{"id":75756,"date":"2024-01-23T20:58:00","date_gmt":"2024-01-23T20:58:00","guid":{"rendered":"https:\/\/discover.restek.com\/?p=75756"},"modified":"2026-01-08T14:25:44","modified_gmt":"2026-01-08T14:25:44","slug":"compound-optimization-in-lc-ms-ms-why-does-it-matter","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/zh-hans\/blogs\/compound-optimization-in-lc-ms-ms-why-does-it-matter\/","title":{"rendered":"Compound optimization in LC-MS\/MS\u2014Why does it matter?"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">Compound optimization, also referred to as tuning, is the process in which mass spectrometry parameters such as the precursor\/product ions, collision energies, and other voltages are optimized for each analyte. Compound optimization can often be overlooked. It might seem easier to find an article or application note and just use the compound settings for a specific analyte, but this shortcut might have consequences for the analyte performance. All mass spectrometers differ from one another; even ones of the same make and model. Because of this variability, transition parameters that work for one instrument may not be optimal for another instrument. This can result in a loss of sensitivity and may impact the detection limits for a particular compound.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">In the example below, three compounds were assessed to see how using optimized compound settings impacted analyte response. Settings for cocaine, morphine, and \u03949-THC were optimized on a Shimadzu LCMS-8045. Another set of settings for these analytes were obtained from an LC-MS\/MS method featured in a journal article<sup>1<\/sup>. The optimized and literature settings are listed below.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Table 1.<\/strong>&nbsp;Analyte settings optimized in-lab on Shimadzu LCMS-8045.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td class=\"has-text-align-center\" data-align=\"center\"><strong>Analyte<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Ion Mode<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Parent<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Product<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Q1 Pre Bias (V)<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>CE<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Q3 Pre Bias (V)<\/strong><\/td><\/tr><tr><td class=\"has-text-align-center\" data-align=\"center\" rowspan=\"2\">Cocaine<\/td><td class=\"has-text-align-center\" data-align=\"center\" rowspan=\"2\">ESI+<\/td><td class=\"has-text-align-center\" data-align=\"center\" rowspan=\"2\">303.90<\/td><td class=\"has-text-align-center\" data-align=\"center\">182.10<\/td><td class=\"has-text-align-center\" data-align=\"center\">-16<\/td><td class=\"has-text-align-center\" data-align=\"center\">-20<\/td><td class=\"has-text-align-center\" data-align=\"center\">-19<\/td><\/tr><tr><td class=\"has-text-align-center\" data-align=\"center\">82.15<\/td><td class=\"has-text-align-center\" data-align=\"center\">-16<\/td><td class=\"has-text-align-center\" data-align=\"center\">-30<\/td><td class=\"has-text-align-center\" data-align=\"center\">-16<\/td><\/tr><tr><td class=\"has-text-align-center\" data-align=\"center\" rowspan=\"2\">Morphine<\/td><td class=\"has-text-align-center\" data-align=\"center\" rowspan=\"2\">ESI+<\/td><td class=\"has-text-align-center\" data-align=\"center\" rowspan=\"2\">285.90<\/td><td class=\"has-text-align-center\" data-align=\"center\">201.05<\/td><td class=\"has-text-align-center\" data-align=\"center\">-15<\/td><td class=\"has-text-align-center\" data-align=\"center\">-27<\/td><td class=\"has-text-align-center\" data-align=\"center\">-21<\/td><\/tr><tr><td class=\"has-text-align-center\" data-align=\"center\">152.15<\/td><td class=\"has-text-align-center\" data-align=\"center\">-11<\/td><td class=\"has-text-align-center\" data-align=\"center\">-55<\/td><td