{"id":45254,"date":"2022-09-06T00:00:00","date_gmt":"2022-09-06T00:00:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/analysis-of-furan-and-alkylfurans-in-food-commodities-by-spme-arrow-gc-ms\/"},"modified":"2026-07-16T12:54:00","modified_gmt":"2026-07-16T12:54:00","slug":"analysis-of-furan-and-alkylfurans-in-food-commodities-by-spme-arrow-gc-ms","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/zh-hans\/application-notes\/fsan3916\/analysis-of-furan-and-alkylfurans-in-food-commodities-by-spme-arrow-gc-ms","title":{"rendered":"Analysis of Furan and Alkylfurans in Food Commodities by SPME Arrow GC-MS"},"content":{"rendered":"\n<h2 class=\"wp-block-heading\">Abstract<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">In this study, SPME Arrow sample preparation was tested for the analysis of furan and alkylfurans in baby formula and coffee samples. Optimized conditions were established for GC-MS analysis using a PAL CTC autosampler and an Rxi-624Sil MS column. Results demonstrated the method may be suitable for diverse sample types containing both high and low concentrations of furan and alkylfurans. This approach could allow food safety labs to improve productivity by using SPME Arrow as a single sample preparation approach compared to methods that use headspace alone or SPME for low-concentration samples and headspace for high-concentration samples.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Introduction<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Analysis of furan and alkylfurans is emerging as a growth area in food safety testing due to increasing recognition of their potential health risks. These compounds form through a diverse array of chemical reactions that occur during heating processes, such as roasting or sterilization, so exposure can occur through a wide range of foods. Exposure has not been well characterized, so in early 2022, the European Commission adopted a recommendation calling for member states to monitor furan, 2-methylfuran, and&nbsp;3-methylfuran in particular foods [1]. The agency also advised monitoring non-methylfurans, such as 2,5-dimethylfuran, 2-pentylfuran, and 2-ethylfuran, if methods are available that have been demonstrated to be reliable for this purpose.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Furan and alkylfurans are volatile compounds that can be analyzed by static headspace (HS) and\/or solid phase microextraction (SPME) in combination with GC-MS. Static headspace GC-MS is typically used with samples that contain high levels of furan and alkylfurans, such as coffee, but it may not offer adequate sensitivity for low concentration samples. SPME is more sensitive and better for low-level samples, but traditional SPME fibers are very fragile and not well suited to high-throughput environments. To meet regulatory expectations efficiently, food safety labs need a simple, fast, cost-effective workflow that performs well for a wide range of analyte concentrations in different sample types.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">For this reason, we decided to evaluate the performance of&nbsp;<a href=\"https:\/\/www.restek.com\/c\/1354\" target=\"_blank\" rel=\"noopener\">SPME Arrows<\/a>&nbsp;to determine if a single approach could be used for the analysis of furan and alkylfurans. SPME Arrows are second-generation SPME devices that are mechanically enhanced with a sheath and inner rod to prevent breakage; they also have higher phase volumes to provide more loading capacity (Figure 1). In this study, based on an initial comparison of sample preparation techniques, we developed an optimized SPME Arrow GC-MS method for the analysis of furan and alkylfurans across a range of concentrations in both coffee and baby formula.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 1:<\/strong>\u00a0SPME Arrows have higher loading capacity than traditional fibers and are also sheathed in metal to prevent breakage.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n\n<div class=\"wp-block-columns is-layout-flex wp-container-core-columns-is-layout-8f761849 wp-block-columns-is-layout-flex\">\n<div class=\"wp-block-column is-layout-flow wp-block-column-is-layout-flow\">\n<p class=\"has-text-align-center wp-block-paragraph\"><strong>SPME Arrows<\/strong><\/p>\n\n\n<style>.kb-image45254_8a9638-e8 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<div class=\"wp-block-kadence-image kb-image45254_8a9638-e8\"><figure class=\"aligncenter size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"850\" height=\"554\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAN3916-01a.jpg\" alt=\"\" class=\"kb-img wp-image-20014\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAN3916-01a.jpg 850w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAN3916-01a-300x196.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAN3916-01a-768x501.jpg 768w\" sizes=\"auto, (max-width: 850px) 100vw, 850px\" \/><\/figure><\/div>\n<\/div>\n\n\n\n<div class=\"wp-block-column is-layout-flow wp-block-column-is-layout-flow\">\n<p class=\"has-text-align-center wp-block-paragraph\"><strong>SPME Fibers<\/strong><\/p>\n\n\n<style>.kb-image45254_4007b6-c8 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<div class=\"wp-block-kadence-image kb-image45254_4007b6-c8\"><figure class=\"aligncenter size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"861\" height=\"554\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAN3916-01b.jpg\" alt=\"\" class=\"kb-img wp-image-20020\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAN3916-01b.jpg 861w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAN3916-01b-300x193.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAN3916-01b-768x494.jpg 768w\" sizes=\"auto, (max-width: 861px) 100vw, 861px\" \/><\/figure><\/div>\n<\/div>\n<\/div>\n\n<\/div><\/div><\/div>\n\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h2 class=\"wp-block-heading\">Experimental<\/h2>\n\n\n\n<h3 class=\"wp-block-heading\">Standard Solutions<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">For each target analyte, 1 mg\/mL solutions were prepared in methanol by spiking pure standards using a glass syringe. Then, 100 ppm stock solutions, one containing all the target analytes and one containing all the internal standards, were prepared in methanol. For the calibration curve covering a low concentration range, 0.1 and 0.5 \u00b5g\/mL working solutions were prepared by spiking 10 and 50 \u00b5L of the 100 ppm stock, respectively, in 10 mL of LC-MS grade water. An internal standard solution of 1 \u00b5g\/mL was prepared by adding 100 \u00b5L of the 100 ppm stock solution to 9.9 mL of water. For the analysis of coffee samples, a high-concentration calibration curve was used. For this purpose, aqueous working solutions at 10 and 25 \u00b5g\/mL, for target analytes and internal standards, respectively, were made by spiking the original individual methanolic solutions directly into water. A 0.5 \u00b5g\/mL solution was prepared by diluting 500 \u00b5L of the 10 \u00b5g\/mL solution in water until reaching a volume of 10 mL.&nbsp;&nbsp;&nbsp;<\/p>\n\n\n\n<h3 class=\"wp-block-heading\">Sample Preparation<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Baby formula and instant coffee samples were purchased at a local grocery store. Preliminary experiments comparing headspace, traditional SPME fibers, and SPME Arrows were run to establish a sample preparation procedure that would be suitable for a wide range of analyte concentrations. Parameters tested included coating chemistry, incubation time and temperature, and extraction time [2,3,4].<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Based on the results of the initial sample preparation experiments, SPME Arrow was selected as the best approach and was used under the optimized conditions shown in Table I for subsequent method evaluation experiments. All samples and calibration solutions were prepared in 20 mL glass vials and extracted and analyzed by automated headspace-SPME using a 120 \u00b5m wide carbon range\/PDMS Restek PAL SPME Arrow (cat.# 27487) assembled in a PAL CTC autosampler.