{"id":51689,"date":"2023-09-26T00:00:00","date_gmt":"2023-09-26T00:00:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/gas-generators-part-4-fid-gas-stations\/"},"modified":"2026-01-28T15:53:00","modified_gmt":"2026-01-28T15:53:00","slug":"gas-generators-part-4-fid-gas-stations","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/zh-hans\/blog-fr\/gnbl4674\/gas-generators-part-4-fid-gas-stations","title":{"rendered":"Gas Generators \u2013 Part 4 \u2013 FID Gas Stations"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">In my final installment of gas generators, I will discuss combination gas generators called FID Gas Stations. &nbsp;For those of you not familiar with gas chromatographs or their detectors, FID refers to a Flame Ionization Detector. &nbsp;Because hydrogen and air are used to produce the FID flame, those are the gases produced by these generators. &nbsp;The hydrogen acts as the fuel gas, and the air supplies the oxidizer (oxygen). &nbsp;The GC\/FID provides the ignition source.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">If you are wondering how these generators operate, I suggest reviewing two previous blog posts I wrote on gas generators, <a href=\"https:\/\/discover.restek.com\/blogs\/GNBL4670\/gas-generators-part-1-hydrogen-generators\" data-type=\"link\" data-id=\"https:\/\/discover.restek.com\/blogs\/GNBL4670\/gas-generators-part-1-hydrogen-generators\"><strong>Gas Generators \u2013 Part 1 \u2013 Hydrogen Generators<\/strong><\/a> and <a href=\"https:\/\/discover.restek.com\/blogs\/GNBL4673\/gas-generators-part-3-zero-air-generators\" data-type=\"link\" data-id=\"https:\/\/discover.restek.com\/blogs\/GNBL4673\/gas-generators-part-3-zero-air-generators\"><strong>Gas Generators \u2013 Part 3 \u2013 Zero Air Generators<\/strong><\/a>. \u00a0The hydrogen section of the FID Gas Station operates similarly to a PEM hydrogen generator. \u00a0Water molecules are split into oxygen and hydrogen and the oxygen is vented. \u00a0The zero-air generator section operates by pushing compressed air through a prefilter (to remove oil, water and larger particulates), then through a catalyst module (which oxidizes hydrocarbons into carbon dioxide and water vapor), and finally through final particulate filter (to remove very small particulates (0.5 micron)).\u00a0<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Restek sells two models, the FID-1000 (Restek part # <strong>20177<\/strong>) and the FID-2500 (Restek custom part # 580073). &nbsp;You can find the Parker Installation, Operation and Maintenance manual here. <a href=\"https:\/\/www.parker.com\/content\/dam\/Parker-com\/Literature\/IGFG\/PDF-Files\/TI_FID1000J_Gas_Stations.pdf\" target=\"_blank\" rel=\"noreferrer noopener\">TI_FID1000J_Gas_Stations.pdf (parker.com)<\/a><\/p>\n\n\n\n<p class=\"wp-block-paragraph\">There are several important specifications which I will mention at this time. &nbsp;<\/p>\n\n\n\n<ol style=\"list-style-type:upper-alpha\" class=\"wp-block-list\">\n<li>These units are bulky, so the recommended installation is on a bench top. &nbsp;Mounting onto a wall is not recommended, nor is suspending these units from a ceiling.<\/li>\n\n\n\n<li>These units should be installed in a climate-controlled environment which has a temperature range between 60\u00b0F and 100\u00b0F.<\/li>\n\n\n\n<li>The water source should be deionized water with a resistivity of at least 5 Mega-Ohm\/cm. &nbsp;Restek part # <strong>22000<\/strong> (Parker HydroGen Mate DI Water System) can provide water which meets specifications. &nbsp;<\/li>\n\n\n\n<li>A source of clean, dry compressed air with a pressure between 40 and 125psi is required.<\/li>\n<\/ol>\n\n\n\n<figure class=\"wp-block-image size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"693\" height=\"693\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-gas-generators-part-4-fid-gas-stations-01.jpg\" alt=\"\" class=\"wp-image-14596\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-gas-generators-part-4-fid-gas-stations-01.jpg 693w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-gas-generators-part-4-fid-gas-stations-01-300x300.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-gas-generators-part-4-fid-gas-stations-01-150x150.jpg 150w\" sizes=\"auto, (max-width: 693px) 100vw, 693px\" \/><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>FID-1000<\/strong><\/p>\n\n\n\n<p class=\"wp-block-paragraph\">The FID-1000 is capable of 90cc\/min of 99.9995% <strong>hydrogen<\/strong> with a maximum output pressure of 60psi. &nbsp;It is capable of 1000cc\/min <strong>zero-grade air<\/strong> with a maximum output pressure of 125psi. &nbsp;Zero-grade air purity is defined as containing less than &lt;0.1ppm methane (hydrocarbon) output. &nbsp;<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>FID-2500<\/strong><\/p>\n\n\n\n<p class=\"wp-block-paragraph\">The FID-2500 is capable of 250cc\/min of 99.9995% <strong>hydrogen<\/strong> with a maximum output pressure of 60psi. &nbsp;It is capable of 2500cc\/min of <strong>zero-grade air<\/strong> with a maximum output pressure of 125psi. &nbsp;As with the FID-1000, zero-grade air purity for the FID-2500 is defined as containing less than &lt;0.1ppm methane (hydrocarbon) output.