class=\"has-text-align-center\" data-align=\"center\">-28<\/td><\/tr><tr><td class=\"has-text-align-center\" data-align=\"center\" rowspan=\"2\">\u03949-THC<\/td><td class=\"has-text-align-center\" data-align=\"center\" rowspan=\"2\">ESI+<\/td><td class=\"has-text-align-center\" data-align=\"center\" rowspan=\"2\">315.10<\/td><td class=\"has-text-align-center\" data-align=\"center\">193.05<\/td><td class=\"has-text-align-center\" data-align=\"center\">-23<\/td><td class=\"has-text-align-center\" data-align=\"center\">-24<\/td><td class=\"has-text-align-center\" data-align=\"center\">-20<\/td><\/tr><tr><td class=\"has-text-align-center\" data-align=\"center\">123.25<\/td><td class=\"has-text-align-center\" data-align=\"center\">-16<\/td><td class=\"has-text-align-center\" data-align=\"center\">-35<\/td><td class=\"has-text-align-center\" data-align=\"center\">-23<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Table 2.<\/strong>&nbsp;Analyte settings obtained from journal article (not optimized in-lab)<sup>1<\/sup>.<\/p>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td class=\"has-text-align-center\" data-align=\"center\"><strong>Analyte<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Ion Mode<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Parent<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Product<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Q1 Pre Bias (V)<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>CE<\/strong><\/td><td class=\"has-text-align-center\" data-align=\"center\"><strong>Q3 Pre Bias (V)<\/strong><\/td><\/tr><tr><td class=\"has-text-align-center\" data-align=\"center\" rowspan=\"2\">Cocaine<\/td><td class=\"has-text-align-center\" data-align=\"center\" rowspan=\"2\">ESI+<\/td><td class=\"has-text-align-center\" data-align=\"center\" rowspan=\"2\">304.00<\/td><td class=\"has-text-align-center\" data-align=\"center\">182.00<\/td><td class=\"has-text-align-center\" data-align=\"center\">-20<\/td><td class=\"has-text-align-center\" data-align=\"center\">-18<\/td><td class=\"has-text-align-center\" data-align=\"center\">-18<\/td><\/tr><tr><td class=\"has-text-align-center\" data-align=\"center\">82.00<\/td><td class=\"has-text-align-center\" data-align=\"center\">-20<\/td><td class=\"has-text-align-center\" data-align=\"center\">-31<\/td><td class=\"has-text-align-center\" data-align=\"center\">-14<\/td><\/tr><tr><td class=\"has-text-align-center\" data-align=\"center\" rowspan=\"2\">Morphine<\/td><td class=\"has-text-align-center\" data-align=\"center\" rowspan=\"2\">ESI+<\/td><td class=\"has-text-align-center\" data-align=\"center\" rowspan=\"2\">286.20<\/td><td class=\"has-text-align-center\" data-align=\"center\">152.20<\/td><td class=\"has-text-align-center\" data-align=\"center\">-20<\/td><td class=\"has-text-align-center\" data-align=\"center\">-55<\/td><td class=\"has-text-align-center\" data-align=\"center\">-30<\/td><\/tr><tr><td class=\"has-text-align-center\" data-align=\"center\">201.20<\/td><td class=\"has-text-align-center\" data-align=\"center\">-20<\/td><td class=\"has-text-align-center\" data-align=\"center\">-25<\/td><td class=\"has-text-align-center\" data-align=\"center\">-12<\/td><\/tr><tr><td class=\"has-text-align-center\" data-align=\"center\" rowspan=\"2\">\u03949-THC<\/td><td class=\"has-text-align-center\" data-align=\"center\" rowspan=\"2\">ESI+<\/td><td class=\"has-text-align-center\" data-align=\"center\" rowspan=\"2\">315.00<\/td><td class=\"has-text-align-center\" data-align=\"center\">193.30<\/td><td class=\"has-text-align-center\" data-align=\"center\">-16<\/td><td class=\"has-text-align-center\" data-align=\"center\">-23<\/td><td class=\"has-text-align-center\" data-align=\"center\">-19<\/td><\/tr><tr><td class=\"has-text-align-center\" data-align=\"center\">123.10<\/td><td class=\"has-text-align-center\" data-align=\"center\">-16<\/td><td class=\"has-text-align-center\" data-align=\"center\">-34<\/td><td class=\"has-text-align-center\" data-align=\"center\">-12<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">A sample containing the analytes at a concentration of 100 ng\/mL was run on two methods, one using the optimized and one using the literature settings. All other method parameters were the same for both methods. The peak heights and areas were compared for each analyte using the optimized versus literature settings.