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Table I:&nbsp;<\/strong>Sample Preparation Conditions for Baby Formula and Coffee Samples<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td><strong>Sample<\/strong><\/td><td><strong>Baby Formula<\/strong><\/td><td><strong>Coffee<\/strong><\/td><\/tr><tr><td><strong>Amount<\/strong><\/td><td colspan=\"2\">0.5 g<\/td><\/tr><tr><td><strong>Volume of 30% NaCl solution added<\/strong><\/td><td>10 mL<\/td><td>5 mL<\/td><\/tr><tr><td><strong>Volume of internal standard solution added<\/strong><\/td><td>50 \u00b5L(1 \u00b5g\/mL solution)<\/td><td>40 \u00b5L(25 \u00b5g\/mL solution)<\/td><\/tr><tr><td><strong>Incubation time<\/strong><\/td><td colspan=\"2\">10 min<\/td><\/tr><tr><td><strong>Incubation and extraction temperature<\/strong><\/td><td colspan=\"2\">50 \u00b0C<\/td><\/tr><tr><td><strong>Agitation<\/strong><\/td><td colspan=\"2\">250 rpm<\/td><\/tr><tr><td><strong>HS-SPME extraction time<\/strong><\/td><td>10 min<\/td><td>1 min<\/td><\/tr><tr><td><strong>Desorption temperature<\/strong><\/td><td colspan=\"2\">280 \u00b0C<\/td><\/tr><tr><td><strong>Desorption time<\/strong><\/td><td colspan=\"2\">1 min<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<h3 class=\"wp-block-heading\">Calibration Curves<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Two calibration curves were prepared in 30% sodium chloride solution: one for low concentration analysis of furan and alkylfurans in baby formula, and one for the quantitation of high concentrations of these target analytes in coffee. For the low calibration range, vials were filled with 10 mL of sodium chloride solution, whereas for the high calibration range, 5 mL of solution were added. Analytes were spiked by adding different volumes of working solutions to the calibration vials, as shown in Table II. Internal standard solutions were added to each calibration vial (50 \u00b5L of the 1 \u00b5g\/mL solution for the low concentration calibration curve, and 40 \u00b5L of the 25 \u00b5g\/mL solution for the high concentration calibration curve). To account for variation in recoveries due to matrix differences, it was essential to use isotopically labeled analogues for almost every analyte (except for 2- and 3-methylfuran because 2-methylfuran-d6 worked for both compounds). Calibration curves were constructed by plotting analyte area\/internal standard average area ratios (n=2) versus spiked concentration.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Table II:&nbsp;<\/strong>Calibration Solutions<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td rowspan=\"2\">&nbsp;<\/td><td colspan=\"2\"><strong>Low Concentration Calibration Curve<\/strong><\/td><td colspan=\"2\"><strong>High Concentration Calibration Curve<\/strong><\/td><\/tr><tr><td><strong>ng of analyte in vial<\/strong><\/td><td><strong>\u00b5L of working solution added (solution concentration)<\/strong><\/td><td><strong>ng of analyte in vial<\/strong><\/td><td><strong>\u00b5L of working solution added (solution concentration)<\/strong><\/td><\/tr><tr><td>Level 1<\/td><td>1.25<\/td><td>12.5 (0.1 \u00b5g\/mL)<\/td><td>25<\/td><td>50 (0.5 \u00b5g\/mL)<\/td><\/tr><tr><td>Level 2<\/td><td>2.5<\/td><td>25 (0.1 \u00b5g\/mL)<\/td><td>50<\/td><td>100 (0.5 \u00b5g\/mL)<\/td><\/tr><tr><td>Level 3<\/td><td>5<\/td><td>10 (0.5 \u00b5g\/mL)<\/td><td>100<\/td><td>10 (10 \u00b5g\/mL)<\/td><\/tr><tr><td>Level 4<\/td><td>10<\/td><td>20 (0.5 \u00b5g\/mL)<\/td><td>200<\/td><td>20 (10 \u00b5g\/mL)<\/td><\/tr><tr><td>Level 5<\/td><td>20<\/td><td>40 (0.5 \u00b5g\/mL)<\/td><td>500<\/td><td>50 (10 \u00b5g\/mL)<\/td><\/tr><tr><td>Level 6<\/td><td>50<\/td><td>100 (0.5 \u00b5g\/mL)<\/td><td>1000<\/td><td>100 (10 \u00b5g\/mL)<\/td><\/tr><tr><td>Level 7<\/td><td>100<\/td><td>200 (0.5 \u00b5g\/mL)<\/td><td>2000<\/td><td>200 (10 \u00b5g\/mL)<\/td><\/tr><tr><td>Level 8<\/td><td>150<\/td><td>300 (0.5 \u00b5g\/mL)<\/td><td>4000<\/td><td>400 (10 \u00b5g\/mL)<\/td><\/tr><tr><td>Level 9<\/td><td>&nbsp;<\/td><td>&nbsp;<\/td><td>8000<\/td><td>800 (10 \u00b5g\/mL)<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<h3 class=\"wp-block-heading\">Method Verification Samples<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">To evaluate method performance, baby formula and coffee samples were spiked at two concentration levels based on the standard method performance requirements listed by the AOAC [6]. Baby formula samples were spiked at 5 and 50 \u00b5g\/kg (n=3), and coffee samples were spiked at 1000 and 4000 \u00b5g\/kg (n=3). Sample blanks spiked only with internal standards were also analyzed (n=3).<\/p>\n\n\n\n<h3 class=\"wp-block-heading\">Chromatographic Method<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">GC-MS analysis of furan and alkylfurans was performed on an Agilent 7890B GC paired with a 5977B MSD. GC-MS method details are shown in Tables III and IV. To enable the analysis of highly concentrated coffee samples with the wide carbon range SPME Arrows, three method parameters were modified: the GC split was set to 100:1 (with a split flow rate of 140 mL\/min); the extraction time was set to 1 min; and the volume of sodium chloride solution was adjusted to 5 mL. Note that a 1.8 mm ID inlet liner was used because the wider bore diameter is necessary to accommodate the width of SPME Arrows.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Table III:<\/strong>&nbsp;GC-MS Instrument Conditions<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td><strong>Column<\/strong><\/td><td>Rxi-624Sil MS, 30 m, 0.25 mm ID, 1.40 \u00b5m (cat.# 13868)<\/td><\/tr><tr><td><strong>Injection Mode<\/strong><\/td><td>Split (10:1 for baby formula and 100:1 for coffee)<\/td><\/tr><tr><td><strong>Liner<\/strong><\/td><td>Topaz 1.8 mm ID SPME\/straight liner (cat.# 23280)<\/td><\/tr><tr><td><strong>Inj. Temp.<\/strong><\/td><td>280 \u00b0C&nbsp;<\/td><\/tr><tr><td><strong>Split Flow<\/strong><\/td><td>\u00b7&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Baby formula: 14.0 mL\/min&nbsp;(10:1)\u00b7&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp; Coffee: 140 mL\/min (100:1)<\/td><\/tr><tr><td><strong>Purge Flow<\/strong><\/td><td>5 mL\/min&nbsp;<\/td><\/tr><tr><td><strong>Oven<\/strong><\/td><td>35 \u00b0C (hold 3 min) to 75 \u00b0C by 8 \u00b0C\/min, then to 200 \u00b0C by 25\u00b0C\/min (hold 1 min)<\/td><\/tr><tr><td><strong>Carrier Gas<\/strong><\/td><td>He, constant flow&nbsp;<\/td><\/tr><tr><td><strong>Flow Rate<\/strong><\/td><td>1.40 mL\/min&nbsp;<\/td><\/tr><tr><td><strong>Analyzer<\/strong><\/td><td>MS&nbsp;(quadrupole)<\/td><\/tr><tr><td><strong>Acquisition Type<\/strong><\/td><td>SIM&nbsp;<\/td><\/tr><tr><td><strong>Ionization Mode<\/strong><\/td><td>EI&nbsp;(70 eV)<\/td><\/tr><tr><td><strong>Transfer Line Temp.<\/strong><\/td><td>280 \u00b0C &nbsp;<\/td><\/tr><tr><td><strong>Source Temp.<\/strong><\/td><td>325 \u00b0C&nbsp;<\/td><\/tr><tr><td><strong>Quadrupole Temp.<\/strong><\/td><td>200 \u00b0C &nbsp;<\/td><\/tr><tr><td><strong>Solvent delay<\/strong><\/td><td>2.2 min<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Table IV:&nbsp;<\/strong>MS Parameters (SIM mode)<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td><strong>Segment Starting Time (min)<\/strong><\/td><td><strong>Compound (t<sub>R<\/sub>, min)<\/strong><\/td><td><strong>Ions<\/strong><\/td><td><strong>Dwell Time (ms)<\/strong><\/td><\/tr><tr><td rowspan=\"4\">2.2<\/td><td rowspan=\"2\">Furan (2.447)<\/td><td>39<\/td><td rowspan=\"4\">50<\/td><\/tr><tr><td>68*<\/td><\/tr><tr><td rowspan=\"2\">Furan-d4 (2.428)<\/td><td>42<\/td><\/tr><tr><td>72*<\/td><\/tr><tr><td rowspan=\"8\">4.2<\/td><td rowspan=\"3\">2-Methylfuran (4.536)<\/td><td>53<\/td><td rowspan=\"3\">30<\/td><\/tr><tr><td>81<\/td><\/tr><tr><td>82*<\/td><\/tr><tr><td rowspan=\"3\">3-Methylfuran (4.846)<\/td><td>53<\/td><td rowspan=\"3\">30<\/td><\/tr><tr><td>81<\/td><\/tr><tr><td>82*<\/td><\/tr><tr><td rowspan=\"2\">2-Methylfuran-d6 (4.464)<\/td><td>58<\/td><td rowspan=\"2\">30<\/td><\/tr><tr><td>88*<\/td><\/tr><tr><td rowspan=\"9\">6.6<\/td><td rowspan=\"3\">2-Ethylfuran (7.100)<\/td><td>53<\/td><td rowspan=\"3\">30<\/td><\/tr><tr><td>81*<\/td><\/tr><tr><td>96<\/td><\/tr><tr><td