{"id":51859,"date":"2014-10-08T14:30:00","date_gmt":"2014-10-08T14:30:00","guid":{"rendered":"https:\/\/discover.restek.com\/uncategorized\/which-gc-injection-port-liner-to-use-for-gas-samples\/"},"modified":"2026-09-23T20:57:21","modified_gmt":"2026-09-23T20:57:21","slug":"which-gc-injection-port-liner-to-use-for-gas-samples","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/zh-hans\/blogs\/gnbl4868\/which-gc-injection-port-liner-to-use-for-gas-samples","title":{"rendered":"Which GC injection port liner to use for gas samples"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">The answer seems fairly straightforward: Use a narrow internal diameter (ID) liner to keep peak shapes sharp by preventing band broadening; but this isn\u2019t always the correct answer. There has to be enough internal volume in the liner to contain all of the sample.&nbsp; Let me explain.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Let\u2019s first look at liners when using&nbsp;<strong>purge &amp; trap (P&amp;T)<\/strong>. All instrumentation that I am familiar with uses a transfer line from the P&amp;T unit to the GC injection port.&nbsp; In this case, you do want to use a narrow ID liner to minimize dead volume and prevent band broadening, especially for the compounds which will not condense and refocus within the capillary column.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Now let\u2019s look at liners when using&nbsp;<strong>headspace (HS)<\/strong>. Headspace units can transfer the sample to the injection port two ways:<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">(1) Using a Gas-Tight syringe&nbsp;to fill a&nbsp;sample loop&nbsp;(which is coupled to a transfer line).<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">(2) Using a Gas-Tight syringe and injecting directly into an injection port liner.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">If the headspace unit uses a&nbsp;<strong>transfer line<\/strong>, then a <strong>small ID<\/strong> liner should be used. However, if a <strong>Gas-Tight syringe<\/strong> is used to inject the gas sample directly into an injection port liner, then a <strong>large ID<\/strong> liner is commonly used.&nbsp;<\/p>\n\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<p class=\"wp-block-paragraph\">&nbsp;(<strong>Splitless<\/strong>&nbsp;injections)<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">The liner below is a Restek Premium 4.0mm ID Single Taper Inlet Liner (23302), which is typically the best choice for gas injections of approximately &gt; 250\u00b5L to 1mL.<\/p>\n\n\n<div class=\"wp-block-image\">\n<figure class=\"aligncenter\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-which-gc-injection-port-liner-to-use-for-gas-samples-01.gif\" alt=\"SG4mmID\" class=\"wp-image-13287\" title=\"-\"><\/figure>\n<\/div>\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<p class=\"wp-block-paragraph\">&nbsp;The liner below is a Restek Premium 2.0mm ID Single Taper Inlet Liner (23315), which is typically the best choice for gas injections less than approximately 250\u00b5L.<\/p>\n\n\n<div class=\"wp-block-image\">\n<figure class=\"aligncenter\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-which-gc-injection-port-liner-to-use-for-gas-samples-02.gif\" alt=\"SG2mmID\" class=\"wp-image-13286\" title=\"-\"><\/figure>\n<\/div>\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<p class=\"wp-block-paragraph\">&nbsp;(<strong>Split<\/strong>&nbsp;injections)<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">If you are injecting gases into a liner using a Gas-Tight syringe and splitting your sample, choose an open-bottom liner (with no gooseneck or other restriction at the bottom), like 23301 (photo below). Restek Premium 4.0 mm ID Straight Inlet Liner<\/p>\n\n\n<div class=\"wp-block-image\">\n<figure class=\"aligncenter\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-which-gc-injection-port-liner-to-use-for-gas-samples-03.gif\" alt=\"a blue rectangle with white text\" class=\"wp-image-13339\" title=\"-\"><\/figure>\n<\/div>\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<p class=\"wp-block-paragraph\">&nbsp;If you are not using a syringe for sample introduction but instead the sample is from a purge &amp; trap unit, headspace unit which uses a transfer line, or gas sampling valve, then the liner I usually recommend is 23333 (photo below).