{"id":75751,"date":"2024-03-26T20:46:00","date_gmt":"2024-03-26T20:46:00","guid":{"rendered":"https:\/\/discover.restek.com\/?p=75751"},"modified":"2026-01-08T14:25:39","modified_gmt":"2026-01-08T14:25:39","slug":"column-selections-for-the-renewable-fuel-standard-rfs-program-biogas-analysis","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/zh-hans\/blogs\/column-selections-for-the-renewable-fuel-standard-rfs-program-biogas-analysis\/","title":{"rendered":"Column selection(s) for the Renewable Fuel Standard (RFS) Program \u2013 Biogas analysis"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">The RFS program seeks to increase energy and fuel production within the USA, while also incentivizing reductions and control of pollutants including sulfurs, siloxanes, greenhouse gases, and an assortment of other toxic compounds<sup>1<\/sup>. In turn, these compounds need to be monitored consistently and reliably, which can be difficult in biogas matrices. Biogas is produced from sites such as landfills, sewage treatment plants, pig farms, and cattle farms, where high amounts of methane are produced and can be repurposed for energy, removing emissions from the atmosphere. These sites are notoriously difficult to sample since substrate materials vary dramatically, and can be moist, promoting side-reactions during the sampling and analysis processes. For this reason, a variety of equipment is often necessary to test different compound classes.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">To help labs get started and standardize their methods, the RFS Program recommends a variety of regulatory methods related to energy production that can be used to analyze different compounds classes, including ASTM D5504, ASTM D8230, EPA 3C, and EPA 18<sup>1,2<\/sup>, though many labs also include TO-15<sup>3<\/sup>&nbsp;or TO-17. While not specified by the RFS Program, TO-15 and TO-17 are used to monitor toxic gases which are also likely to be present in biogas and may be hazardous to the environment rather than to energy production.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">With such a wide range of sample types possible for this program, the specific columns dimensions and selections are likely to vary from industry to industry, but this blog will help identify \u201cstarting points\u201d in column-selection for different compound classes.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">ASTM D5504-20<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Sulfur-containing volatiles compounds are hazards to both the environment and combustion equipment. Hydrogen sulfide, specifically, is a very reactive compound which can readily decompose to H<sub>2<\/sub>SO<sub>4<\/sub>, poisoning catalysts or corroding biogas engine and turbines<sup>4<\/sup>. Other potential degradation byproducts such as sulfur dioxide and trioxide are environmentally hazardous<sup>4<\/sup>. This compound is monitored primarily to determine the cost-benefit of using the biogas, where companies can determine how much wear and tear there may be on the equipment, prior to burning.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">ASTM D5504-20 recommends a 60m x 0.53mmID PDMS phase column<sup>5<\/sup>, such as Restek\u2019s Rtx-1, though shorter, narrower ID columns have been demonstrated to work as well, with slightly lower sample capacity. The ASTM method shows an example of retention times possible using a 30m x 0.32mmID x 4\u00b5m equivalent to Restek\u2019s Rtx-1 (cat.# 10198), suggesting it is a good starting point. Below is an example method showing separation of some sulfur compounds on the\u00a0<a href=\"https:\/\/www.restek.com\/chromatogram-detail\/GC_PC1271\" target=\"_blank\" rel=\"noopener\">Rtx-1<\/a>:<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\">Figure 0: Sulfur Compounds on Rtx-1<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_pc1271.png\" alt=\"Sulfur Compounds on Rtx-1\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_PC1271<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-6 col-12 peak-50\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 50px\">Conc.<br \/>(\u00b5g\/mL)<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethyl mercaptan\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-08-1\/Ethyl mercaptan\" target=\"_blank\" rel=\"noopener\">Ethyl mercaptan<\/a><\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dimethyl sulfide\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-18-3\/Dimethyl sulfide\" target=\"_blank\" rel=\"noopener\">Dimethyl sulfide<\/a><\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Carbon disulfide\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-15-0\/Carbon disulfide\" target=\"_blank\" rel=\"noopener\">Carbon disulfide<\/a><\/td><td class=\"oth\">20<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Isopropyl mercaptan\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-33-2\/Isopropyl mercaptan\" target=\"_blank\" rel=\"noopener\">Isopropyl mercaptan<\/a><\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for t-Butyl mercaptan\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-66-1\/t-Butyl mercaptan\" target=\"_blank\" rel=\"noopener\">t-Butyl mercaptan<\/a><\/td><td class=\"oth\">50<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for n-Propyl mercaptan\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/107-03-9\/n-Propyl mercaptan\" target=\"_blank\" rel=\"noopener\">n-Propyl mercaptan<\/a><\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethyl methyl sulfide\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/624-89-5\/Ethyl methyl sulfide\" target=\"_blank\" rel=\"noopener\">Ethyl methyl sulfide<\/a><\/td><td class=\"oth\">50<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for sec-Butyl mercaptan\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/513-53-1\/sec-Butyl mercaptan\" target=\"_blank\" rel=\"noopener\">sec-Butyl mercaptan<\/a><\/td><td class=\"oth\">50<\/td><\/tr>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Thiophene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/110-02-1\/Thiophene\" target=\"_blank\" rel=\"noopener\">Thiophene<\/a><\/td><td class=\"oth\">100<\/td><\/tr>\n<\/tbody><\/table>\n<table class=\"peaks col-lg-6 col-12 peak-50\"><thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 50px\">Conc.<br \/>(\u00b5g\/mL)<\/th><\/tr><\/thead><tbody>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Isobutyl mercaptan\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/513-44-0\/Isobutyl mercaptan\" target=\"_blank\" rel=\"noopener\">Isobutyl mercaptan<\/a><\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">11.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Diethyl sulfide\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/352-93-2\/Diethyl sulfide\" target=\"_blank\" rel=\"noopener\">Diethyl sulfide<\/a><\/td><td class=\"oth\">75<\/td><\/tr>\n<tr><td class=\"num\">12.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for n-Butyl mercaptan\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/109-79-5\/n-Butyl mercaptan\" target=\"_blank\" rel=\"noopener\">n-Butyl mercaptan<\/a><\/td><td class=\"oth\">50<\/td><\/tr>\n<tr><td class=\"num\">13.