rowspan=\"2\">2-Ethylfuran-d5 (7.001)<\/td><td>55<\/td><td rowspan=\"2\">30<\/td><\/tr><tr><td>101*<\/td><\/tr><tr><td rowspan=\"2\">2,5-Dimethylfuran (7.243)<\/td><td>67<\/td><td rowspan=\"2\">30<\/td><\/tr><tr><td>95*<\/td><\/tr><tr><td rowspan=\"2\">2,5-Dimethylfuran-d3 (7.179)<\/td><td>84<\/td><td rowspan=\"2\">30<\/td><\/tr><tr><td>99*<\/td><\/tr><tr><td rowspan=\"4\">10.6<\/td><td rowspan=\"2\">2-Pentylfuran (11.570)<\/td><td>81<\/td><td>30<\/td><\/tr><tr><td>138*<\/td><td>30<\/td><\/tr><tr><td rowspan=\"2\">2-Pentylfuran-d11 (11.501)<\/td><td>83<\/td><td>30<\/td><\/tr><tr><td>149*<\/td><td>30<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">*Quantifier ions<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Results and Discussion<\/h2>\n\n\n\n<h3 class=\"wp-block-heading\">Comparison of Headspace, SPME Fiber, and SPME Arrow Sample Preparation<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">A comparison of various sampling strategies was conducted at the beginning of the study. SPME Arrow (wide carbon range coating); traditional SPME fibers (wide carbon range coating); and headspace sampling were evaluated in terms of area counts after sampling from 30% sodium chloride solution (10 mL) spiked with all analytes (20 ng of each analyte). As can be seen in Figure 2, the SPME Arrow provided significantly higher responses for all target analytes compared to headspace or traditional SPME fibers. Compared to SPME fibers, SPME Arrows have the additional advantage of a stainless-steel sheath and stabilizing inner rod, which prevent breakage during use. Based on the response data and rugged, steel construction, SPME Arrow was chosen for the development of a single sample preparation GC-MS method.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 2:\u00a0<\/strong>Analyte response was significantly higher with SPME Arrows than with headspace or traditional SPME fibers.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<style>.kb-image45254_392dcf-79 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<div class=\"wp-block-kadence-image kb-image45254_392dcf-79\"><figure class=\"aligncenter size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"635\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAN3916-05-1024x635.jpg\" alt=\"\" class=\"kb-img wp-image-20026\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAN3916-05-1024x635.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAN3916-05-300x186.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAN3916-05-768x477.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAN3916-05-1536x953.jpg 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAN3916-05.jpg 1800w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure><\/div>\n\n<\/div><\/div><\/div>\n\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h3 class=\"wp-block-heading\">Chromatographic Performance<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">An Rxi-624Sil MS column (30 m, 0.25 mm ID, 1.40 \u00b5m [cat.# 13868]) was used for the analysis of furan and alkylfurans because it is highly selective for volatile organics and works well for both SPME and headspace methods. The column and optimized conditions provided good chromatographic separation of all compounds, and example chromatograms for a solvent standard and spiked samples are provided in Figures 3-5. Of particular note, chromatographic separation between 2,5-dimethylfuran and 2-ethylfuran (isomeric compounds) and between 2,5-dimethylfuran-d3 and 2-ethylfuran was achieved, which allows accurate identification and quantitation (Figure 4). Chromatographic separation is important because 2,5-dimethylfuran-d3 and 2-ethylfuran share most of the ions of their mass spectra (except for ions m\/z 95 and 96) and, if they are not separated, 2,5-dimethylfuran-d3 can contribute to the most abundant qualifier ions of 2-ethylfuran (m\/z 81 and 53). All compounds were adequately separated in a 14-minute analysis.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 3:<\/strong>\u00a0Analysis of furan and alkylfurans with internal standards in a 10 ng\/mL standard. Good separation of 2-ethylfuran, 2,5-dimethylfuran-d3, and 2,5-dimethylfuran was achieved on an Rxi-624Sil MS column using the optimized gradient conditions.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_gn1218.png\" alt=\"Furan and Alkylfurans Standards and Internal Standards in Solvent on Rxi-624Sil MS\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_GN1218<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12 peak-50\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Ion (Quantifier)<\/th><th style=\"text-align: center;width: 75px\">Ion (Qualifier)<\/th><th style=\"text-align: center;width: 75px\">Dwell Time (ms)<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Furan-d4\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/6142-90-1\/Furan-d4\" target=\"_blank\" rel=\"noopener\">Furan-d4<\/a><\/td><td class=\"oth\">2.428<\/td><td class=\"oth\">72<\/td><td class=\"oth\">42<\/td><td class=\"oth\">50<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Furan\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/110-00-9\/Furan\" target=\"_blank\" rel=\"noopener\">Furan<\/a><\/td><td class=\"oth\">2.447<\/td><td class=\"oth\">68<\/td><td class=\"oth\">39<\/td><td class=\"oth\">50<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\">2-Methylfuran-d6 <\/td><td class=\"oth\">4.464<\/td><td class=\"oth\">88<\/td><td class=\"oth\">58<\/td><td class=\"oth\">30<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Methylfuran\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/534-22-5\/2-Methylfuran\" target=\"_blank\" rel=\"noopener\">2-Methylfuran<\/a><\/td><td class=\"oth\">4.536<\/td><td class=\"oth\">82<\/td><td class=\"oth\">53<\/td><td class=\"oth\">30<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 3-Methylfuran\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/930-27-8\/3-Methylfuran\" target=\"_blank\" rel=\"noopener\">3-Methylfuran<\/a><\/td><td class=\"oth\">4.846<\/td><td class=\"oth\">82<\/td><td class=\"oth\">53<\/td><td class=\"oth\">30<\/td><\/tr>\n<\/tbody><\/table>\n<table class=\"peaks col-lg-6 col-12 peak-50\"><thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Ion (Quantifier)<\/th><th style=\"text-align: center;width: 75px\">Ion (Qualifier)<\/th><th style=\"text-align: center;width: 75px\">Dwell Time (ms)<\/th><\/tr><\/thead><tbody>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\">2-Ethylfuran-d5 <\/td><td class=\"oth\">7.001<\/td><td class=\"oth\">101<\/td><td class=\"oth\">55<\/td><td class=\"oth\">30<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Ethylfuran\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/3208-16-0\/2-Ethylfuran\" target=\"_blank\" rel=\"noopener\">2-Ethylfuran<\/a><\/td><td class=\"oth\">7.100<\/td><td class=\"oth\">81<\/td><td class=\"oth\">96<\/td><td class=\"oth\">30<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\">2,5-Dimethylfuran-d3 <\/td><td class=\"oth\">7.179<\/td><td class=\"oth\">99<\/td><td class=\"oth\">84<\/td><td class=\"oth\">30<\/td><\/tr>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,5-Dimethylfuran\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/625-86-5\/2,5-Dimethylfuran\" target=\"_blank\" rel=\"noopener\">2,5-Dimethylfuran<\/a><\/td><td class=\"oth\">7.243<\/td><td class=\"oth\">95<\/td><td class=\"oth\">67<\/td><td class=\"oth\">30<\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\">2-Pentylfuran-d11 <\/td><td class=\"oth\">11.501<\/td><td class=\"oth\">149<\/td><td class=\"oth\">83<\/td><td class=\"oth\">30<\/td><\/tr>\n<tr><td class=\"num\">11.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Pentylfuran\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/3777-69-3\/2-Pentylfuran\" target=\"_blank\" rel=\"noopener\">2-Pentylfuran<\/a><\/td><td class=\"oth\">11.570<\/td><td class=\"oth\">138<\/td><td class=\"oth\">81<\/td><td class=\"oth\">30<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Rxi-624Sil MS, 30 m, 0.25 mm ID, 1.40 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/13868\" rel=\"noopener\">cat.