<\/p>\n\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 1.<\/strong>\u00a0TIC for cocaine using optimized settings.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<style>.kb-image75756_6836a5-1d:not(.kb-image-is-ratio-size) .kb-img, .kb-image75756_6836a5-1d.kb-image-is-ratio-size{padding-bottom:var(--global-kb-spacing-xs, 1rem);}.kb-image75756_6836a5-1d .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<div class=\"wp-block-kadence-image kb-image75756_6836a5-1d\"><figure class=\"aligncenter size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"1230\" height=\"320\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-compound-optimization-in-lc-msms-why-does-it-matter-01.png\" alt=\"\" class=\"kb-img wp-image-75763\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-compound-optimization-in-lc-msms-why-does-it-matter-01.png 1230w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-compound-optimization-in-lc-msms-why-does-it-matter-01-300x78.png 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-compound-optimization-in-lc-msms-why-does-it-matter-01-1024x266.png 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-compound-optimization-in-lc-msms-why-does-it-matter-01-768x200.png 768w\" sizes=\"auto, (max-width: 1230px) 100vw, 1230px\" \/><\/figure><\/div>\n\n<\/div><\/div><\/div>\n\n<div class=\"wp-block-custom-chromatogram-article-top\"><div class=\"chromatogram-article-placeholder\"><div class=\"figure-heading\"><strong>Figure 2.<\/strong>\u00a0TIC for cocaine using literature settings.<\/div><div class='chromatogram-article-inner-full'><div class=\"chromatogram-article-inner\">\n<style>.kb-image75756_16c2bb-01:not(.kb-image-is-ratio-size) .kb-img, .kb-image75756_16c2bb-01.kb-image-is-ratio-size{padding-bottom:var(--global-kb-spacing-xs, 1rem);}.kb-image75756_16c2bb-01 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<div class=\"wp-block-kadence-image kb-image75756_16c2bb-01\"><figure class=\"aligncenter size-full\"><img loading=\"lazy\" decoding=\"async\" width=\"1230\" height=\"320\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-compound-optimization-in-lc-msms-why-does-it-matter-02.png\" alt=\"\" class=\"kb-img wp-image-75757\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-compound-optimization-in-lc-msms-why-does-it-matter-02.png 1230w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-compound-optimization-in-lc-msms-why-does-it-matter-02-300x78.png 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-compound-optimization-in-lc-msms-why-does-it-matter-02-1024x266.png 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-compound-optimization-in-lc-msms-why-does-it-matter-02-768x200.png 768w\" sizes=\"auto, (max-width: 1230px) 100vw, 1230px\" \/><\/figure><\/div>\n\n<\/div><\/div><\/div>\n\n\n<p class=\"wp-block-paragraph\" style=\"padding-top:var(--wp--preset--spacing--30);padding-bottom:0\"><strong>Table 3.<\/strong>&nbsp;Comparison of peak area for optimized versus literature settings.