&nbsp;&nbsp;Restek Premium 1.0mm ID Straight Inlet Liner<\/p>\n\n\n<div class=\"wp-block-image\">\n<figure class=\"aligncenter\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-which-gc-injection-port-liner-to-use-for-gas-samples-04.gif\" alt=\"1mmID\" class=\"wp-image-13288\" title=\"-\"><\/figure>\n<\/div>\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<p class=\"wp-block-paragraph\">&nbsp;<em><strong>Please note<\/strong><\/em>&nbsp;that these are representative liners for an Agilent GC with a split\/splitless injection port; the catalog number of the liner you would select depends upon your&nbsp;<em>specific instrument and injection port<\/em>.&nbsp;&nbsp;<em><strong>In addition<\/strong><\/em>, make sure the injection port liner internal diameter (ID) is large enough for a transfer line to fit into the liner (if applicable).&nbsp; For example, I have heard that some transfer lines have an outside diameter (OD) that is larger than 1mm OD; in this case, you would need to use a liner with a 2mm ID or larger opening, and not the 1mm ID liner shown above.&nbsp; As an example, I recently found out that the Agilent 7697 Headspace Sampler requires a 2mmID liner like Restek&nbsp;23313.<\/p>\n\n\n<div class=\"wp-block-image\">\n<figure class=\"aligncenter\"><img decoding=\"async\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-which-gc-injection-port-liner-to-use-for-gas-samples-05.gif\" alt=\"decorative\" class=\"wp-image-79657\" title=\"-\"><\/figure>\n<\/div>\n\n\n<div style=\"height:20px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<p class=\"wp-block-paragraph\">&nbsp;In summary, liner selection for gas samples will depend on your particular instrumentation and how the sample is introduced into the injection port liner. If your instrument uses a transfer line, then a small ID liner will likely provide you the best chromatography.&nbsp; If a Gas-Tight syringe is used to inject the gas sample, a larger internal diameter liner will usually provide you the best results.<\/p>\n\n\n\n<div style=\"height:50px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n\n<div class=\"wp-block-columns are-vertically-aligned-center has-border-color has-background is-layout-flex wp-container-core-columns-is-layout-8f761849 wp-block-columns-is-layout-flex\" style=\"border-color:#d8e5f0;border-radius:10px;background-color:#d8e5f0\">\n<div class=\"wp-block-column is-vertically-aligned-center is-layout-flow wp-block-column-is-layout-flow\" style=\"flex-basis:66.66%\">\n<h3 class=\"wp-block-heading has-text-color has-link-color wp-elements-1\" style=\"color:#0f6da0\">Restek&#8217;s Topaz GC Inlet Liners<\/h3>\n\n\n\n<p class=\"wp-block-paragraph\">Restek\u2019s Topaz GC inlet liners feature revolutionary technology and inertness to deliver the next level of True Blue Performance:<\/p>\n\n\n\n<ul style=\"margin-bottom:20px\" class=\"wp-block-list\">\n<li><strong>Deactivation<\/strong>\u2014unbelievably low breakdown for accurate and precise low-level GC analyses.<\/li>\n\n\n\n<li><strong>Reproducibility<\/strong>\u2014unbeatable manufacturing controls and QC testing for superior reliability across compound classes.<\/li>\n\n\n\n<li><strong>Productivity<\/strong>\u2014unparalleled cleanliness for maximized GC uptime and lab throughput.<\/li>\n<\/ul>\n\n\n<style>.wp-block-kadence-advancedbtn.kb-btns104379_7af9ae-b4{gap:var(--global-kb-gap-md, 2rem);justify-content:flex-start;align-items:center;}.kt-btns104379_7af9ae-b4 .kt-button{font-weight:normal;font-style:normal;}.kt-btns104379_7af9ae-b4 .kt-btn-wrap-0{margin-right:5px;}.wp-block-kadence-advancedbtn.kt-btns104379_7af9ae-b4 .kt-btn-wrap-0 .kt-button{color:#555555;border-color:#555555;}.wp-block-kadence-advancedbtn.kt-btns104379_7af9ae-b4 .kt-btn-wrap-0 .kt-button:hover, .wp-block-kadence-advancedbtn.kt-btns104379_7af9ae-b4 .kt-btn-wrap-0 .kt-button:focus{color:#ffffff;border-color:#444444;}.wp-block-kadence-advancedbtn.kt-btns104379_7af9ae-b4 .kt-btn-wrap-0 .kt-button::before{display:none;}.wp-block-kadence-advancedbtn.kt-btns104379_7af9ae-b4 .kt-btn-wrap-0 .kt-button:hover, .wp-block-kadence-advancedbtn.kt-btns104379_7af9ae-b4 .kt-btn-wrap-0 .kt-button:focus{background:#444444;}<\/style>\n<div class=\"wp-block-kadence-advancedbtn kb-buttons-wrap kb-btns104379_7af9ae-b4\"><style>ul.menu .wp-block-kadence-advancedbtn .kb-btn104379_ba1339-09.kb-button{width:initial;}.wp-block-kadence-advancedbtn .kb-btn104379_ba1339-09.kb-button{background:#008cd4;border-top-left-radius:10px;border-top-right-radius:10px;border-bottom-right-radius:10px;border-bottom-left-radius:10px;padding:var(--kb-btn-padding);margin:var(--kb-btn-margin);box-shadow:none;}.wp-block-kadence-advancedbtn .kb-btn104379_ba1339-09.kb-button:hover, .wp-block-kadence-advancedbtn .kb-btn104379_ba1339-09.kb-button:focus{background:var(--global-palette2, #2B6CB0);box-shadow:var(--kb-btn-shadow-hover, none);}<\/style><a class=\"kb-button kt-button button kb-btn104379_ba1339-09 kt-btn-size-standard kt-btn-width-type-auto kb-btn-global-fill kt-btn-has-text-true kt-btn-has-svg-false wp-block-kadence-singlebtn\" href=\"https:\/\/discover.restek.com\/topaz-liners\/\"><span class=\"kt-btn-inner-text\"><strong>LEARN MORE<\/strong><\/span><\/a>\n\n<style>ul.menu .wp-block-kadence-advancedbtn .kb-btn104379_614e75-66.kb-button{width:initial;}.wp-block-kadence-advancedbtn .kb-btn104379_614e75-66.kb-button{background:#008cd4;border-top-left-radius:10px;border-top-right-radius:10px;border-bottom-right-radius:10px;border-bottom-left-radius:10px;padding:var(--kb-btn-padding);margin:var(--kb-btn-margin);box-shadow:none;}.wp-block-kadence-advancedbtn .kb-btn104379_614e75-66.kb-button:hover, .wp-block-kadence-advancedbtn .kb-btn104379_614e75-66.kb-button:focus{background:var(--global-palette2, #2B6CB0);box-shadow:var(--kb-btn-shadow-hover, none);}<\/style><a class=\"kb-button kt-button button kb-btn104379_614e75-66 kt-btn-size-standard kt-btn-width-type-auto kb-btn-global-fill kt-btn-has-text-true kt-btn-has-svg-false wp-block-kadence-singlebtn\" href=\"https:\/\/www.restek.com\/c\/1484\" target=\"_blank\" rel=\"noopener\"><span class=\"kt-btn-inner-text\"><strong>SHOP LINERS<\/strong><\/span><\/a><\/div>\n<\/div>\n\n\n\n<div class=\"wp-block-column is-vertically-aligned-center is-layout-flow wp-block-column-is-layout-flow\" style=\"flex-basis:33.33%\"><div class=\"wp-block-image\">\n<figure class=\"aligncenter size-large is-resized\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"799\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/Topaz-1024x799.png\" alt=\"\" class=\"wp-image-29812\" style=\"width:170px;height:auto\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/Topaz-1024x799.png 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/Topaz-300x234.png 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/Topaz-768x599.png 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/Topaz-1536x1198.png 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/Topaz.png 1632w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure>\n<\/div><\/div>\n<\/div>\n\n\n\n<div style=\"height:50px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n        <div class=\"cpb\">\n            <h3 class=\"cpb-heading\">\u76f8\u5173\u4ea7\u54c1<\/h3>\n            <hr class=\"cpb-heading-underline\" \/>\n            <div class=\"cpb-list\">\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23302\">                                \u8d27\u53f7\uff1a 23302                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Topaz, Single Taper Inlet Liner, 4.0 mm x 6.5 x 78.5, for Agilent GCs, Premium Deactivation, 5-pk.<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23302\">\u67e5\u770b\u4ea7\u54c1<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23315\">                                \u8d27\u53f7\uff1a 23315                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Topaz, Single Taper Inlet Liner, 2.0 mm x 6.5 x 78.5, for Agilent GCs, Premium Deactivation, 5-pk.<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23315\">\u67e5\u770b\u4ea7\u54c1<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23301\">                                \u8d27\u53f7\uff1a 23301                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Topaz, Straight Inlet Liner, 4.0 mm x 6.3 x 78.5, for Agilent GCs, Premium Deactivation, 5-pk.