# 13868<\/a>)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>N\/A<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>10 ng\/mL<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td>split (split ratio 10:1)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>Topaz 1.8 mm ID straight\/SPME inlet liner (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23280\" rel=\"noopener\">cat.# 23280<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Temp.:<\/th><td>280 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Split Vent Flow Rate:<\/th><td>14 mL\/min<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>35 &#176;C (hold 3 min) to 75 &#176;C at 8 &#176;C\/min to 200 &#176;C at 25 &#176;C\/min (hold 1 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Flow Rate:<\/th><td>1.4 mL\/min<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>MS<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Transfer Line Temp.:<\/th><td>280 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Analyzer Type:<\/th><td>Quadrupole <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Source Temp.:<\/th><td>325 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Quad Temp.:<\/th><td>200 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Electron Energy:<\/th><td>70 eV<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Tune Type:<\/th><td>PFTBA <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ionization Mode:<\/th><td>EI <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Agilent 7890B GC &amp; 5977B MSD<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Sample Preparation<\/th><td>Data was collected by extracting via HS-SPME from a 20 mL vial (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23083\" rel=\"noopener\">cat.# 23083<\/a>) capped with a magnetic screw-thread cap (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23091\" rel=\"noopener\">cat.# 23091<\/a>). The vial contained 10 mL of sodium chloride solution spiked with 100 ng of each analyte and internal standard. A Restek PAL SPME Arrow (120 \u00b5m carbon wide range [WR]\/PDMS; cat.# 27487) was used.<br \/><br \/>SPME Arrow sampling conditions: 10 min extraction time, 10 min incubation at 50 \u00b0C, 1 min desorption at 280 \u00b0C, agitation at 250 rpm.<br \/><br \/><\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Notes<\/th><td>A native furan\/alkylfurans standard (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/33334\" rel=\"noopener\">cat.# 33334<\/a>) is now available. Two labeled furan\/alkylfurans standard mixes are also available (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/33335 &amp; cat.# 33336\" rel=\"noopener\">cat.# 33335 &amp; cat.# 33336<\/a>).<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>13868<\/li><li>23280<\/li><li>23083<\/li><li>23091<\/li><li>33334<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>Alkylfurans<\/li><li>Baby formula<\/li><li>Coffee<\/li><li>Furan<\/li><li>HS-SPME<\/li><li>SPME Arrow<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_gn1218.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 4:<\/strong>\u00a0Analysis of furan and alkylfurans in a baby formula sample spiked at 5 \u00b5g\/kg.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_gn1219.png\" alt=\"Furan and Alkylfurans in Baby Formula Spiked at 5 \u00b5g\/kg on Rxi-624Sil MS\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_GN1219<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12 peak-50\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Ion (Quantifier)<\/th><th style=\"text-align: center;width: 75px\">Ion (Qualifier)<\/th><th style=\"text-align: center;width: 75px\">Dwell Time (ms)<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Furan-d4\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/6142-90-1\/Furan-d4\" target=\"_blank\" rel=\"noopener\">Furan-d4<\/a><\/td><td class=\"oth\">2.428<\/td><td class=\"oth\">72<\/td><td class=\"oth\">42<\/td><td class=\"oth\">50<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Furan\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/110-00-9\/Furan\" target=\"_blank\" rel=\"noopener\">Furan<\/a><\/td><td class=\"oth\">2.447<\/td><td class=\"oth\">68<\/td><td class=\"oth\">39<\/td><td class=\"oth\">50<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\">2-Methylfuran-d6 <\/td><td class=\"oth\">4.464<\/td><td class=\"oth\">88<\/td><td class=\"oth\">58<\/td><td class=\"oth\">30<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Methylfuran\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/534-22-5\/2-Methylfuran\" target=\"_blank\" rel=\"noopener\">2-Methylfuran<\/a><\/td><td class=\"oth\">4.536<\/td><td class=\"oth\">82<\/td><td class=\"oth\">53<\/td><td class=\"oth\">30<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 3-Methylfuran\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/930-27-8\/3-Methylfuran\" target=\"_blank\" rel=\"noopener\">3-Methylfuran<\/a><\/td><td class=\"oth\">4.846<\/td><td class=\"oth\">82<\/td><td class=\"oth\">53<\/td><td class=\"oth\">30<\/td><\/tr>\n<\/tbody><\/table>\n<table class=\"peaks col-lg-6 col-12 peak-50\"><thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Ion (Quantifier)<\/th><th style=\"text-align: center;width: 75px\">Ion (Qualifier)<\/th><th style=\"text-align: center;width: 75px\">Dwell Time (ms)<\/th><\/tr><\/thead><tbody>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\">2-Ethylfuran-d5 <\/td><td class=\"oth\">7.001<\/td><td class=\"oth\">101<\/td><td class=\"oth\">55<\/td><td class=\"oth\">30<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Ethylfuran\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/3208-16-0\/2-Ethylfuran\" target=\"_blank\" rel=\"noopener\">2-Ethylfuran<\/a><\/td><td class=\"oth\">7.100<\/td><td class=\"oth\">81<\/td><td class=\"oth\">96<\/td><td class=\"oth\">30<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\">2,5-Dimethylfuran-d3 <\/td><td class=\"oth\">7.179<\/td><td class=\"oth\">99<\/td><td class=\"oth\">84<\/td><td class=\"oth\">30<\/td><\/tr>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,5-Dimethylfuran\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/625-86-5\/2,5-Dimethylfuran\" target=\"_blank\" rel=\"noopener\">2,5-Dimethylfuran<\/a><\/td><td class=\"oth\">7.243<\/td><td class=\"oth\">95<\/td><td class=\"oth\">67<\/td><td class=\"oth\">30<\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\">2-Pentylfuran-d11 <\/td><td class=\"oth\">11.501<\/td><td class=\"oth\">149<\/td><td class=\"oth\">83<\/td><td class=\"oth\">30<\/td><\/tr>\n<tr><td class=\"num\">11.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Pentylfuran\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/3777-69-3\/2-Pentylfuran\" target=\"_blank\" rel=\"noopener\">2-Pentylfuran<\/a><\/td><td class=\"oth\">11.570<\/td><td class=\"oth\">138<\/td><td class=\"oth\">81<\/td><td class=\"oth\">30<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Rxi-624Sil MS, 30 m, 0.25 mm ID, 1.40 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/13868\" rel=\"noopener\">cat.# 13868<\/a>)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>N\/A<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>5  \u00b5g\/kg<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td>split (split ratio 10:1)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>Topaz 1.8 mm ID straight\/SPME inlet liner (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23280\" rel=\"noopener\">cat.# 23280<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Temp.:<\/th><td>280 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Split Vent Flow Rate:<\/th><td>14 mL\/min<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>35 &#176;C (hold 3 min) to 75 &#176;C at 8 &#176;C\/min to 200 &#176;C at 25 &#176;C\/min (hold 1 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Flow Rate:<\/th><td>1.4 mL\/min<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>MS<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Transfer Line Temp.:<\/th><td>280 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Analyzer Type:<\/th><td>Quadrupole <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Source Temp.:<\/th><td>325 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Quad Temp.:<\/th><td>200 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Electron Energy:<\/th><td>70 eV<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Tune Type:<\/th><td>PFTBA <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ionization Mode:<\/th><td>EI <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Agilent 7890B GC &amp; 5977B MSD<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Sample Preparation<\/th><td>Data was collected by extracting via HS-SPME from a 20 mL vial (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23083\" rel=\"noopener\">cat.