{"id":51777,"date":"2020-05-20T14:30:00","date_gmt":"2020-05-20T14:30:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen\/"},"modified":"2026-01-28T15:58:56","modified_gmt":"2026-01-28T15:58:56","slug":"proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/zh-hans\/blogs\/gnbl4772\/proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen","title":{"rendered":"ProEZGC Chromatogram Modeler \u2013 there is much more to the program than just the Welcome Screen"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">When you log into the ProEZGC Chromatogram Modeler software, you will see the screen below, which is what I refer to as the Welcome Screen. This is the starting point for all, and unfortunately all some customers will think there is to the ProEZGC software.<\/p>\n\n\n\n<figure class=\"wp-block-image size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"800\" height=\"624\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen-01.jpg\" alt=\"\" class=\"wp-image-16570\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen-01.jpg 800w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen-01-300x234.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen-01-768x599.jpg 768w\" sizes=\"auto, (max-width: 800px) 100vw, 800px\" \/><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">For those of you who have used ProEZGC, you know that all you need to do is to add a compound name or CAS# into the box titled&nbsp;<strong>Search by Name or CAS#<\/strong>&nbsp;to get started (remember, only one compound name or CAS# per line).<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">For this post, I would like you all to see that there is more to ProEZGC than just the screen above and the screen below. In this example, I typed in the names of four common compounds\/solvents and clicked the blue&nbsp;<strong>Solve<\/strong>&nbsp;button.<\/p>\n\n\n<div class=\"wp-block-image\">\n<figure class=\"aligncenter size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"800\" height=\"654\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen-02.jpg\" alt=\"\" class=\"wp-image-16576\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen-02.jpg 800w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen-02-300x245.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen-02-768x628.jpg 768w\" sizes=\"auto, (max-width: 800px) 100vw, 800px\" \/><\/figure>\n<\/div>\n\n\n<p class=\"wp-block-paragraph\">After reviewing the search results for all stationary phases, I selected the results for the Rtx-502.2 column.<\/p>\n\n\n<div class=\"wp-block-image\">\n<figure class=\"aligncenter size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"800\" height=\"654\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen-03.jpg\" alt=\"\" class=\"wp-image-16582\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen-03.jpg 800w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen-03-300x245.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen-03-768x628.jpg 768w\" sizes=\"auto, (max-width: 800px) 100vw, 800px\" \/><\/figure>\n<\/div>\n\n\n<p class=\"wp-block-paragraph\">I then clicked on the&nbsp;<strong>Conditions<\/strong>&nbsp;tab (the dark blue tab between&nbsp;<strong>Compounds<\/strong>&nbsp;and&nbsp;<strong>My EZGC<\/strong>) and were presented with a new page (see below).<\/p>\n\n\n<div class=\"wp-block-image\">\n<figure class=\"aligncenter size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"800\" height=\"608\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen-04.jpg\" alt=\"\" class=\"wp-image-16588\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen-04.jpg 800w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen-04-300x228.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen-04-768x584.jpg 768w\" sizes=\"auto, (max-width: 800px) 100vw, 800px\" \/><\/figure>\n<\/div>\n\n\n<p class=\"wp-block-paragraph\">From this page you can choose different dimensions of the same column (in this case, the Rtx-502.2), modify the GC oven temperature program, and if you select&nbsp;<strong>Custom<\/strong>&nbsp;at the bottom of the page in the&nbsp;<strong>Results<\/strong>&nbsp;section, you can change carrier gas flow rates (see screen below) and a few other parameters.<\/p>\n\n\n<div class=\"wp-block-image\">\n<figure class=\"aligncenter size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"800\" height=\"614\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen-05.jpg\" alt=\"\" class=\"wp-image-16594\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen-05.jpg 800w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen-05-300x230.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen-05-768x589.jpg 768w\" sizes=\"auto, (max-width: 800px) 100vw, 800px\" \/><\/figure>\n<\/div>\n\n\n<p class=\"wp-block-paragraph\">Notice the blue arrow under&nbsp;<strong>Control Parameters<\/strong>&nbsp;and to the right of&nbsp;<strong>Column Flow&nbsp;<\/strong>in the screen above?&nbsp; So what does the arrow represent? It allows you to change the remaining top parameters (<strong>Column Flow, Average Velocity, Holdup Time<\/strong>&nbsp;and\/or&nbsp;<strong>Inlet Pressure<\/strong>&nbsp;&#8211; whichever parameters the arrow is&nbsp;<em>not<\/em>&nbsp;pointing to) in the&nbsp;<strong>Control Parameters<\/strong>&nbsp;section without affecting the parameter the arrow&nbsp;<em>is<\/em>&nbsp;pointing to.&nbsp; See example below.&nbsp; Notice that I moved the arrow from&nbsp;<strong>Column Flow<\/strong>&nbsp;to&nbsp;<strong>Average Velocity<\/strong>&nbsp;by double-clicking in the empty space to the left of 39.70 cm\/sec.