<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23301\">\u67e5\u770b\u4ea7\u54c1<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23333\">                                \u8d27\u53f7\uff1a 23333                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Topaz, Straight Inlet Liner, 1.0 mm x 6.3 x 78.5, for Agilent GCs, Premium Deactivation, 5-pk.<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23333\">\u67e5\u770b\u4ea7\u54c1<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23313\">                                \u8d27\u53f7\uff1a 23313                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Topaz, Straight Inlet Liner, 2.0 mm x 6.5 x 78.5, for Agilent GCs, Premium Deactivation, 5-pk.<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/23313\">\u67e5\u770b\u4ea7\u54c1<\/a>\n                                                    <\/div>\n                    <\/div>\n                            <\/div>\n        <\/div>\n        \n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>The answer seems fairly straightforward: Use a narrow internal diameter (ID) liner to keep peak shapes sharp by preventing band broadening; but this isn&rsquo;t always the correct answer. There has to be enough internal volume in the liner to contain&#8230;<\/p>\n","protected":false},"author":14,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[9],"tags":[2794],"industries-application":[],"post-badge":[],"resource-type":[567],"product-library":[],"resource-technique":[],"ppma_author":[582],"class_list":["post-51859","post","type-post","status-publish","format-standard","hentry","category-blogs","tag-topaz-liners","resource-type-selection-guides"],"acf":[],"taxonomy_info":{"category":[{"value":9,"label":"Blogs"}],"post_tag":[{"value":2794,"label":"Topaz Liners"}],"resource-type":[{"value":567,"label":"Selection Guides"}]},"featured_image_src_large":false,"author_info":{"display_name":"Daniel","author_link":"https:\/\/discover.restek.com\/zh-hans\/author\/daniel-voralia\/"},"comment_info":0,"category_info":[{"term_id":9,"name":"Blogs","slug":"blogs","term_group":0,"term_taxonomy_id":9,"taxonomy":"category","description":"","parent":0,"count":457,"filter":"raw","cat_ID":9,"category_count":457,"category_description":"","cat_name":"Blogs","category_nicename":"blogs","category_parent":0}],"tag_info":[{"term_id":2794,"name":"Topaz Liners","slug":"topaz-liners","term_group":0,"term_taxonomy_id":2794,"taxonomy":"post_tag","description":"Explore Topaz Liner Resources","parent":0,"count":28,"filter":"raw"}],"authors":[{"term_id":582,"user_id":0,"is_guest":1,"slug":"alan-sensue","display_name":"Alan Sensue","avatar_url":"https:\/\/secure.gravatar.com\/avatar\/?s=96&d=mm&r=g","author_category":"1","first_name":"Alan","last_name":"Sensue","user_url":"","job_title":"","description":""}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts\/51859","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/users\/14"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/comments?post=51859"}],"version-history":[{"count":3,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts\/51859\/revisions"}],"predecessor-version":[{"id":104401,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts\/51859\/revisions\/104401"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/media?parent=51859"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/categories?post=51859"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/tags?post=51859"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/industries-application?post=51859"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/post-badge?post=51859"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/resource-type?post=51859"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/product-library?post=51859"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/resource-technique?post=51859"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/ppma_author?post=51859"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}