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dimethyl disulfide\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/624-92-0\/Dimethyl disulfide\" target=\"_blank\" rel=\"noopener\">Dimethyl disulfide<\/a><\/td><td class=\"oth\">20<\/td><\/tr>\n<tr><td class=\"num\">14.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Methylthiophene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/554-14-3\/2-Methylthiophene\" target=\"_blank\" rel=\"noopener\">2-Methylthiophene<\/a><\/td><td class=\"oth\">75<\/td><\/tr>\n<tr><td class=\"num\">15.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 3-Methylthiophene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/616-44-4\/3-Methylthiophene\" target=\"_blank\" rel=\"noopener\">3-Methylthiophene<\/a><\/td><td class=\"oth\">100<\/td><\/tr>\n<tr><td class=\"num\">16.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Diethyl disulfide\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/110-81-6\/Diethyl disulfide\" target=\"_blank\" rel=\"noopener\">Diethyl disulfide<\/a><\/td><td class=\"oth\">20<\/td><\/tr>\n<tr><td class=\"num\">17.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzothiophene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/95-15-8\/Benzothiophene\" target=\"_blank\" rel=\"noopener\">Benzothiophene<\/a><\/td><td class=\"oth\">75<\/td><\/tr>\n<tr><td class=\"num\">18.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 3-Methylbenzothiophene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1455-18-1\/3-Methylbenzothiophene\" target=\"_blank\" rel=\"noopener\">3-Methylbenzothiophene<\/a><\/td><td class=\"oth\">100<\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Rtx-1, 30 m, 0.32 mm ID, 4.00 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/10198\" rel=\"noopener\">cat.# 10198<\/a>)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td>ASTM D5623 standards 1 &amp; 2 from DCG Partnership 1, Ltd.<\/td><\/tr>\n<tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>0.2 \u00b5L split (split ratio 10:1)<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Liner:<\/th><td>4 mm ID cup inlet liner with Siltek deactivation<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Temp.:<\/th><td>275 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Split Vent Flow Rate:<\/th><td>24 mL\/min<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>30 &#176;C (hold 2 min) to 200 &#176;C at 10 &#176;C\/min (hold 10 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Flow Rate:<\/th><td>2.4 mL\/min<tr><th class=\"sub conditions_header\" scope=\"row\">Linear Velocity:<\/th><td>38 cm\/sec @ 30 \u00b0C<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>DP SCD @ 800 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Burner Vacuum:<\/th><td>363 torr<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Hydrogen Flow Rate:<\/th><td>40 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Air Flow Rate:<\/th><td>60 mL\/min<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>Agilent 7890B GC<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Notes<\/th><td>Solvent standard of sulfur components found in gasoline samples utilizing ASTM  D5623.  <br \/><\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>10198<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>ASTM D5623<\/li><li>D5623<\/li><li>astm-gc-library<\/li><li>gasoline<\/li><li>scd<\/li><li>sulfur<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_pc1271.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<h2 class=\"wp-block-heading\">ASTM D8230-19<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Silicon-containing compounds, primarily polysiloxanes, are present primarily landfills and sewage, from sources such as lubricants and hygienic products<sup>6<\/sup>. While volatile methyl siloxanes have the potential to be converted to methane for biogas production, polysiloxanes are more likely to be monitored for their potential to damage equipment<sup>4,6<\/sup>. Polysiloxanes, when burnt, leave behind abrasive silica dioxide (SiO2) which abrades and blocks equipment. In addition, some polysiloxanes such as D4 (octamethylcyclotetrasiloxane) and D5 (decamethylcyclopentasiloxane) are known to be hazardous to the environment<sup>6<\/sup>.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">ASTM D8230-19 provides GC-MS, GC-ICP-MS, and GC-AED methodologies for the analysis of silicon-containing volatiles<sup>7<\/sup>. While there is no specific column phase recommendation, the method specifies a low bleed column is necessary, either composed of diphenyl dimethyl polysiloxane<sup>7<\/sup>&nbsp;(i.e. Rtx-5) or 1,4- bis(dimethylsiloxy)phenylene di-methyl polysiloxane<sup>7<\/sup>&nbsp;(Rxi-5Sil MS). Column bleed is composed of polysiloxanes which can of course interfere with analysis of polysiloxanes, so column inertness is a top priority.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">However, the need for a thick film will work against another critical need \u2013 low bleed. A balance therefore must be struck in column dimensions to keep the film thick enough for the necessary retention of volatiles, but thin enough to control bleed. Clark\u00a0<em>et al<\/em>., interested in polysiloxanes D4 and D5 in landfill gas, used a 30m x 0.32mmID x 1\u00b5m column<sup>6<\/sup>, equivalent to the Rxi-5Sil MS (cat.#13654) With Restek\u2019s exceptionally inert Rxi deactivation, this column is a great starting point for your polysiloxane analysis. As an alternative, the\u00a0<a href=\"https:\/\/www.restek.com\/chromatogram-detail\/GC_CH00360\" target=\"_blank\" rel=\"noopener\">Rtx-200<\/a>\u00a0(cat.# 15020) is also well suited to separating polysiloxanes with a unique chemistry that may benefit separation from other components of a mixture.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">EPA 3C<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Permanent gas analysis is important for understanding the combustion efficiency of biogas. While biogas is primarily methane, there is likely a high presence of other non-combustible gases such as CO2, CO, N2, and O2 which can interfere with energy production efficiency. This is especially important if catalysts are involved for conversation of non-combustible gases to more combustible gases (usually methane), where the presence of certain contaminants may affect catalyst efficiency too.<sup>4<\/sup><\/p>\n\n\n\n<p class=\"wp-block-paragraph\">EPA method 3C does not recommend a specific column, but encourages separation of methane, CO2, N2, O2, and \u201cother gas components that may be present in the sample\u201d<sup>8<\/sup>. Molecular sieves are prone to adsorption of water and CO2, which is why Restek\u2019s ShinCarbon ST column (cat.# 19808) may be a better option. Below is an example of the separation of permanent gases achievable on the\u00a0<a href=\"https:\/\/www.restek.com\/chromatogram-detail\/GC_PC00665\" target=\"_blank\" rel=\"noopener\">ShinCarbon ST<\/a>.