<\/p>\n\n\n<figure class=\"wp-block-table\">\n<table class=\"has-fixed-layout\">\n<tbody>\n<tr style=\"background-color: #4472c4; color: #ffffff;\">\n<td style=\"text-align: center;\"><strong>Analyte<\/strong><\/td>\n<td style=\"text-align: center;\"><strong>Peak Area<br \/>(Optimized Settings)<\/strong><\/td>\n<td style=\"text-align: center;\"><strong>Peak Area<br \/>(Unoptimized Settings)<\/strong><\/td>\n<td style=\"text-align: center;\"><strong>% Decrease<\/strong><\/td>\n<\/tr>\n<tr style=\"background-color: #cfd5ea;\">\n<td style=\"text-align: center;\">Cocaine<\/td>\n<td style=\"text-align: center;\">12293511<\/td>\n<td style=\"text-align: center;\">8656042<\/td>\n<td style=\"text-align: center;\">-29.58%<\/td>\n<\/tr>\n<tr style=\"background-color: #e9ebf5;\">\n<td style=\"text-align: center;\">Morphine<\/td>\n<td style=\"text-align: center;\">436044<\/td>\n<td style=\"text-align: center;\">238450<\/td>\n<td style=\"text-align: center;\">-45.31%<\/td>\n<\/tr>\n<tr style=\"background-color: #cfd5ea;\">\n<td style=\"text-align: center;\">\u03949-THC<\/td>\n<td style=\"text-align: center;\">597953<\/td>\n<td style=\"text-align: center;\">521493<\/td>\n<td style=\"text-align: center;\">-12.78%<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<\/figure>\n\n\n<p class=\"wp-block-paragraph\"><strong>Table 4.<\/strong>&nbsp;Comparison of peak height for optimized versus literature settings.<\/p>\n\n\n<figure class=\"wp-block-table\">\n<table class=\"has-fixed-layout\">\n<tbody>\n<tr style=\"background-color: #4472c4; color: #ffffff;\">\n<td style=\"text-align: center;\"><strong>Analyte<\/strong><\/td>\n<td style=\"text-align: center;\"><strong>Peak Height<br \/>(Optimized Settings)<\/strong><\/td>\n<td style=\"text-align: center;\"><strong>Peak Height<br \/>(Unoptimized Settings)<\/strong><\/td>\n<td style=\"text-align: center;\"><strong>% Decrease<\/strong><\/td>\n<\/tr>\n<tr style=\"background-color: #cfd5ea;\">\n<td style=\"text-align: center;\">Cocaine<\/td>\n<td style=\"text-align: center;\">4690398<\/td>\n<td style=\"text-align: center;\">3341265<\/td>\n<td style=\"text-align: center;\">-28.76%<\/td>\n<\/tr>\n<tr style=\"background-color: #e9ebf5;\">\n<td style=\"text-align: center;\">Morphine<\/td>\n<td style=\"text-align: center;\">149075<\/td>\n<td style=\"text-align: center;\">81472<\/td>\n<td style=\"text-align: center;\">-45.34%<\/td>\n<\/tr>\n<tr style=\"background-color: #cfd5ea;\">\n<td style=\"text-align: center;\">\u03949-THC<\/td>\n<td style=\"text-align: center;\">239200<\/td>\n<td style=\"text-align: center;\">211382<\/td>\n<td style=\"text-align: center;\">-11.63%<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<\/figure>\n\n\n<p class=\"wp-block-paragraph\">When using the literature settings, peak area and height significantly decreased for all analytes compared to when using the optimized settings. This loss of sensitivity would have definite consequences for achieving low analytical detection limits such as limit of detection (LOD) or limit of quantification (LOQ) for these analytes. This experiment demonstrates the importance of optimizing compound settings per individual LC-MS\/MS instruments.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Literature values can still provide useful information. In some instances, it may be acceptable to use the literature settings as is. You might use literature values if they meet your performance goals, as a starting point for compound optimization, or simply as confirmation that the settings you are using are appropriate.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">While compound optimization might add an extra step to the method development process, your detection limits will suffer without it. If you aren\u2019t sure how to effectively tune compounds on your lab\u2019s LC-MS\/MS, instrument manufacturers can provide step-by-step instructions on how to optimize compounds using your particular system. Happy tuning!