; Rinse with 3x 1 mL 5 mM ammonium acetate + 0.1% acetic acid.<br \/>   &#8211; Dry cartridge with vacuum for 10 minutes.<br \/>   &#8211; Elute with 3 mL acetonitrile followed by 3 mL butyl chloride.<br \/><br \/><i>4) Concentrate sample:<\/i><br \/>   &#8211; Evaporate sample to dryness under nitrogen at 40 \u00b0C.<br \/>   &#8211; Reconstitute in 0.5 mL water + 0.05% acetic acid:acetonitrile + 0.05% acetic acid (50:50).<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Notes<\/th><td>Resprep SPE cartridge cat.# 24052 was used to produce this chromatogram, but has since been discontinued. For assistance choosing a replacement for this application, contact Restek Technical Service or your local Restek representative.<br \/>&#8211; &#8211; &#8211; &#8211; &#8211; &#8211;<br \/>Since multiple transitions are shared between analytes, only 5 transitions are shown to simplify viewing. The transitions shown are: 344.1\/155.1 (blue trace), 358.1\/155.1 (green trace), 372.1\/155.2 (grey trace), 376.1\/155.2 (purple trace &#8211; internal standard), 351.3\/127.0 (red trace &#8211; internal standard).<br \/><br \/>CAD Gas was set to 4 psi.<br \/><br \/>For 1 ng\/mL calibration level, see chromatogram LC_CF0530.<br \/>For 500 ng\/mL calibration level, see chromatogram LC_CF0532.<br \/><\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Acknowledgement<\/th><td>Special thanks to Cayman Chemical for reference standards.<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>9109552<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/lc_cf0531.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h2 class=\"wp-block-heading\">Conclusion<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Labs analyzing synthetic cannabinoids and their metabolites are under increasing pressure to add new compounds to their analytical testing services. While this can be done through new method development, adding new compounds to an existing method can save time and resources. The method shown here demonstrates the advantages of the Raptor Biphenyl SPP column for this analysis; due to the column\u2019s highly retentive, selective characteristics, 35 drugs (including isomers and emerging compounds) can be analyzed in just 5 minutes.<\/p>\n\n\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n        <div class=\"cpb\">\n            <h3 class=\"cpb-heading\">\u76f8\u5173\u4ea7\u54c1<\/h3>\n            <hr class=\"cpb-heading-underline\" \/>\n            <div class=\"cpb-list\">\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/9309A5E\">                                \u8d27\u53f7\uff1a 9309A5E                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Raptor Biphenyl, 2.7 \u00b5m, 50 x 3.0 mm HPLC Column<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/9309A5E\">\u67e5\u770b\u4ea7\u54c1<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/9309A0253\">                                \u8d27\u53f7\uff1a 9309A0253                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Raptor Biphenyl EXP Guard Column Cartridge, 2.7 \u00b5m, 5 x 3.0 mm, 3-pk.<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/9309A0253\">\u67e5\u770b\u4ea7\u54c1<\/a>\n                                                    <\/div>\n                    <\/div>\n                            <\/div>\n        <\/div>\n        \n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>The analysis of synthetic cannabinoids and their metabolites can be a difficult and challenging task. Keeping up with the ever-growing list of synthetic cannabinoids that illicit drugmakers produce further complicates the analysis. As shown here, the retention and selectivity of the Raptor Biphenyl column allows new drugs to be added to an existing method, providing labs with an important vehicle for improving efficiency and productivity.<\/p>\n","protected":false},"author":3,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":""},"categories":[2805],"tags":[],"industries-application":[2280,2223,2226,2239,2221],"post-badge":[],"resource-type":[],"product-library":[2499,2481,2483,2510],"resource-technique":[2326,2363,2329],"ppma_author":[606,586],"class_list":["post-43537","post","type-post","status-publish","format-standard","hentry","category-application-notes","industries-application-cannabis-de","industries-application-klinisch","industries-application-klinische-toxikologie","industries-application-forensik","industries-application-medizinisches-nicht-medizinisches-cannabis","product-library-lc-saulen","product-library-flussigchromatographie-produkte","product-library-produkte-zur-probenvorbereitung","product-library-spe-sle-produkte","resource-technique-flussigchromatografie","resource-technique-ms-ms-de","resource-technique-festphasenextraktion-spe"],"acf":[],"taxonomy_info":{"category":[{"value":2805,"label":"Application Notes"}],"industries-application":[{"value":2280,"label":"Cannabis"},{"value":2223,"label":"Klinisch"},{"value":2226,"label":"Klinische Toxikologie"},{"value":2239,"label":"Forensik"},{"value":2221,"label":"Medizinisches oder nicht-medizinisches Cannabis"}],"product-library":[{"value":2499,"label":"LC S\u00e4ulen"},{"value":2481,"label":"Fl\u00fcssigchromatographie Produkte"},{"value":2483,"label":"Produkte zur Probenvorbereitung"},{"value":2510,"label":"SPE &amp; SLE"}],"resource-technique":[{"value":2326,"label":"Fl\u00fcssigchromatografie"},{"value":2363,"label":"MS\/MS"},{"value":2329,"label":"Festphasenextraktion (SPE)"}]},"featured_image_src_large":false,"author_info":{"display_name":"Audrey Crater","author_link":"https:\/\/discover.restek.com\/zh-hans\/author\/audrey-craterrestek-com\/"},"comment_info":0,"category_info":[{"term_id":2805,"name":"Application Notes","slug":"application-notes","term_group":0,"term_taxonomy_id":2805,"taxonomy":"category","description":"","parent":0,"count":79,"filter":"raw","cat_ID":2805,"category_count":79,"category_description":"","cat_name":"Application Notes","category_nicename":"application-notes","category_parent":0}],"tag_info":false,"authors":[{"term_id":606,"user_id":0,"is_guest":1,"slug":"sharon-lupo","display_name":"Sharon Lupo","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-lupo-sharon.png","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-lupo-sharon.png"},"author_category":"1","first_name":"Sharon","last_name":"Lupo","user_url":"","job_title":"","description":""},{"term_id":586,"user_id":0,"is_guest":1,"slug":"frances-carroll","display_name":"Frances Carroll","avatar_url":"https:\/\/secure.gravatar.com\/avatar\/?s=96&d=mm&r=g","author_category":"1","first_name":"Frances","last_name":"Carroll","user_url":"","job_title":"","description":""}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts\/43537","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/users\/3"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/comments?post=43537"}],"version-history":[{"count":3,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts\/43537\/revisions"}],"predecessor-version":[{"id":73275,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts\/43537\/revisions\/73275"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/media?parent=43537"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/categories?post=43537"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/tags?post=43537"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/industries-application?post=43537"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/post-badge?post=43537"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/resource-type?post=43537"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/product-library?post=43537"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/resource-technique?post=43537"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/ppma_author?post=43537"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}