<\/p>\n\n\n\n<figure class=\"wp-block-image size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"509\" height=\"509\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-gas-generators-part-4-fid-gas-stations-02.jpg\" alt=\"\" class=\"wp-image-14602\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-gas-generators-part-4-fid-gas-stations-02.jpg 509w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-gas-generators-part-4-fid-gas-stations-02-300x300.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-gas-generators-part-4-fid-gas-stations-02-150x150.jpg 150w\" sizes=\"auto, (max-width: 509px) 100vw, 509px\" \/><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">Please review page 21 of the <a href=\"https:\/\/www.parker.com\/content\/dam\/Parker-com\/Literature\/IGFG\/PDF-Files\/TI_FID1000J_Gas_Stations.pdf\" target=\"_blank\" rel=\"noopener\">manual<\/a> for System Specifications and Replacement Parts. &nbsp;Maintenance for both models include the following:<\/p>\n\n\n\n<ol class=\"wp-block-list\">\n<li>Refilling the water reservoir once a week with deionized water. &nbsp;Consider using Restek part # <strong>22000<\/strong> (Parker model # <a href=\"https:\/\/www.parker.com\/content\/dam\/Parker-com\/Literature\/S3.2.016a---HydroGenMate_72-230.pdf\" target=\"_blank\" rel=\"noreferrer noopener\"><strong>72-230<\/strong><\/a>) as the source of the deionized water. &nbsp;The <strong>annual<\/strong> maintenance kit for the Parker HydroGen Mate is Restek <strong>22001<\/strong> (Parker 72236). &nbsp;This kit contains one prefilter, two resin cartridges, and one final filter.<\/li>\n\n\n\n<li>Replacing the resin bed cartridge approximately every 6-months, or when the color changes 75%. &nbsp;The Restek part # for the resin bed cartridge is <strong>24914<\/strong> (Parker # B02-0323).<\/li>\n\n\n\n<li>Changing the desiccant cartridge when the orange-colored beads become clear. &nbsp;The Restek part # for the desiccant cartridge is <strong>21671<\/strong> (Parker # 1647727)<\/li>\n\n\n\n<li>Replacement of filters annually. &nbsp;Restek kit <strong>24915<\/strong> (Parker MKFID1000) contains the resin bed cartridge listed in #2 (above), the desiccant cartridge listed in #3 (above), and a prefilter and a final filter.<\/li>\n\n\n\n<li>Replacement of the catalyst tower every 2-years. &nbsp;The Parker part number for the FID-1000 catalyst tower is either B02-0298 or B02-0437 (depending on the unit\u2019s serial number), and B02-0361 for the FID-2500.<\/li>\n<\/ol>\n\n\n\n<p class=\"wp-block-paragraph\">Just a few more comments before I wrap up this post, and my series on gas generators.<\/p>\n\n\n\n<ol style=\"list-style-type:upper-alpha\" class=\"wp-block-list\">\n<li>The hydrogen produced from the FID Gas Stations is not recommended for use as the GC carrier gas. &nbsp;You will want to select a <a href=\"https:\/\/www.restek.com\/search?Ntt=PEMPD&amp;Ns=bestMatch\" target=\"_blank\" rel=\"noreferrer noopener\"><strong>PEMPD<\/strong><\/a> hydrogen generator instead. &nbsp;<\/li>\n\n\n\n<li>Restek may be able to provide, as a custom part number, any product in this post in which only a Parker part number is listed (such as with the catalyst towers). &nbsp;Please <a href=\"https:\/\/www.restek.com\/contact-us\" target=\"_blank\" rel=\"noopener\"><strong>contact<\/strong><\/a> your Restek sales representative or distributor to obtain a quote.<\/li>\n\n\n\n<li>If either the hydrogen or zero-grade air section of this generator needs serviced, you may not have a source of that gas available until the unit is repaired. &nbsp;As a result, you should always have a back-up plan, as you would with any critical gas supply, whether you keep a few gas cylinders available onsite, or have a separate generator for each gas. &nbsp;Some customers prefer owning two separate gas generators for this reason.&nbsp;<\/li>\n\n\n\n<li>As with all gas generators, if you need assistance with selection, please have the following information available.\n<ol class=\"wp-block-list\">\n<li>Gas required.<\/li>\n\n\n\n<li>Purity required.<\/li>\n\n\n\n<li>Flow rate required.<\/li>\n\n\n\n<li>Output pressure required.<\/li>\n<\/ol>\n<\/li>\n<\/ol>\n\n\n\n<p class=\"wp-block-paragraph\">Thank you for reading my series on gas generators. &nbsp;Let us know if you have any questions.