# 23083<\/a>) capped with a magnetic screw-thread cap (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23091\" rel=\"noopener\">cat.# 23091<\/a>). The vial contained 0.5 g of baby formula and 10 mL of sodium chloride solution spiked with 2.5 ng of each analyte (5 ng\/g final concentration). Internal standards were spiked at 100 ng\/g. A Restek PAL SPME Arrow (120 \u00b5m carbon wide range [WR]\/PDMS; cat.# 27487) was used.<br \/><br \/>SPME Arrow sampling conditions: 10 min extraction time, 10 min incubation at 50 \u00b0C, 1 min desorption at 280 \u00b0C, agitation at 250 rpm.<br \/><br \/><br \/><\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Notes<\/th><td>A native furan\/alkylfurans standard (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/33334\" rel=\"noopener\">cat.# 33334<\/a>) is now available. Two labeled furan\/alkylfurans standard mixes are also available (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/33335 &amp; cat.# 33336\" rel=\"noopener\">cat.# 33335 &amp; cat.# 33336<\/a>).<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>13868<\/li><li>23280<\/li><li>23083<\/li><li>23091<\/li><li>33334<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>Alkylfurans<\/li><li>Baby formula<\/li><li>Furan<\/li><li>HS-SPME<\/li><li>SPME Arrow<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_gn1219.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 5:<\/strong>\u00a0Analysis of furan and alkylfurans in an instant coffee sample spiked at 1000 \u00b5g\/kg.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_gn1220.png\" alt=\"Furan and Alkylfurans in Instant Coffee Spiked at 1000 \u00b5g\/kg on Rxi-624Sil MS\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_GN1220<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12 peak-50\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Ion (Quantifier)<\/th><th style=\"text-align: center;width: 75px\">Ion (Qualifier)<\/th><th style=\"text-align: center;width: 75px\">Dwell Time (ms)<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Furan-d4\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/6142-90-1\/Furan-d4\" target=\"_blank\" rel=\"noopener\">Furan-d4<\/a><\/td><td class=\"oth\">2.428<\/td><td class=\"oth\">72<\/td><td class=\"oth\">42<\/td><td class=\"oth\">50<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Furan\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/110-00-9\/Furan\" target=\"_blank\" rel=\"noopener\">Furan<\/a><\/td><td class=\"oth\">2.447<\/td><td class=\"oth\">68<\/td><td class=\"oth\">39<\/td><td class=\"oth\">50<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\">2-Methylfuran-d6 <\/td><td class=\"oth\">4.464<\/td><td class=\"oth\">88<\/td><td class=\"oth\">58<\/td><td class=\"oth\">30<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Methylfuran\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/534-22-5\/2-Methylfuran\" target=\"_blank\" rel=\"noopener\">2-Methylfuran<\/a><\/td><td class=\"oth\">4.536<\/td><td class=\"oth\">82<\/td><td class=\"oth\">53<\/td><td class=\"oth\">30<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 3-Methylfuran\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/930-27-8\/3-Methylfuran\" target=\"_blank\" rel=\"noopener\">3-Methylfuran<\/a><\/td><td class=\"oth\">4.846<\/td><td class=\"oth\">82<\/td><td class=\"oth\">53<\/td><td class=\"oth\">30<\/td><\/tr>\n<\/tbody><\/table>\n<table class=\"peaks col-lg-6 col-12 peak-50\"><thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 75px\">t<sub>R<\/sub> (min)<\/th><th style=\"text-align: center;width: 75px\">Ion (Quantifier)<\/th><th style=\"text-align: center;width: 75px\">Ion (Qualifier)<\/th><th style=\"text-align: center;width: 75px\">Dwell Time (ms)<\/th><\/tr><\/thead><tbody>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\">2-Ethylfuran-d5 <\/td><td class=\"oth\">7.001<\/td><td class=\"oth\">101<\/td><td class=\"oth\">55<\/td><td class=\"oth\">30<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Ethylfuran\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/3208-16-0\/2-Ethylfuran\" target=\"_blank\" rel=\"noopener\">2-Ethylfuran<\/a><\/td><td class=\"oth\">7.100<\/td><td class=\"oth\">81<\/td><td class=\"oth\">96<\/td><td class=\"oth\">30<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\">2,5-Dimethylfuran-d3 <\/td><td class=\"oth\">7.179<\/td><td class=\"oth\">99<\/td><td class=\"oth\">84<\/td><td class=\"oth\">30<\/td><\/tr>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2,5-Dimethylfuran\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/625-86-5\/2,5-Dimethylfuran\" target=\"_blank\" rel=\"noopener\">2,5-Dimethylfuran<\/a><\/td><td class=\"oth\">7.243<\/td><td class=\"oth\">95<\/td><td class=\"oth\">67<\/td><td class=\"oth\">30<\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\">2-Pentylfuran-d11 <\/td><td class=\"oth\">11.501<\/td><td class=\"oth\">149<\/td><td class=\"oth\">83<\/td><td class=\"oth\">30<\/td><\/tr>\n<tr><td class=\"num\">11.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Pentylfuran\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/3777-69-3\/2-Pentylfuran\" target=\"_blank\" rel=\"noopener\">2-Pentylfuran<\/a><\/td><td class=\"oth\">11.570<\/td><td class=\"oth\">138<\/td><td class=\"oth\">81<\/td><td class=\"oth\">30<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Rxi-624Sil MS, 30 m, 0.25 mm ID, 1.40 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/13868\" rel=\"noopener\">cat.# 13868<\/a>)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>N\/A<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>1000 \u00b5g\/kg <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td>split (split ratio 100:1)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>Topaz 1.8 mm ID straight\/SPME inlet liner (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23280\" rel=\"noopener\">cat.# 23280<\/a>)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Temp.:<\/th><td>280 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Split Vent Flow Rate:<\/th><td>140 mL\/min<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>35 &#176;C (hold 3 min) to 75 &#176;C at 8 &#176;C\/min to 200 &#176;C at 25 &#176;C\/min (hold 1 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Flow Rate:<\/th><td>1.4 mL\/min<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>MS<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Transfer Line Temp.:<\/th><td>280 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Analyzer Type:<\/th><td>Quadrupole <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Source Temp.:<\/th><td>325 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Quad Temp.:<\/th><td>200 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Electron Energy:<\/th><td>70 eV<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Tune Type:<\/th><td>PFTBA <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ionization Mode:<\/th><td>EI <\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Agilent 7890B GC &amp; 5977B MSD<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Sample Preparation<\/th><td>Data was collected by extracting via HS-SPME from a 20 mL vial (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23083\" rel=\"noopener\">cat.# 23083<\/a>) capped with a magnetic screw-thread cap (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/23091\" rel=\"noopener\">cat.# 23091<\/a>). The vial contained 0.5 g of instant coffee and 5 mL of sodium chloride solution spiked with 500 ng of each analyte (1000 ng\/g final concentration). Internal standards were spiked at 2000 ng\/g. A Restek PAL SPME Arrow (120 \u00b5m carbon wide range [WR]\/PDMS; cat.