<\/p>\n\n\n<div class=\"wp-block-image\">\n<figure class=\"aligncenter size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"800\" height=\"609\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen-06.jpg\" alt=\"\" class=\"wp-image-16600\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen-06.jpg 800w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen-06-300x228.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen-06-768x585.jpg 768w\" sizes=\"auto, (max-width: 800px) 100vw, 800px\" \/><\/figure>\n<\/div>\n\n\n<p class=\"wp-block-paragraph\">By keeping the <strong>Average Velocity<\/strong>&nbsp;the same and changing the length of the capillary column, notice how the&nbsp;<strong>Column Flow<\/strong>&nbsp;(carrier gas flow rate value) increases, as do the Holdup Time* and Inlet Pressure.<\/p>\n\n\n\n<ul class=\"wp-block-list\">\n<li>Holdup Time is the time it takes for a non-retained compound to travel through the column.<\/li>\n<\/ul>\n\n\n<div class=\"wp-block-image\">\n<figure class=\"aligncenter size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"800\" height=\"611\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen-07.jpg\" alt=\"\" class=\"wp-image-16606\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen-07.jpg 800w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen-07-300x229.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen-07-768x587.jpg 768w\" sizes=\"auto, (max-width: 800px) 100vw, 800px\" \/><\/figure>\n<\/div>\n\n\n<p class=\"wp-block-paragraph\">One last comment; under&nbsp;<strong>Results<\/strong>&nbsp;you have three selections to pick from.&nbsp; The default is&nbsp;<strong>Speed<\/strong>, and for most analysis this is fine.&nbsp; As the name implies, when this button is selected, analysis times are minimized.&nbsp; I showed you all how to select the&nbsp;<strong>Custom<\/strong>&nbsp;button for the most control of carrier gas parameters.&nbsp; My favorite button, however, is&nbsp;<strong>Efficiency<\/strong>.&nbsp; Analysis times may be a little longer, but separations are usually improved, and that is my goal, to show a customer maximum compound separation with reasonable analysis times.<\/p>\n\n\n<div class=\"wp-block-image\">\n<figure class=\"aligncenter size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"800\" height=\"626\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen-08.jpg\" alt=\"\" class=\"wp-image-16612\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen-08.jpg 800w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen-08-300x235.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-proezgc-chromatogram-modeler-there-is-much-more-to-the-program-than-just-the-welcome-screen-08-768x601.jpg 768w\" sizes=\"auto, (max-width: 800px) 100vw, 800px\" \/><\/figure>\n<\/div>\n\n\n<p class=\"wp-block-paragraph\">If you have never seen or tried changing conditions by selecting the <strong>Conditions<\/strong>&nbsp;tab, I encourage you to do so.&nbsp;To learn more about Pro EZGC, visit the&nbsp;Help Section&nbsp;and the&nbsp;Restek Video Library. Let us know if you have any questions.<\/p>\n\n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>When you log into the ProEZGC Chromatogram Modeler software, you will see the screen below, which is what I refer to as the Welcome Screen. This is the starting point for all, and unfortunately all some customers will think there&#8230;<\/p>\n","protected":false},"author":2,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[9],"tags":[],"industries-application":[],"post-badge":[],"resource-type":[],"product-library":[],"resource-technique":[2294],"ppma_author":[582],"class_list":["post-51777","post","type-post","status-publish","format-standard","hentry","category-blogs","resource-technique-gas-chromatography-gc"],"acf":[],"taxonomy_info":{"category":[{"value":9,"label":"Blogs"}],"resource-technique":[{"value":2294,"label":"Gas Chromatography (GC)"}]},"featured_image_src_large":false,"author_info":{"display_name":"Patrick Gallagher","author_link":"https:\/\/discover.restek.com\/zh-hans\/author\/patrick-gallagherrestek-com\/"},"comment_info":0,"category_info":[{"term_id":9,"name":"Blogs","slug":"blogs","term_group":0,"term_taxonomy_id":9,"taxonomy":"category","description":"","parent":0,"count":457,"filter":"raw","cat_ID":9,"category_count":457,"category_description":"","cat_name":"Blogs","category_nicename":"blogs","category_parent":0}],"tag_info":false,"authors":[{"term_id":582,"user_id":0,"is_guest":1,"slug":"alan-sensue","display_name":"Alan Sensue","avatar_url":"https:\/\/secure.gravatar.com\/avatar\/?s=96&d=mm&r=g","author_category":"1","first_name":"Alan","last_name":"Sensue","user_url":"","job_title":"","description":""}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts\/51777","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/users\/2"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/comments?post=51777"}],"version-history":[{"count":3,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts\/51777\/revisions"}],"predecessor-version":[{"id":85434,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts\/51777\/revisions\/85434"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/media?parent=51777"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/categories?post=51777"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/tags?post=51777"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/industries-application?post=51777"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/post-badge?post=51777"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/resource-type?post=51777"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/product-library?post=51777"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/resource-technique?post=51777"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/ppma_author?post=51777"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}