<\/p>\n\n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n        <div class=\"cpb\">\n            <h3 class=\"cpb-heading\">\u76f8\u5173\u4ea7\u54c1<\/h3>\n            <hr class=\"cpb-heading-underline\" \/>\n            <div class=\"cpb-list\">\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/20177\">                                \u8d27\u53f7\uff1a 20177                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Gas Station\u00a0for Parker Model FID-1000<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/20177\">\u67e5\u770b\u4ea7\u54c1<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/22000\">                                \u8d27\u53f7\uff1a 22000                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">HydroGen Mate DI Water System, Parker, Whatman Model 72-230<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/22000\">\u67e5\u770b\u4ea7\u54c1<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/22001\">                                \u8d27\u53f7\uff1a 22001                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Maintenance Kit for Parker HydroGen Mate DI Water System, Whatman Model 72-230<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/22001\">\u67e5\u770b\u4ea7\u54c1<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/24914\">                                \u8d27\u53f7\uff1a 24914                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Resin Bed Cartridge, for Hydrogen Generators in FID-1000 and FID-2500 Gas Stations<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/24914\">\u67e5\u770b\u4ea7\u54c1<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/21671\">                                \u8d27\u53f7\uff1a 21671                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Replacement Desiccant Cartridge for Parker FID-1000 and FID-2500 Gas Stations<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/21671\">\u67e5\u770b\u4ea7\u54c1<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/24915\">                                \u8d27\u53f7\uff1a 24915                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">FID Gas Station Annual Maintenance Kit<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/24915\">\u67e5\u770b\u4ea7\u54c1<\/a>\n                                                    <\/div>\n                    <\/div>\n                            <\/div>\n        <\/div>\n        \n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>In this final installment of our gas generator series, we look at FID gas stations.<\/p>\n","protected":false},"author":5,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[791],"tags":[],"industries-application":[],"post-badge":[],"resource-type":[],"product-library":[],"resource-technique":[2294],"ppma_author":[582],"class_list":["post-51689","post","type-post","status-publish","format-standard","hentry","category-blog-fr","resource-technique-gas-chromatography-gc"],"acf":[],"taxonomy_info":{"category":[{"value":791,"label":"Blog"}],"resource-technique":[{"value":2294,"label":"Gas Chromatography (GC)"}]},"featured_image_src_large":false,"author_info":{"display_name":"Brenda Seprish","author_link":"https:\/\/discover.restek.com\/zh-hans\/author\/brenda-coganrestek-com\/"},"comment_info":0,"category_info":[{"term_id":791,"name":"Blog","slug":"blog-fr","term_group":0,"term_taxonomy_id":791,"taxonomy":"category","description":"","parent":0,"count":457,"filter":"raw","cat_ID":791,"category_count":457,"category_description":"","cat_name":"Blog","category_nicename":"blog-fr","category_parent":0}],"tag_info":false,"authors":[{"term_id":582,"user_id":0,"is_guest":1,"slug":"alan-sensue","display_name":"Alan Sensue","avatar_url":"https:\/\/secure.gravatar.com\/avatar\/?s=96&d=mm&r=g","author_category":"1","first_name":"Alan","last_name":"Sensue","user_url":"","job_title":"","description":""}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts\/51689","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/users\/5"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/comments?post=51689"}],"version-history":[{"count":7,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts\/51689\/revisions"}],"predecessor-version":[{"id":85305,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts\/51689\/revisions\/85305"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/media?parent=51689"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/categories?post=51689"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/tags?post=51689"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/industries-application?post=51689"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/post-badge?post=51689"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/resource-type?post=51689"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/product-library?post=51689"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/resource-technique?post=51689"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/ppma_author?post=51689"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}