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\">Figure 1: Permanent Gases and C1+C2 Hydrocarbons on ShinCarbon ST (micropacked)<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image regular-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_pc00665.png\" alt=\"Permanent Gases and C1+C2 Hydrocarbons on ShinCarbon ST (micropacked)\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_PC00665<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-4 col-12\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Oxygen\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/7782-44-7\/Oxygen\" target=\"_blank\" rel=\"noopener\">Oxygen<\/a><\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Nitrogen\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/7727-37-9\/Nitrogen\" target=\"_blank\" rel=\"noopener\">Nitrogen<\/a><\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Carbon monoxide\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/630-08-0\/Carbon monoxide\" target=\"_blank\" rel=\"noopener\">Carbon monoxide<\/a><\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/74-82-8\/Methane\" target=\"_blank\" rel=\"noopener\">Methane<\/a><\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Carbon dioxide\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/124-38-9\/Carbon dioxide\" target=\"_blank\" rel=\"noopener\">Carbon dioxide<\/a><\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetylene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/74-86-2\/Acetylene\" target=\"_blank\" rel=\"noopener\">Acetylene<\/a><\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethylene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/74-85-1\/Ethylene\" target=\"_blank\" rel=\"noopener\">Ethylene<\/a><\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/74-84-0\/Ethane\" target=\"_blank\" rel=\"noopener\">Ethane<\/a><\/td><\/tr>\n<\/tbody><\/table><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>100\/120 mesh, ShinCarbon ST, 2 m, 1\/16 in. OD, 1.0 mm ID (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/19808\" rel=\"noopener\">cat.# 19808<\/a>)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td>Permanent gases-C1+C2 hydrocarbons<\/td><\/tr>\n<tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Vol.:<\/th><td>5 \u00b5L packed on-column<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Inj. Temp.:<\/th><td>100 \u00b0C<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>40 &#176;C (hold 3 min) to 250 &#176;C at 8 &#176;C\/min (hold 10 min)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Flow Rate:<\/th><td>10 mL\/min<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>HID @ 200 \u00b0C<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>19808<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>packed column<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_pc00665.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\">Since CO and CO2 do not burn very well, the addition of a methanizer&nbsp;may be helpful to improve detection limits relative to a TCD.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">EPA 18, TO-15 &amp; TO-17<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">While considered \u201cminor impurities\u201d in biogas, there is a wide variety volatile organic compounds (VOC\u2019s) that are produced in biogas sites, many of which are hazardous to the environment, equipment, or both. The VOC\u2019s are produced either from waste itself (ex: solvents) or from the anaerobic breakdown of waste. Since the waste substrate is highly variable, the compounds produced are also highly variable, which means analytical methods need to be flexible as well. It is well established that biogas sites contain hazardous solvents such benzene and vinyl chloride, alongside more benign malodorous compounds like hydrocarbons and oxygenates<sup>4,12<\/sup>. To control odor, chlorinated products may be added to landfills, contributing to the generation of hazardous chlorinated halocarbons as well<sup>4,13<\/sup>.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Because of the overlap with toxic compounds, EPA Methods 18, TO-15, or TO-17 can be performed. Method 18 is not very specific about compounds or analytical methods (intentionally)<sup>9<\/sup>, while TO-15 and TO-17 provide more details for analysts to work with<sup>10,11<\/sup>. However, TO-15 and TO-17 do not include all compounds that may be hazardous to biogas equipment, such as terpenes<sup>3,12<\/sup>&nbsp;so methods and equipment may need to be augmented for this wider range of compounds.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Restek has many options for separating TO-15 compounds, but with the addition of more polar terpenes, the\u00a0<a href=\"https:\/\/www.restek.com\/chromatogram-detail\/GC_AR1148\" target=\"_blank\" rel=\"noopener\">Rxi-624Sil MS<\/a>\u00a0is likely the best starting point. Authors Orecchio and Marti both used 60m 624-type columns, such as\u00a0cat.# 13869\u00a0and\u00a0cat.# 13872<sup>3,13<\/sup>.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\">Figure 2: TO-15 65 Component Mix on Rxi-624Sil MS (30 m)<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<div class=\"wp-block-custom-chromatogram-article\"><div class=\"wp-block-custom-chromatogram-article\"><div class=\"chromatogram-image wide-image\"><img decoding=\"async\" src=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ar1148.png\" alt=\"TO-15 65 Component Mix on Rxi-624Sil MS (30 m)\" title=\"-\"><\/div><p class=\"article-id\" style=\"text-align:center\"> GC_AR1148<\/p><div class=\"chromatogram-peaks\"><h4>Peaks<\/h4><table class=\"peaks col-lg-4 col-12 peak-33\">\n<thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 50px\">t<sub>R<\/sub> (min)<\/th><\/tr><\/thead>\n<tbody><tr><td class=\"num\">1.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Propylene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/115-07-1\/Propylene\" target=\"_blank\" rel=\"noopener\">Propylene<\/a><\/td><td class=\"oth\">1.42<\/td><\/tr>\n<tr><td class=\"num\">2.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dichlorodifluoromethane (CFC-12)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-71-8\/Dichlorodifluoromethane\" target=\"_blank\" rel=\"noopener\">Dichlorodifluoromethane (CFC-12)<\/a><\/td><td class=\"oth\">1.46<\/td><\/tr>\n<tr><td class=\"num\">3.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,2-Dichlorotetrafluoroethane (CFC-114)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/76-14-2\/1,2-Dichlorotetrafluoroethane\" target=\"_blank\" rel=\"noopener\">1,2-Dichlorotetrafluoroethane (CFC-114)<\/a><\/td><td class=\"oth\">1.58<\/td><\/tr>\n<tr><td class=\"num\">4.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Chloromethane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/74-87-3\/Chloromethane\" target=\"_blank\" rel=\"noopener\">Chloromethane<\/a><\/td><td class=\"oth\">1.62<\/td><\/tr>\n<tr><td class=\"num\">5.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Vinyl chloride\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-01-4\/Vinyl chloride\" target=\"_blank\" rel=\"noopener\">Vinyl chloride<\/a><\/td><td class=\"oth\">1.73<\/td><\/tr>\n<tr><td class=\"num\">6.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,3-Butadiene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/106-99-0\/1,3-Butadiene\" target=\"_blank\" rel=\"noopener\">1,3-Butadiene<\/a><\/td><td class=\"oth\">1.77<\/td><\/tr>\n<tr><td class=\"num\">7.