# 27487) was used.<br \/><br \/>SPME Arrow sampling conditions: 1 min extraction time, 10 min incubation at 50 \u00b0C, 1 min desorption at 280 \u00b0C, agitation at 250 rpm.<br \/><\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Notes<\/th><td>A native furan\/alkylfurans standard (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/33334\" rel=\"noopener\">cat.# 33334<\/a>) is now available. Two labeled furan\/alkylfurans standard mixes are also available (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/33335 &amp; cat.# 33336\" rel=\"noopener\">cat.# 33335 &amp; cat.# 33336<\/a>).<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>13868<\/li><li>23280<\/li><li>23083<\/li><li>23091<\/li><li>33334<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>Alkylfurans<\/li><li>Coffee<\/li><li>Furan<\/li><li>HS-SPME<\/li><li>SPME Arrow<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_gn1220.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h3 class=\"wp-block-heading\">Linearity<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Calibration curves were constructed for all the target analytes in sodium chloride solution (30%) using appropriate deuterated analogues for each compound. The use of isotopically labeled internal standards was essential to account for differences in analyte recovery from sample matrices versus from spiked sodium chloride solution. The quantitation strategy followed in this work was a slightly modified version of the method described by Frank et al. in 2020 [5]. The low concentration calibration curve ranged from 1.25 to 150 ng of analytes in vial (8 calibration points) while the high concentration calibration curve ranged from 25 to 8000 ng of analyte in vial (9 calibration points). As can be seen in Figure 6, all analytes showed good linearity within the tested ranges, and all the coefficients of determination were above 0.99. For pentylfuran, the low concentration calibration curve ranged from 1.25 to 100 ng, and the high concentration calibration curve was linear from 25 to 2000 ng in vial.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 6:\u00a0<\/strong>Calibration curves for the analysis of furans and alkylfurans via SPME Arrow.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<style>.kb-image45254_e4cb27-d6 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<div class=\"wp-block-kadence-image kb-image45254_e4cb27-d6\"><figure class=\"aligncenter size-medium_large\"><img loading=\"lazy\" decoding=\"async\" width=\"768\" height=\"753\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAN3916-06-768x753.jpg\" alt=\"\" class=\"kb-img wp-image-20032\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAN3916-06-768x753.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAN3916-06-300x294.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAN3916-06-1024x1004.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAN3916-06-1536x1505.jpg 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/figure-article-FSAN3916-06.jpg 1800w\" sizes=\"auto, (max-width: 768px) 100vw, 768px\" \/><\/figure><\/div>\n\n<\/div><\/div><\/div>\n\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h3 class=\"wp-block-heading\">Method Verification in Baby Formula and Coffee Samples<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Baby formula and coffee both had significant concentrations of some of the analytes of interest in the blank samples. Baby formula was analyzed using the low concentration calibration curve, whereas instant coffee was analyzed using the high concentration calibration curve. Table V and VI present a summary of the collected results. All results for baby formula passed the 80-110% recovery criteria, except for 3-methylfuran, which had a recovery of 113% in the low spike. For coffee, all low concentration spikes passed, but some high bias was observed for the high concentration spikes. While further investigation into sample size, sample handling, and dilution could be done in future experiments, recoveries were still fairly close to the 110% upper limit. This indicates that with some modification SPME Arrow is likely to be an effective approach for high concentration samples as well as low concentration samples.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Table V:<\/strong>&nbsp;Analysis of furan and alkylfurans in baby formula.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td rowspan=\"2\"><strong>Analyte<\/strong><\/td><td colspan=\"2\"><strong>Blank (n=3)<\/strong><\/td><td colspan=\"2\"><strong>Low concentration (n=3),<\/strong><strong>5 \u00b5g\/kg*<\/strong><\/td><td colspan=\"2\"><strong>High concentration (n=3),<\/strong><strong>50 \u00b5g\/kg*<\/strong><\/td><\/tr><tr><td><strong>Concentration, \u00b5g\/kg<\/strong><\/td><td><strong>RSD, %<\/strong><\/td><td><strong>Accuracy, %<\/strong><\/td><td><strong>RSD, %<\/strong><\/td><td><strong>Accuracy, %<\/strong><\/td><td><strong>RSD, %<\/strong><\/td><\/tr><tr><td>Furan<\/td><td>16<\/td><td>1<\/td><td>110<\/td><td>2<\/td><td>94<\/td><td>4<\/td><\/tr><tr><td>2-Methylfuran<\/td><td>60<\/td><td>2<\/td><td>97<\/td><td>2<\/td><td>100<\/td><td>3<\/td><\/tr><tr><td>3-Methylfuran<\/td><td>&#8211;<\/td><td>&#8211;<\/td><td>113<\/td><td>3<\/td><td>107<\/td><td>6<\/td><\/tr><tr><td>2-Ethylfuran<\/td><td>67<\/td><td>2<\/td><td>93<\/td><td>3<\/td><td>105<\/td><td>5<\/td><\/tr><tr><td>2,5-Dimethylfuran<\/td><td>&#8211;<\/td><td>&#8211;<\/td><td>101<\/td><td>4<\/td><td>97<\/td><td>14<\/td><\/tr><tr><td>2-Pentylfuran<\/td><td>219<\/td><td>3<\/td><td>87<\/td><td>11<\/td><td>108<\/td><td>12<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">*Accuracy was determined as follows: ((measured concentration \u2013 concentration in blank)\/spiked concentration)*100<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Table VI:<\/strong>&nbsp;Analysis of furan and alkylfurans in instant coffee.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td rowspan=\"2\"><strong>Analyte<\/strong><\/td><td colspan=\"2\"><strong>Blank (n=3)<\/strong><\/td><td colspan=\"2\"><strong>Low concentration (n=3),<\/strong><strong>1000 \u00b5g\/kg*<\/strong><\/td><td colspan=\"2\"><strong>High concentration (n=3),<\/strong><strong>4000 \u00b5g\/kg*<\/strong><\/td><\/tr><tr><td><strong>Concentration, \u00b5g\/kg<\/strong><\/td><td><strong>RSD, %<\/strong><\/td><td><strong>Accuracy, %<\/strong><\/td><td><strong>RSD, %<\/strong><\/td><td><strong>Accuracy, %<\/strong><\/td><td><strong>RSD, %<\/strong><\/td><\/tr><tr><td>Furan<\/td><td>394<\/td><td>10<\/td><td>87<\/td><td>2<\/td><td>125<\/td><td>3<\/td><\/tr><tr><td>2-Methylfuran<\/td><td>843<\/td><td>10<\/td><td>94<\/td><td>3<\/td><td>116<\/td><td>2<\/td><\/tr><tr><td>3-Methylfuran<\/td><td>96<\/td><td>10<\/td><td>87<\/td><td>2<\/td><td>119<\/td><td>3<\/td><\/tr><tr><td>2-Ethylfuran<\/td><td>29<\/td><td>11<\/td><td>93<\/td><td>2<\/td><td>115<\/td><td>5<\/td><\/tr><tr><td>2,5-Dimethylfuran<\/td><td>46<\/td><td>11<\/td><td>98<\/td><td>2<\/td><td>120<\/td><td>4<\/td><\/tr><tr><td>2-Pentylfuran<\/td><td>&#8211;<\/td><td>&#8211;<\/td><td>83<\/td><td>3<\/td><td>83<\/td><td>11<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">*Accuracy was determined as follows: ((measured concentration \u2013 concentration in blank)\/spiked concentration)*100<\/p>\n\n\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h2 class=\"wp-block-heading\">Conclusion<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Analysis of furan and alkylfurans is complicated by a diverse array of sample types that can contain the target compounds at a wide range of concentrations. Headspace and traditional SPME fibers can be used for high and low concentration samples respectively, but this study sought to establish a single approach that could be adapted for all analyte levels. The method developed here, based on SPME Arrow sample preparation, was evaluated in matrices spiked at both low and high concentration levels: 5 and 50 \u00b5g\/kg for baby formula, and 1000 and 4000 \u00b5g\/kg for coffee. Satisfactory results in terms of linearity, accuracy, and precision were obtained in most cases. It is noted that for the analysis of highly concentrated samples; different split (1:100); extraction time (1 min); and sodium chloride solution (30%) (5 mL) conditions should be used, and further optimization may be necessary. Use of this SPME Arrow GC-MS method may allow labs to improve productivity by using a single sample preparation technique that can be adjusted to accommodate both high and low concentration samples.