<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">References<\/h2>\n\n\n\n<ol class=\"wp-block-list\">\n<li>Marilia S Cardoso, Rafael Lanaro, Raul C Dolores, Damila R Morais, Ana Carolina Furiozo Arantes, Karina Diniz Oliveira, Jose Luiz Costa, Determination of Drugs of Abuse in Hair by LC\u2013MS-MS: Application to Suicide Attempts Investigation, Journal of Analytical Toxicology, Volume 46, Issue 5, June 2022, Pages 577\u2013581, https:\/\/doi.org\/10.1093\/jat\/bkab058<\/li>\n<\/ol>\n\n\n<style>.wp-block-kadence-spacer.kt-block-spacer-75756_cb9cf6-41 .kt-block-spacer{height:60px;}.wp-block-kadence-spacer.kt-block-spacer-75756_cb9cf6-41 .kt-divider{border-top-width:1px;height:1px;border-top-color:#eee;width:80%;border-top-style:solid;}<\/style>\n<div class=\"wp-block-kadence-spacer aligncenter kt-block-spacer-75756_cb9cf6-41\"><div class=\"kt-block-spacer kt-block-spacer-halign-center\"><hr class=\"kt-divider\"\/><\/div><\/div>\n","protected":false},"excerpt":{"rendered":"<p>Compound optimization, also referred to as tuning, is the process in which mass spectrometry parameters such as the precursor\/product ions, collision energies, and other voltages are optimized for each analyte. Compound optimization can often be overlooked. It might seem easier to find an article or application note and just use the compound settings for a&#8230;<\/p>\n","protected":false},"author":13,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[9],"tags":[],"industries-application":[],"post-badge":[],"resource-type":[],"product-library":[],"resource-technique":[],"ppma_author":[432],"class_list":["post-75756","post","type-post","status-publish","format-standard","hentry","category-blogs"],"acf":[],"taxonomy_info":{"category":[{"value":9,"label":"Blogs"}]},"featured_image_src_large":false,"author_info":{"display_name":"Haley Berkland, MS","author_link":"https:\/\/discover.restek.com\/zh-hans\/author\/haley-berkland-ms\/"},"comment_info":0,"category_info":[{"term_id":9,"name":"Blogs","slug":"blogs","term_group":0,"term_taxonomy_id":9,"taxonomy":"category","description":"","parent":0,"count":457,"filter":"raw","cat_ID":9,"category_count":457,"category_description":"","cat_name":"Blogs","category_nicename":"blogs","category_parent":0}],"tag_info":false,"authors":[{"term_id":432,"user_id":13,"is_guest":0,"slug":"haley-berkland-ms","display_name":"Haley Berkland, MS","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-berkland-haley.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/people-berkland-haley.jpg"},"author_category":"1","first_name":"Haley","last_name":"Berkland","user_url":"https:\/\/www.linkedin.com\/in\/haley-berkland-60b66517b\/","job_title":"LC Applications Scientist","description":"Haley is an LC applications scientist at Restek. She attended Duquesne University, receiving her bachelor's degree in biochemistry and a master's degree in forensic science and law. As a graduate student, she performed research on the detection of drugs of abuse in vitreous humor by LC-MS\/MS. Before joining Restek in 2023, Haley spent four years working as a forensic toxicologist. While in this role, she performed analysis of postmortem toxicology casework, identification of seized drug evidence, and development\/validation of new assays by LC-MS\/MS, GC-MS, and GC-FID."}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts\/75756","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/users\/13"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/comments?post=75756"}],"version-history":[{"count":5,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts\/75756\/revisions"}],"predecessor-version":[{"id":75811,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts\/75756\/revisions\/75811"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/media?parent=75756"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/categories?post=75756"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/tags?post=75756"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/industries-application?post=75756"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/post-badge?post=75756"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/resource-type?post=75756"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/product-library?post=75756"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/resource-technique?post=75756"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/ppma_author?post=75756"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}