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Bromomethane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/74-83-9\/Bromomethane\" target=\"_blank\" rel=\"noopener\">Bromomethane<\/a><\/td><td class=\"oth\">2.02<\/td><\/tr>\n<tr><td class=\"num\">8.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Chloroethane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-00-3\/Chloroethane\" target=\"_blank\" rel=\"noopener\">Chloroethane<\/a><\/td><td class=\"oth\">2.12<\/td><\/tr>\n<tr><td class=\"num\">9.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Trichlorofluoromethane (CFC-11)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-69-4\/Trichlorofluoromethane\" target=\"_blank\" rel=\"noopener\">Trichlorofluoromethane (CFC-11)<\/a><\/td><td class=\"oth\">2.34<\/td><\/tr>\n<tr><td class=\"num\">10.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethanol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/64-17-5\/Ethanol\" target=\"_blank\" rel=\"noopener\">Ethanol<\/a><\/td><td class=\"oth\">2.54<\/td><\/tr>\n<tr><td class=\"num\">11.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acrolein\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/107-02-8\/Acrolein\" target=\"_blank\" rel=\"noopener\">Acrolein<\/a><\/td><td class=\"oth\">2.71<\/td><\/tr>\n<tr><td class=\"num\">12.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,1-Dichloroethene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-35-4\/1,1-Dichloroethene\" target=\"_blank\" rel=\"noopener\">1,1-Dichloroethene<\/a><\/td><td class=\"oth\">2.78<\/td><\/tr>\n<tr><td class=\"num\">13.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,1,2-Trichlorotrifluoroethane (CFC-113)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/76-13-1\/1,1,2-Trichlorotrifluoroethane\" target=\"_blank\" rel=\"noopener\">1,1,2-Trichlorotrifluoroethane (CFC-113)<\/a><\/td><td class=\"oth\">2.80<\/td><\/tr>\n<tr><td class=\"num\">14.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetone\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-64-1\/Acetone\" target=\"_blank\" rel=\"noopener\">Acetone<\/a><\/td><td class=\"oth\">2.82<\/td><\/tr>\n<tr><td class=\"num\">15.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Carbon disulfide\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-15-0\/Carbon disulfide\" target=\"_blank\" rel=\"noopener\">Carbon disulfide<\/a><\/td><td class=\"oth\">2.95<\/td><\/tr>\n<tr><td class=\"num\">16.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Isopropyl alcohol\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-63-0\/Isopropyl alcohol\" target=\"_blank\" rel=\"noopener\">Isopropyl alcohol<\/a><\/td><td class=\"oth\">2.99<\/td><\/tr>\n<tr><td class=\"num\">17.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Acetonitrile (contaminant)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-05-8\/Acetonitrile\" target=\"_blank\" rel=\"noopener\">Acetonitrile (contaminant)<\/a><\/td><td class=\"oth\">3.05<\/td><\/tr>\n<tr><td class=\"num\">18.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methylene chloride\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-09-2\/Methylene chloride\" target=\"_blank\" rel=\"noopener\">Methylene chloride<\/a><\/td><td class=\"oth\">3.20<\/td><\/tr>\n<tr><td class=\"num\">19.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for trans-1,2-Dichloroethene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/156-60-5\/trans-1,2-Dichloroethene\" target=\"_blank\" rel=\"noopener\">trans-1,2-Dichloroethene<\/a><\/td><td class=\"oth\">3.43<\/td><\/tr>\n<tr><td class=\"num\">20.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methyl tert-butyl ether (MTBE)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/1634-04-4\/Methyl tert-butyl ether\" target=\"_blank\" rel=\"noopener\">Methyl tert-butyl ether (MTBE)<\/a><\/td><td class=\"oth\">3.44<\/td><\/tr>\n<tr><td class=\"num\">21.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Hexane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/110-54-3\/Hexane\" target=\"_blank\" rel=\"noopener\">Hexane<\/a><\/td><td class=\"oth\">3.67<\/td><\/tr>\n<tr><td class=\"num\">22.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,1-Dichloroethane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-34-3\/1,1-Dichloroethane\" target=\"_blank\" rel=\"noopener\">1,1-Dichloroethane<\/a><\/td><td class=\"oth\">3.80<\/td><\/tr>\n<tr><td class=\"num\">23.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Vinyl acetate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/108-05-4\/Vinyl acetate\" target=\"_blank\" rel=\"noopener\">Vinyl acetate<\/a><\/td><td class=\"oth\">3.82<\/td><\/tr>\n<\/tbody><\/table>\n<table class=\"peaks col-lg-4 col-12 peak-33\"><thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 50px\">t<sub>R<\/sub> (min)<\/th><\/tr><\/thead><tbody>\n<tr><td class=\"num\">24.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Butanone (MEK)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/78-93-3\/2-Butanone\" target=\"_blank\" rel=\"noopener\">2-Butanone (MEK)<\/a><\/td><td class=\"oth\">4.29<\/td><\/tr>\n<tr><td class=\"num\">25.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for cis-1,2-Dichloroethene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/156-59-2\/cis-1,2-Dichloroethene\" target=\"_blank\" rel=\"noopener\">cis-1,2-Dichloroethene<\/a><\/td><td class=\"oth\">4.29<\/td><\/tr>\n<tr><td class=\"num\">26.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethyl acetate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/141-78-6\/Ethyl acetate\" target=\"_blank\" rel=\"noopener\">Ethyl acetate<\/a><\/td><td class=\"oth\">4.34<\/td><\/tr>\n<tr><td class=\"num\">27.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Bromochloromethane (IS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/74-97-5\/Bromochloromethane\" target=\"_blank\" rel=\"noopener\">Bromochloromethane (IS)<\/a><\/td><td class=\"oth\">4.50<\/td><\/tr>\n<tr><td class=\"num\">28.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Tetrahydrofuran\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/109-99-9\/Tetrahydrofuran\" target=\"_blank\" rel=\"noopener\">Tetrahydrofuran<\/a><\/td><td class=\"oth\">4.52<\/td><\/tr>\n<tr><td class=\"num\">29.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Chloroform\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/67-66-3\/Chloroform\" target=\"_blank\" rel=\"noopener\">Chloroform<\/a><\/td><td class=\"oth\">4.60<\/td><\/tr>\n<tr><td class=\"num\">30.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,1,1-Trichloroethane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/71-55-6\/1,1,1-Trichloroethane\" target=\"_blank\" rel=\"noopener\">1,1,1-Trichloroethane<\/a><\/td><td class=\"oth\">4.74<\/td><\/tr>\n<tr><td class=\"num\">31.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Cyclohexane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/110-82-7\/Cyclohexane\" target=\"_blank\" rel=\"noopener\">Cyclohexane<\/a><\/td><td class=\"oth\">4.80<\/td><\/tr>\n<tr><td class=\"num\">32.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Carbon tetrachloride\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/56-23-5\/Carbon tetrachloride\" target=\"_blank\" rel=\"noopener\">Carbon tetrachloride<\/a><\/td><td class=\"oth\">4.88<\/td><\/tr>\n<tr><td class=\"num\">33.