<\/p>\n\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<h3 class=\"wp-block-heading\">References<\/h3>\n\n\n\n<ol class=\"wp-block-list\">\n<li>European Commission recommendation 2022\/495, On monitoring the presence of furan and alkylfurans in food, 2022.&nbsp;<a href=\"https:\/\/eur-lex.europa.eu\/legal-content\/EN\/TXT\/HTML\/?uri=CELEX:32022H0495&amp;from=EN#ntr1-L_2022100EN.01006001-E0001\" target=\"_blank\" rel=\"noopener\">https:\/\/eur-lex.europa.eu\/legal-content\/EN\/TXT\/HTML\/?uri=CELEX:32022H0495&amp;from=EN#ntr1-L_2022100EN.01006001-E0001<\/a><\/li>\n\n\n\n<li>Reyes-Garc\u00e9s,&nbsp;<a href=\"https:\/\/discover.restek.com\/GNBL4573\/analysis-of-furans-and-alkylfurans-in-food-samples-part-1\/\" data-type=\"post\" data-id=\"51601\">Analysis of furan and alkylfurans in food (part 1): choosing GC-MS conditions and a sample preparation technique<\/a>, ChromaBLOGraphy, Restek Corporation, 2022.&nbsp;<\/li>\n\n\n\n<li>Reyes-Garc\u00e9s,&nbsp;<a href=\"https:\/\/discover.restek.com\/GNBL4573\/analysis-of-furans-and-alkylfurans-in-food-samples-part-2\/\" data-type=\"link\" data-id=\"https:\/\/discover.restek.com\/GNBL4573\/analysis-of-furans-and-alkylfurans-in-food-samples-part-2\/\">Analysis of furan and alkylfurans in food (part 2): optimizing experimental conditions for the analysis of food matrices using the SPME Arrow<\/a>, ChromaBLOGraphy, Restek Corporation, 2022.<\/li>\n\n\n\n<li>Reyes-Garc\u00e9s,&nbsp;Analysis of furan and alkylfurans in food samples (part 3): Adjusting the experimental conditions for the analysis of samples with high concentrations of analytes and preparing calibration curves, ChromaBLOGraphy, Restek Corporation, 2022.<\/li>\n\n\n\n<li>Frank, M. Dubois, J.F. Huertas P\u00e9rez, Detection of furan and five alkylfurans, including 2-pentylfuran, in various food matrices, J. Chromatogr. A, 1622 (2020) 461119.&nbsp;<a href=\"https:\/\/doi.org\/10.1016\/j.chroma.2020.461119\" target=\"_blank\" rel=\"noopener\">https:\/\/doi.org\/10.1016\/j.chroma.2020.461119<\/a><\/li>\n\n\n\n<li>AOAC SMPR 2019.004, Standard method performance requirements (SMPRs) for furan and alkyl furans in coffee, baby foods, infant formula, cereals, and fruit juices, final version: October 3, 2019, AOAC International, Rockville, MD, U.S.&nbsp;<a href=\"https:\/\/www.aoac.org\/wp-content\/uploads\/2019\/09\/SMPR_Furans_v6-For-Comment.pdf\" target=\"_blank\" rel=\"noopener\">https:\/\/www.aoac.org\/wp-content\/uploads\/2019\/09\/SMPR_Furans_v6-For-Comment.pdf<\/a><\/li>\n<\/ol>\n\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n<style>.kb-row-layout-id45254_2e5dc7-30 > .kt-row-column-wrap{align-content:start;}:where(.kb-row-layout-id45254_2e5dc7-30 > .kt-row-column-wrap) > .wp-block-kadence-column{justify-content:start;}.kb-row-layout-id45254_2e5dc7-30 > .kt-row-column-wrap{column-gap:var(--global-kb-gap-md, 2rem);row-gap:var(--global-kb-gap-md, 2rem);padding-top:var(--global-kb-spacing-sm, 1.5rem);padding-bottom:var(--global-kb-spacing-sm, 1.5rem);grid-template-columns:minmax(0, 2fr) minmax(0, 1fr);}.kb-row-layout-id45254_2e5dc7-30{border-top-left-radius:10px;border-top-right-radius:10px;border-bottom-right-radius:10px;border-bottom-left-radius:10px;overflow:clip;isolation:isolate;}.kb-row-layout-id45254_2e5dc7-30 > .kt-row-layout-overlay{border-top-left-radius:10px;border-top-right-radius:10px;border-bottom-right-radius:10px;border-bottom-left-radius:10px;}.kb-row-layout-id45254_2e5dc7-30{background-color:#fffaf1;}.kb-row-layout-id45254_2e5dc7-30 > .kt-row-layout-overlay{opacity:0.30;}@media all and (max-width: 1024px){.kb-row-layout-id45254_2e5dc7-30 > .kt-row-column-wrap{grid-template-columns:minmax(0, 2fr) minmax(0, 1fr);}}@media all and (max-width: 767px){.kb-row-layout-id45254_2e5dc7-30 > .kt-row-column-wrap{grid-template-columns:minmax(0, 1fr);}}<\/style><div class=\"kb-row-layout-wrap kb-row-layout-id45254_2e5dc7-30 alignnone kt-row-has-bg wp-block-kadence-rowlayout\"><div class=\"kt-row-column-wrap kt-has-2-columns kt-row-layout-left-golden kt-tab-layout-inherit kt-mobile-layout-row kt-row-valign-top\">\n<style>.kadence-column45254_f24283-93 > .kt-inside-inner-col{display:flex;}.kadence-column45254_f24283-93 > .kt-inside-inner-col{padding-top:var(--global-kb-spacing-xs, 1rem);padding-right:var(--global-kb-spacing-xs, 1rem);padding-bottom:var(--global-kb-spacing-xs, 1rem);padding-left:var(--global-kb-spacing-xs, 1rem);}.kadence-column45254_f24283-93 > .kt-inside-inner-col,.kadence-column45254_f24283-93 > .kt-inside-inner-col:before{border-top-left-radius:0px;border-top-right-radius:0px;border-bottom-right-radius:0px;border-bottom-left-radius:0px;}.kadence-column45254_f24283-93 > .kt-inside-inner-col{column-gap:var(--global-kb-gap-sm, 1rem);}.kadence-column45254_f24283-93 > .kt-inside-inner-col{flex-direction:column;justify-content:center;}.kadence-column45254_f24283-93 > .kt-inside-inner-col > .aligncenter{width:100%;}.kt-row-column-wrap > .kadence-column45254_f24283-93{align-self:center;}.kt-inner-column-height-full:not(.kt-has-1-columns) > .wp-block-kadence-column.kadence-column45254_f24283-93{align-self:auto;}.kt-inner-column-height-full:not(.kt-has-1-columns) > .wp-block-kadence-column.kadence-column45254_f24283-93 > .kt-inside-inner-col{flex-direction:column;justify-content:center;}.kadence-column45254_f24283-93 > .kt-inside-inner-col:before{opacity:0.3;}.kadence-column45254_f24283-93{position:relative;}@media all and (max-width: 1024px){.kt-row-column-wrap > .kadence-column45254_f24283-93{align-self:center;}}@media all and (max-width: 1024px){.kt-inner-column-height-full:not(.kt-has-1-columns) > .wp-block-kadence-column.kadence-column45254_f24283-93{align-self:auto;}}@media all and (max-width: 1024px){.kt-inner-column-height-full:not(.kt-has-1-columns) > .wp-block-kadence-column.kadence-column45254_f24283-93 > .kt-inside-inner-col{flex-direction:column;justify-content:center;}}@media all and (max-width: 1024px){.kadence-column45254_f24283-93 > .kt-inside-inner-col{flex-direction:column;justify-content:center;}}@media all and (max-width: 767px){.kt-row-column-wrap > .kadence-column45254_f24283-93{align-self:center;}.kt-inner-column-height-full:not(.kt-has-1-columns) > .wp-block-kadence-column.kadence-column45254_f24283-93{align-self:auto;}.kt-inner-column-height-full:not(.kt-has-1-columns) > .wp-block-kadence-column.kadence-column45254_f24283-93 > .kt-inside-inner-col{flex-direction:column;justify-content:center;}.kadence-column45254_f24283-93 > .kt-inside-inner-col{flex-direction:column;justify-content:center;}}<\/style>\n<div class=\"wp-block-kadence-column kadence-column45254_f24283-93\"><div class=\"kt-inside-inner-col\">\n<h3 class=\"wp-block-heading has-text-color has-link-color wp-elements-a63340580c5d302b34f73ac16aec943d\" style=\"color:#de772c\">Looking for furan\/alkylfurans standards?<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">A native furan\/alkylfurans standard (<a href=\"https:\/\/www.restek.com\/p\/33334\" target=\"_blank\" rel=\"noreferrer noopener\">cat.# 33334<\/a>) is now available. Two labeled furan\/alkylfurans standard mixes are also available (<a href=\"https:\/\/www.restek.com\/p\/33335\" data-type=\"link\" data-id=\"https:\/\/www.restek.com\/p\/33335\" target=\"_blank\" rel=\"noopener\">cat.# 33335<\/a> &amp; <a href=\"https:\/\/www.restek.com\/p\/33336\" data-type=\"link\" data-id=\"https:\/\/www.restek.com\/p\/33336\" target=\"_blank\" rel=\"noopener\">cat.# 33336<\/a>).