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/71-43-2\/Benzene\" target=\"_blank\" rel=\"noopener\">Benzene<\/a><\/td><td class=\"oth\">5.06<\/td><\/tr>\n<tr><td class=\"num\">34.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,2-Dichloroethane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/107-06-2\/1,2-Dichloroethane\" target=\"_blank\" rel=\"noopener\">1,2-Dichloroethane<\/a><\/td><td class=\"oth\">5.12<\/td><\/tr>\n<tr><td class=\"num\">35.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Heptane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/142-82-5\/Heptane\" target=\"_blank\" rel=\"noopener\">Heptane<\/a><\/td><td class=\"oth\">5.35<\/td><\/tr>\n<tr><td class=\"num\">36.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,4-Difluorobenzene (IS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/540-36-3\/1,4-Difluorobenzene\" target=\"_blank\" rel=\"noopener\">1,4-Difluorobenzene (IS)<\/a><\/td><td class=\"oth\">5.47<\/td><\/tr>\n<tr><td class=\"num\">37.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Trichloroethylene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/79-01-6\/Trichloroethylene\" target=\"_blank\" rel=\"noopener\">Trichloroethylene<\/a><\/td><td class=\"oth\">5.69<\/td><\/tr>\n<tr><td class=\"num\">38.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,2-Dichloropropane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/78-87-5\/1,2-Dichloropropane\" target=\"_blank\" rel=\"noopener\">1,2-Dichloropropane<\/a><\/td><td class=\"oth\">5.94<\/td><\/tr>\n<tr><td class=\"num\">39.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Methyl methacrylate\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/80-62-6\/Methyl methacrylate\" target=\"_blank\" rel=\"noopener\">Methyl methacrylate<\/a><\/td><td class=\"oth\">6.01<\/td><\/tr>\n<tr><td class=\"num\">40.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,4-Dioxane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/123-91-1\/1,4-Dioxane\" target=\"_blank\" rel=\"noopener\">1,4-Dioxane<\/a><\/td><td class=\"oth\">6.02<\/td><\/tr>\n<tr><td class=\"num\">41.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Bromodichloromethane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-27-4\/Bromodichloromethane\" target=\"_blank\" rel=\"noopener\">Bromodichloromethane<\/a><\/td><td class=\"oth\">6.20<\/td><\/tr>\n<tr><td class=\"num\">42.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for cis-1,3-Dichloropropene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/10061-01-5\/cis-1,3-Dichloropropene\" target=\"_blank\" rel=\"noopener\">cis-1,3-Dichloropropene<\/a><\/td><td class=\"oth\">6.64<\/td><\/tr>\n<tr><td class=\"num\">43.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 4-Methyl-2-pentanone (MIBK)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/108-10-1\/4-Methyl-2-pentanone\" target=\"_blank\" rel=\"noopener\">4-Methyl-2-pentanone (MIBK)<\/a><\/td><td class=\"oth\">6.80<\/td><\/tr>\n<tr><td class=\"num\">44.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Toluene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/108-88-3\/Toluene\" target=\"_blank\" rel=\"noopener\">Toluene<\/a><\/td><td class=\"oth\">6.99<\/td><\/tr>\n<tr><td class=\"num\">45.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for trans-1,3-Dichloropropene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/10061-02-6\/trans-1,3-Dichloropropene\" target=\"_blank\" rel=\"noopener\">trans-1,3-Dichloropropene<\/a><\/td><td class=\"oth\">7.26<\/td><\/tr>\n<tr><td class=\"num\">46.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,1,2-Trichloroethane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/79-00-5\/1,1,2-Trichloroethane\" target=\"_blank\" rel=\"noopener\">1,1,2-Trichloroethane<\/a><\/td><td class=\"oth\">7.47<\/td><\/tr>\n<\/tbody><\/table>\n<table class=\"peaks col-lg-4 col-12 peak-33\"><thead><tr><th><\/th><th style=\"text-align: left;width: 75px\">Peaks<\/th><th style=\"text-align: center;width: 50px\">t<sub>R<\/sub> (min)<\/th><\/tr><\/thead><tbody>\n<tr><td class=\"num\">47.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Tetrachloroethene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/127-18-4\/Tetrachloroethene\" target=\"_blank\" rel=\"noopener\">Tetrachloroethene<\/a><\/td><td class=\"oth\">7.55<\/td><\/tr>\n<tr><td class=\"num\">48.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 2-Hexanone (MBK)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/591-78-6\/2-Hexanone\" target=\"_blank\" rel=\"noopener\">2-Hexanone (MBK)<\/a><\/td><td class=\"oth\">7.72<\/td><\/tr>\n<tr><td class=\"num\">49.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Dibromochloromethane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/124-48-1\/Dibromochloromethane\" target=\"_blank\" rel=\"noopener\">Dibromochloromethane<\/a><\/td><td class=\"oth\">7.88<\/td><\/tr>\n<tr><td class=\"num\">50.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,2-Dibromoethane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/106-93-4\/1,2-Dibromoethane\" target=\"_blank\" rel=\"noopener\">1,2-Dibromoethane<\/a><\/td><td class=\"oth\">8.01<\/td><\/tr>\n<tr><td class=\"num\">51.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Chlorobenzene-d5 (IS)\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/3114-55-4\/Chlorobenzene-d5\" target=\"_blank\" rel=\"noopener\">Chlorobenzene-d5 (IS)<\/a><\/td><td class=\"oth\">8.54<\/td><\/tr>\n<tr><td class=\"num\">52.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Chlorobenzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/108-90-7\/Chlorobenzene\" target=\"_blank\" rel=\"noopener\">Chlorobenzene<\/a><\/td><td class=\"oth\">8.58<\/td><\/tr>\n<tr><td class=\"num\">53.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Ethylbenzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/100-41-4\/Ethylbenzene\" target=\"_blank\" rel=\"noopener\">Ethylbenzene<\/a><\/td><td class=\"oth\">8.71<\/td><\/tr>\n<tr><td class=\"num\">54.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for m-Xylene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/108-38-3\/m-Xylene\" target=\"_blank\" rel=\"noopener\">m-Xylene<\/a><\/td><td class=\"oth\">8.87<\/td><\/tr>\n<tr><td class=\"num\">55.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for p-Xylene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/106-42-3\/p-Xylene\" target=\"_blank\" rel=\"noopener\">p-Xylene<\/a><\/td><td class=\"oth\">8.87<\/td><\/tr>\n<tr><td class=\"num\">56.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for o-Xylene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/95-47-6\/o-Xylene\" target=\"_blank\" rel=\"noopener\">o-Xylene<\/a><\/td><td class=\"oth\">9.32<\/td><\/tr>\n<tr><td class=\"num\">57.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Styrene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/100-42-5\/Styrene\" target=\"_blank\" rel=\"noopener\">Styrene<\/a><\/td><td class=\"oth\">9.35<\/td><\/tr>\n<tr><td class=\"num\">58.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Bromoform\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/75-25-2\/Bromoform\" target=\"_blank\" rel=\"noopener\">Bromoform<\/a><\/td><td class=\"oth\">9.56<\/td><\/tr>\n<tr><td class=\"num\">59.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 4-Bromofluorobenzene*\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/460-00-4\/4-Bromofluorobenzene\" target=\"_blank\" rel=\"noopener\">4-Bromofluorobenzene*<\/a><\/td><td class=\"oth\">9.99<\/td><\/tr>\n<tr><td class=\"num\">60.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,1,2,2-Tetrachloroethane\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/79-34-5\/1,1,2,2-Tetrachloroethane\" target=\"_blank\" rel=\"noopener\">1,1,2,2-Tetrachloroethane<\/a><\/td><td class=\"oth\">10.18<\/td><\/tr>\n<tr><td class=\"num\">61.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 4-Ethyltoluene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/622-96-8\/4-Ethyltoluene\" target=\"_blank\" rel=\"noopener\">4-Ethyltoluene<\/a><\/td><td class=\"oth\">10.47<\/td><\/tr>\n<tr><td class=\"num\">62.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,3,5-Trimethylbenzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/108-67-8\/1,3,5-Trimethylbenzene\" target=\"_blank\" rel=\"noopener\">1,3,5-Trimethylbenzene<\/a><\/td><td class=\"oth\">10.55<\/td><\/tr>\n<tr><td class=\"num\">63.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,2,4-Trimethylbenzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/95-63-6\/1,2,4-Trimethylbenzene\" target=\"_blank\" rel=\"noopener\">1,2,4-Trimethylbenzene<\/a><\/td><td class=\"oth\">11.02<\/td><\/tr>\n<tr><td class=\"num\">64.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,3-Dichlorobenzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/541-73-1\/1,3-Dichlorobenzene\" target=\"_blank\" rel=\"noopener\">1,3-Dichlorobenzene<\/a><\/td><td class=\"oth\">11.35<\/td><\/tr>\n<tr><td class=\"num\">65.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,4-Dichlorobenzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/106-46-7\/1,4-Dichlorobenzene\" target=\"_blank\" rel=\"noopener\">1,4-Dichlorobenzene<\/a><\/td><td class=\"oth\">11.49<\/td><\/tr>\n<tr><td class=\"num\">66.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Benzyl chloride\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/100-44-7\/Benzyl chloride\" target=\"_blank\" rel=\"noopener\">Benzyl chloride<\/a><\/td><td class=\"oth\">11.65<\/td><\/tr>\n<tr><td class=\"num\">67.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,2-Dichlorobenzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/95-50-1\/1,2-Dichlorobenzene\" target=\"_blank\" rel=\"noopener\">1,2-Dichlorobenzene<\/a><\/td><td class=\"oth\">11.93<\/td><\/tr>\n<tr><td class=\"num\">68.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for 1,2,4-Trichlorobenzene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/120-82-1\/1,2,4-Trichlorobenzene\" target=\"_blank\" rel=\"noopener\">1,2,4-Trichlorobenzene<\/a><\/td><td class=\"oth\">13.47<\/td><\/tr>\n<tr><td class=\"num\">69.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Hexachlorobutadiene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/87-68-3\/Hexachlorobutadiene\" target=\"_blank\" rel=\"noopener\">Hexachlorobutadiene<\/a><\/td><td class=\"oth\">13.59<\/td><\/tr>\n<tr><td class=\"num\">70.<\/td><td class=\"cmpd\"><a class=\"cmpd_link\" title=\"View compound information for Naphthalene\" href=\"https:\/\/ez.restek.com\/compound\/view\/en\/91-20-3\/Naphthalene\" target=\"_blank\" rel=\"noopener\">Naphthalene<\/a><\/td><td class=\"oth\">13.67<\/td><\/tr>\n<\/tbody><\/table><div class=\"peaknotes\">*Tuning standard<\/div><\/div><div class=\"chromatogram-conditions\"><h4>Conditions<\/h4><div class=\"conditions-container container-fluid\"><div class=\"row\"><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Column<\/th><td>Rxi-624Sil MS, 30 m, 0.32 mm ID, 1.80 \u00b5m (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/13870\" rel=\"noopener\">cat.# 13870<\/a>)<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Standard\/Sample<\/th><td><\/td><\/tr><tr><td><\/td><td>TO-15 65 component mix (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/34436\" rel=\"noopener\">cat.# 34436<\/a>)<\/td><\/tr><tr><td><\/td><td>TO-14A internal standard\/tuning mix (<a class=\"ga-cgram-prod-link\" target=\"_blank\" href=\"https:\/\/www.restek.com\/p\/34408\" rel=\"noopener\">cat.# 34408<\/a>)<\/td><\/tr>\n<tr><th class=\"sub conditions_header\" scope=\"row\">Diluent:<\/th><td>Nitrogen<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Conc.:<\/th><td>10.0  ppbv 200 cc injection<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Injection<\/th><td>Direct<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\" colspan=\"2\">Oven<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Oven Temp.:<\/th><td>32 &#176;C (hold 1 min) to 150 &#176;C at 11 &#176;C\/min to 230 &#176;C at 33 &#176;C\/min<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Carrier Gas<\/th><td>He, constant flow<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Flow Rate:<\/th><td>2.0 mL\/min<tr><th class=\"sub conditions_header\" scope=\"row\">Linear Velocity:<\/th><td>51 cm\/sec @ 32 \u00b0C<\/td><\/tr><\/table><table class=\"conditions col-lg-6 col-12\"><tr><th class=\"conditions_header\" scope=\"row\">Detector<\/th><td>MS<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Mode:<\/th><td>Scan<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Transfer Line Temp.:<\/th><td>230 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Analyzer Type:<\/th><td>Quadrupole <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Source Temp.:<\/th><td>230 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Quad Temp.:<\/th><td>150 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Electron Energy:<\/th><td>69.9 eV<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Solvent Delay Time:<\/th><td>1.0  min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Tune Type:<\/th><td>BFB <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Ionization Mode:<\/th><td>EI <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Scan Range:<\/th><td>35 &#8211; 250 amu<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">Scan Rate:<\/th><td>3.32 scans\/sec<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Preconcentrator<\/th><td>Nutech 8900DS<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">&nbsp;&nbsp;Trap 1 Settings<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">&nbsp;&nbsp;Type\/Sorbent :<\/th><td>Glass Beads <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">&nbsp;&nbsp;Cooling temp:<\/th><td>-155 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">&nbsp;&nbsp;Preheat temp:<\/th><td>5 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">&nbsp;&nbsp;Preheat time:<\/th><td>0 sec<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">&nbsp;&nbsp;Desorb temp:<\/th><td>20 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">&nbsp;&nbsp;Desorb flow:<\/th><td>5 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">&nbsp;&nbsp;Desorb time:<\/th><td>360 sec<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">&nbsp;&nbsp;Bakeout temp:<\/th><td>200 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">&nbsp;&nbsp;Flush flow:<\/th><td>120 