<\/p>\n<\/div><\/div>\n\n\n<style>.kadence-column45254_df90e6-d5 > .kt-inside-inner-col{display:flex;}.kadence-column45254_df90e6-d5 > .kt-inside-inner-col,.kadence-column45254_df90e6-d5 > .kt-inside-inner-col:before{border-top-left-radius:0px;border-top-right-radius:0px;border-bottom-right-radius:0px;border-bottom-left-radius:0px;}.kadence-column45254_df90e6-d5 > .kt-inside-inner-col{column-gap:var(--global-kb-gap-sm, 1rem);}.kadence-column45254_df90e6-d5 > .kt-inside-inner-col{flex-direction:column;justify-content:center;}.kadence-column45254_df90e6-d5 > .kt-inside-inner-col > .aligncenter{width:100%;}.kt-row-column-wrap > .kadence-column45254_df90e6-d5{align-self:center;}.kt-inner-column-height-full:not(.kt-has-1-columns) > .wp-block-kadence-column.kadence-column45254_df90e6-d5{align-self:auto;}.kt-inner-column-height-full:not(.kt-has-1-columns) > .wp-block-kadence-column.kadence-column45254_df90e6-d5 > .kt-inside-inner-col{flex-direction:column;justify-content:center;}.kadence-column45254_df90e6-d5 > .kt-inside-inner-col:before{opacity:0.3;}.kadence-column45254_df90e6-d5{position:relative;}@media all and (max-width: 1024px){.kt-row-column-wrap > .kadence-column45254_df90e6-d5{align-self:center;}}@media all and (max-width: 1024px){.kt-inner-column-height-full:not(.kt-has-1-columns) > .wp-block-kadence-column.kadence-column45254_df90e6-d5{align-self:auto;}}@media all and (max-width: 1024px){.kt-inner-column-height-full:not(.kt-has-1-columns) > .wp-block-kadence-column.kadence-column45254_df90e6-d5 > .kt-inside-inner-col{flex-direction:column;justify-content:center;}}@media all and (max-width: 1024px){.kadence-column45254_df90e6-d5 > .kt-inside-inner-col{flex-direction:column;justify-content:center;}}@media all and (max-width: 767px){.kt-row-column-wrap > .kadence-column45254_df90e6-d5{align-self:center;}.kt-inner-column-height-full:not(.kt-has-1-columns) > .wp-block-kadence-column.kadence-column45254_df90e6-d5{align-self:auto;}.kt-inner-column-height-full:not(.kt-has-1-columns) > .wp-block-kadence-column.kadence-column45254_df90e6-d5 > .kt-inside-inner-col{flex-direction:column;justify-content:center;}.kadence-column45254_df90e6-d5 > .kt-inside-inner-col{flex-direction:column;justify-content:center;}}<\/style>\n<div class=\"wp-block-kadence-column kadence-column45254_df90e6-d5\"><div class=\"kt-inside-inner-col\"><style>.kb-image45254_51474a-14:not(.kb-image-is-ratio-size) .kb-img, .kb-image45254_51474a-14.kb-image-is-ratio-size{padding-top:var(--global-kb-spacing-xs, 1rem);padding-right:var(--global-kb-spacing-xs, 1rem);padding-bottom:var(--global-kb-spacing-xs, 1rem);padding-left:var(--global-kb-spacing-xs, 1rem);}.kb-image45254_51474a-14 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<div class=\"wp-block-kadence-image kb-image45254_51474a-14\"><figure class=\"aligncenter size-medium\"><img loading=\"lazy\" decoding=\"async\" width=\"300\" height=\"92\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/teaser-product-reference-standards-furan-alkylfurans-300x92.png\" alt=\"\" class=\"kb-img wp-image-29782\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/teaser-product-reference-standards-furan-alkylfurans-300x92.png 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/teaser-product-reference-standards-furan-alkylfurans.png 384w\" sizes=\"auto, (max-width: 300px) 100vw, 300px\" \/><\/figure><\/div>\n<\/div><\/div>\n\n<\/div><\/div>\n\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n        <div class=\"cpb\">\n            <h3 class=\"cpb-heading\">\u76f8\u5173\u4ea7\u54c1<\/h3>\n            <hr class=\"cpb-heading-underline\" \/>\n            <div class=\"cpb-list\">\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/13868\">                                \u8d27\u53f7\uff1a 13868                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Rxi-624Sil MS GC Capillary Column, 30 m, 0.25 mm ID, 1.40 \u00b5m<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/13868\">\u67e5\u770b\u4ea7\u54c1<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23280\">                                \u8d27\u53f7\uff1a 23280                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Topaz, Straight\/SPME Inlet Liner, 1.8 mm x 6.5 x 78.5, for Agilent GCs, Premium Deactivation, 5-pk.<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23280\">\u67e5\u770b\u4ea7\u54c1<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23083\">                                \u8d27\u53f7\uff1a 23083                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Headspace Screw-Thread Vials, 18 mm, 20 mL, Clear, 1000-pk.<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23083\">\u67e5\u770b\u4ea7\u54c1<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23091\">                                \u8d27\u53f7\uff1a 23091                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Magnetic Screw-Thread Headspace Caps, 18 mm and Blue PTFE\/Silicone, 1.5 mm thick Septa for SPME, 1000-pk.<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23091\">\u67e5\u770b\u4ea7\u54c1<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/33334\">                                \u8d27\u53f7\uff1a 33334                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Furan\/Alkylfurans Standard, 100 \u00b5g\/mL, P&amp;T Methanol, 1 mL\/ampul<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/33334\">\u67e5\u770b\u4ea7\u54c1<\/a>\n                                                    <\/div>\n                    <\/div>\n                            <\/div>\n        <\/div>\n        \n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>Food safety labs can significantly improve productivity by using SPME Arrow to test diverse sample types containing furan and alkylfurans at different levels compared to using legacy headspace GC methods or methods that require both headspace and SPME.<\/p>\n","protected":false},"author":14,"featured_media":6967,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[13],"tags":[],"industries-application":[],"post-badge":[],"resource-type":[],"product-library":[],"resource-technique":[],"ppma_author":[603],"class_list":["post-45254","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-application-notes"],"acf":[],"taxonomy_info":{"category":[{"value":13,"label":"Application Notes"}]},"featured_image_src_large":["https:\/\/discover.restek.com\/wp-content\/uploads\/feature-FSAN3916-1024x536.jpg",1024,536,true],"author_info":{"display_name":"Daniel","author_link":"https:\/\/discover.restek.com\/zh-hans\/author\/daniel-voralia\/"},"comment_info":0,"category_info":[{"term_id":13,"name":"Application Notes","slug":"application-notes","term_group":0,"term_taxonomy_id":13,"taxonomy":"category","description":"","parent":0,"count":88,"filter":"raw","cat_ID":13,"category_count":88,"category_description":"","cat_name":"Application Notes","category_nicename":"application-notes","category_parent":0}],"tag_info":false,"authors":[{"term_id":603,"user_id":0,"is_guest":1,"slug":"nathaly-reyes-garces-phd","display_name":"Nathaly Reyes-Garc\u00e9s, PhD","avatar_url":"https:\/\/secure.gravatar.com\/avatar\/?s=96&d=mm&r=g","author_category":"1","first_name":"Nathaly","last_name":"Reyes-Garc\u00e9s","user_url":"","job_title":"","description":""}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts\/45254","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/users\/14"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/comments?post=45254"}],"version-history":[{"count":20,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts\/45254\/revisions"}],"predecessor-version":[{"id":98426,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts\/45254\/revisions\/98426"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/media\/6967"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/media?parent=45254"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/categories?post=45254"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/tags?post=45254"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/industries-application?post=45254"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/post-badge?post=45254"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/resource-type?post=45254"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/product-library?post=45254"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/resource-technique?post=45254"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/ppma_author?post=45254"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}