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">&nbsp;&nbsp;Flush time:<\/th><td>60 sec<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">&nbsp;&nbsp;Sweep flow:<\/th><td>120 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">&nbsp;&nbsp;Sweep time:<\/th><td>60 sec<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">&nbsp;&nbsp;Trap 2 Settings<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">&nbsp;&nbsp;Type\/Sorbent:<\/th><td>Tenax <\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">&nbsp;&nbsp;Cooling temp:<\/th><td>-35 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">&nbsp;&nbsp;Desorb temp:<\/th><td>190 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">&nbsp;&nbsp;Desorb time:<\/th><td>30 sec<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">&nbsp;&nbsp;Bakeout temp:<\/th><td>200 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">&nbsp;&nbsp;Bakeout time:<\/th><td>10 sec<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">&nbsp;&nbsp;Cryofocuser<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">&nbsp;&nbsp;Cooling temp:<\/th><td>-160 \u00b0C<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">&nbsp;&nbsp;Inject time:<\/th><td>140 sec<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">&nbsp;&nbsp;Internal Standard<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">&nbsp;&nbsp;Purge flow:<\/th><td>100 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">&nbsp;&nbsp;Purge time:<\/th><td>6 sec<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">&nbsp;&nbsp;Vol.:<\/th><td>20 mL<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">&nbsp;&nbsp;ISTD flow:<\/th><td>100 mL\/min<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">&nbsp;&nbsp;Standard<\/th><td><\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">&nbsp;&nbsp;Size:<\/th><td>200 mL<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">&nbsp;&nbsp;Purge flow:<\/th><td>100 mL\/min<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">&nbsp;&nbsp;Purge time:<\/th><td>6 sec<\/td><\/tr><tr><th class=\"sub conditions_header\" scope=\"row\">&nbsp;&nbsp;Sample flow:<\/th><td>100 mL\/min<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Instrument<\/th><td>HP6890  GC &amp; 5973 MSD<\/td><\/tr><tr class=\"cgram_header_row\"><th class=\"conditions_header\" scope=\"row\">Acknowledgement<\/th><td>Nutech<\/td><\/tr><\/table><\/div><\/div><\/div><div class=\"chromatogram-related-skus\"><h4>Related SKUs<\/h4><ul><li>34436<\/li><li>34408<\/li><li>13870<\/li><\/ul><\/div><div class=\"chromatogram-related-searches\"><h4>Related Searches<\/h4><ul><li>Air<\/li><li>TO-15<\/li><\/ul><\/div><div class=\"chromatogram-pdf-link\"><a href=\"https:\/\/ez.restek.com\/images\/cgram\/gc_ar1148.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"><svg xmlns=\"http:\/\/www.w3.org\/2000\/svg\" width=\"18\" height=\"18\" viewBox=\"0 0 18 18\"><g data-name=\"Group 2996\"><path data-name=\"Rectangle 1246\" d=\"M0 0h18v18H0z\" style=\"fill: none;\"><\/path><\/g><g data-name=\"Group 2997\"><path data-name=\"Path 729\" d=\"M13.412 11.4v2.017H5.345V11.4H4v2.017a1.349 1.349 0 0 0 1.345 1.345h8.068a1.349 1.349 0 0 0 1.345-1.345V11.4zm-.672-2.694-.948-.948-1.741 1.735V4H8.706v5.493L6.965 7.758l-.948.948 3.361 3.361z\" transform=\"translate(-.437 -.414)\" style=\"fill: rgb(13, 123, 196);\"><\/path><\/g><\/svg>Download PDF<\/a><\/div><\/div><\/div>\n<\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\">Alternative columns that may be suitable, depending on the specific compounds being investigated, include&nbsp;<a href=\"https:\/\/www.restek.com\/chromatogram-detail\/GC_AR1156\" target=\"_blank\" rel=\"noopener\">Rxi-1ms<\/a>,&nbsp;<a href=\"https:\/\/www.restek.com\/chromatogram-detail\/GC_AR1157\" target=\"_blank\" rel=\"noopener\">Rxi-5Sil MS<\/a>, and&nbsp;<a href=\"https:\/\/www.restek.com\/chromatogram-detail\/GC_AR1153\" target=\"_blank\" rel=\"noopener\">Rtx-VMS<\/a>. To help with the selection, a variety of compounds that may be of interest are included in Restek\u2019s&nbsp;<a href=\"https:\/\/ez.restek.com\/proezgc\" target=\"_blank\" rel=\"noopener\">Pro<em>EZ<\/em>GC<\/a>&nbsp;libraries. Column phases, dimensions, and oven parameters can be adjusted to determine which column is best for your specific VOC analysis.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Additional resources:<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Alan Sensue has additional recommendations for sampling equipment related to your biogas analysis:\u00a0<a href=\"https:\/\/discover.restek.com\/gnbl4785\/sampling-and-analysis-of-landfill-gas-biogas-an-overview\/\" data-type=\"post\" data-id=\"51783\">Sampling and analysis of landfill gas biogas an overview (restek.com)<\/a><\/p>\n\n\n\n<h2 class=\"wp-block-heading\">References:<\/h2>\n\n\n\n<ol class=\"wp-block-list\">\n<li>Renewable fuel standards Program:\u00a0<a href=\"https:\/\/www.epa.gov\/renewable-fuel-standard-program\" target=\"_blank\" rel=\"noreferrer noopener\">Renewable Fuel Standard Program | US EPA<\/a><\/li>\n\n\n\n<li>40 CFR Parts 80 and 1090 (pg 44535):\u00a0<a href=\"https:\/\/www.govinfo.gov\/content\/pkg\/FR-2023-07-12\/pdf\/2023-13462.pdf\" target=\"_blank\" rel=\"noreferrer noopener\">https:\/\/www.govinfo.gov\/content\/pkg\/FR-2023-07-12\/pdf\/2023-13462.pdf<\/a><\/li>\n\n\n\n<li>Orecchio\u00a0<em>et al<\/em>.:\u00a0<a href=\"https:\/\/doi.org\/10.3390\/ijerph14020195\" target=\"_blank\" rel=\"noreferrer noopener\">https:\/\/doi.org\/10.3390\/ijerph14020195<\/a><\/li>\n\n\n\n<li>Braganca\u00a0<em>et al<\/em>.:\u00a0<a href=\"https:\/\/doi.org\/10.1016\/j.biombioe.2020.105878\" target=\"_blank\" rel=\"noreferrer noopener\">https:\/\/doi.org\/10.1016\/j.biombioe.2020.105878<\/a><\/li>\n\n\n\n<li>ASTM D5504:\u00a0<a href=\"https:\/\/www.astm.org\/d5504-20.html\" target=\"_blank\" rel=\"noreferrer noopener\">D5504 Standard Test Method for Determination of Sulfur Compounds in Natural Gas and Gaseous Fuels by Gas Chromatography and Chemiluminescence (astm.org)<\/a><\/li>\n\n\n\n<li>Clark\u00a0<em>et al<\/em>.:\u00a0<a href=\"https:\/\/doi.org\/10.1139\/l2012-051\" target=\"_blank\" rel=\"noreferrer noopener\">https:\/\/doi.org\/10.1139\/l2012-051<\/a><\/li>\n\n\n\n<li>ASTM D8230:\u00a0<a href=\"https:\/\/www.astm.org\/d8230-19.html\" target=\"_blank\" rel=\"noreferrer noopener\">D8230 Standard Test Method for Measurement of Volatile Silicon-Containing Compounds in a Gaseous Fuel Sample Using Gas Chromatography with Spectroscopic Detection (astm.org)<\/a><\/li>\n\n\n\n<li>EPA 3C:\u00a0<a href=\"https:\/\/www.epa.gov\/sites\/default\/files\/2017-08\/documents\/method_3c.pdf\" target=\"_blank\" rel=\"noreferrer noopener\">METHOD 3C\u2014DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES (epa.gov)<\/a><\/li>\n\n\n\n<li>EPA 18:\u00a0<a href=\"https:\/\/www.epa.gov\/sites\/default\/files\/2019-06\/documents\/method_18_0.pdf\" target=\"_blank\" rel=\"noreferrer noopener\">method_18_0.pdf (epa.gov)<\/a><\/li>\n\n\n\n<li>EPA TO-15:\u00a0<a href=\"https:\/\/www.epa.gov\/sites\/default\/files\/2019-12\/documents\/to-15a_vocs.pdf\" target=\"_blank\" rel=\"noreferrer noopener\">https:\/\/www.epa.gov\/sites\/default\/files\/2019-12\/documents\/to-15a_vocs.pdf<\/a><\/li>\n\n\n\n<li>EPA TO-17:\u00a0<a href=\"https:\/\/www.epa.gov\/sites\/default\/files\/2019-11\/documents\/to-17r.pdf\" target=\"_blank\" rel=\"noreferrer noopener\">Method TO-17 &#8211; Determination of Volatile Organic Compounds in Ambient Air Using Active Sampling Onto Sorbent Tubes (epa.gov)<\/a><\/li>\n\n\n\n<li>Duan\u00a0<em>et al<\/em>.:\u00a0<a href=\"https:\/\/doi.org\/10.1016\/j.wasman.2020.09.015\" target=\"_blank\" rel=\"noreferrer noopener\">https:\/\/doi.org\/10.1016\/j.wasman.2020.09.015<\/a><\/li>\n\n\n\n<li>Marti\u00a0<em>et al<\/em>.:\u00a0<a href=\"https:\/\/doi.org\/10.1016\/j.apgeochem.2014.06.018\" target=\"_blank\" rel=\"noreferrer noopener\">https:\/\/doi.org\/10.1016\/j.apgeochem.2014.06.018<\/a><\/li>\n<\/ol>\n\n\n\n<div style=\"height:30px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n\n\n        <div class=\"cpb\">\n            <h3 class=\"cpb-heading\">Related Products<\/h3>\n            <hr class=\"cpb-heading-underline\" \/>\n            <div class=\"cpb-list\">\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/10198\">                                \u8d27\u53f7\uff1a 10198                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Rtx-1 GC Capillary Column, 30 m, 0.32 mm ID, 4.00 \u00b5m<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/10198\">\u67e5\u770b\u4ea7\u54c1<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/13654\">                                \u8d27\u53f7\uff1a 13654                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Rxi-5Sil MS GC Capillary Column, 30 m, 0.32 mm ID, 1.0 \u00b5m<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/13654\">\u67e5\u770b\u4ea7\u54c1<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/19808\">                                \u8d27\u53f7\uff1a 19808                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">ShinCarbon ST Micropacked GC Column, 100\/120, 2 m, 1 mm ID, 1\/16&quot; OD, Silco<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/19808\">\u67e5\u770b\u4ea7\u54c1<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/15020\">                                \u8d27\u53f7\uff1a 15020                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Rtx-200 GC Capillary Column, 15 m, 0.25 mm ID, 0.25 \u00b5m<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/15020\">\u67e5\u770b\u4ea7\u54c1<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/13869\">                                \u8d27\u53f7\uff1a 13869                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Rxi-624Sil MS GC Capillary Column, 60 m, 0.25 mm ID, 1.40 \u00b5m<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/13869\">\u67e5\u770b\u4ea7\u54c1<\/a>\n                                                    <\/div>\n                    <\/div>\n                                    <div class=\"cpb-item\">\n                        <div class=\"cpb-col cpb-col--left\">\n                            <a class=\"cpb-catalog\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/13872\">                                \u8d27\u53f7\uff1a 13872                            <\/a>                        <\/div>\n                        <div class=\"cpb-col cpb-col--middle\">\n                            <div class=\"cpb-title\">Rxi-624Sil MS GC Capillary Column, 60 m, 0.32 mm ID, 1.80 \u00b5m<\/div>\n                        <\/div>\n                        <div class=\"cpb-col cpb-col--right\">\n                                                            <a class=\"cpb-view-btn\" target=\"_blank\" rel=\"noopener noreferrer\" href=\"https:\/\/www.restek.com\/p\/13872\">\u67e5\u770b\u4ea7\u54c1<\/a>\n                                                    <\/div>\n                    <\/div>\n                            <\/div>\n        <\/div>\n        \n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>The RFS program seeks to increase energy and fuel production within the USA, while also incentivizing reductions and control of pollutants including sulfurs, siloxanes, greenhouse gases, and an assortment of other toxic compounds1. In turn, these compounds need to be monitored consistently and reliably, which can be difficult in biogas matrices. Biogas is produced from&#8230;<\/p>\n","protected":false},"author":34,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[9],"tags":[],"industries-application":[],"post-badge":[],"resource-type":[],"product-library":[],"resource-technique":[],"ppma_author":[454],"class_list":["post-75751","post","type-post","status-publish","format-standard","hentry","category-blogs"],"acf":[],"taxonomy_info":{"category":[{"value":9,"label":"Blogs"}]},"featured_image_src_large":false,"author_info":{"display_name":"Erica Pack, PhD","author_link":"https:\/\/discover.restek.com\/zh-hans\/author\/erica-pack-phd\/"},"comment_info":0,"category_info":[{"term_id":9,"name":"Blogs","slug":"blogs","term_group":0,"term_taxonomy_id":9,"taxonomy":"category","description":"","parent":0,"count":457,"filter":"raw","cat_ID":9,"category_count":457,"category_description":"","cat_name":"Blogs","category_nicename":"blogs","category_parent":0}],"tag_info":false,"authors":[{"term_id":454,"user_id":34,"is_guest":0,"slug":"erica-pack-phd","display_name":"Erica Pack, PhD","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-pack-erica.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-pack-erica.jpg"},"author_category":"1","first_name":"Erica","last_name":"Pack","user_url":"https:\/\/www.linkedin.com\/in\/ericapack\/","job_title":"Senior Scientist, GC Applied","description":"Erica is an interdisciplinary GC applications and technologies scientist at Restek. She obtained her bachelor's degree in forensic biology from The Pennsylvania State University, and her doctorate from Virginia Tech in plant pathology, physiology, and weed science. Since joining Restek in 2021, she has worked with a wide variety of GC columns, including fused silica, MXT, PLOT, and packed columns as well as accessories, such as liners, valves, and methanizers."}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts\/75751","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/users\/34"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/comments?post=75751"}],"version-history":[{"count":3,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts\/75751\/revisions"}],"predecessor-version":[{"id":75755,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts\/75751\/revisions\/75755"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/media?parent=75751"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/categories?post=75751"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/tags?post=75751"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/industries-application?post=75751"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/post-badge?post=75751"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/resource-type?post=75751"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/product-library?post=75751"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/resource-technique?post=75751"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/ppma_author?post=75751"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}