{"id":76393,"date":"2022-01-20T00:00:00","date_gmt":"2022-01-20T00:00:00","guid":{"rendered":"https:\/\/discover.restek.com\/?p=76393"},"modified":"2025-12-23T15:40:02","modified_gmt":"2025-12-23T15:40:02","slug":"adding-thc-o-acetate-to-your-potency-method","status":"publish","type":"post","link":"https:\/\/discover.restek.com\/zh-hans\/blogs\/gnbl5462\/adding-thc-o-acetate-to-your-potency-method","title":{"rendered":"Adding THC-O Acetate to your Potency Method"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">The cannabis industry is always innovating with a constant revolving door of hot new products. In the past we\u2019ve seen buzz around products containing delta-8-THC, delta-10-THC, and THCP.&nbsp;<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Recently there has been a lot of interest in tetrahydrocannabinol acetate, or THC-O acetate. THC-O acetate is a psychoactive synthetic compound produced by acetylating THC. As far as legality, there is arguably some grey area and varies by state. There are claims of THC-O acetate being stronger than delta-9-THC and many products containing THC-O acetate are becoming available on the market. As such, labs are interested in including this compound into their cannabinoid potency methods.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Herein, three methods have been developed to monitor 12 cannabinoids, including delta-10-THC and THC-O acetate by LC-UV. Since labs have different requirements, different methods were developed so the user has the flexibility to use an unbuffered or a buffered mobile phase system as well as methanol or acetonitrile in order to best fit their needs. For this analysis, a Raptor ARC-18 150 x 3 mm, 2.7 \u00b5m column was chosen to achieve desired resolution while keeping the analysis time under 12 minutes.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Figure 1<\/strong>. Method conditions and extracted chromatogram for the analysis of 13 cannabinoids by LC-UV using acetonitrile as an organic modifier and buffered aqueous mobile phase.<\/p>\n\n\n<style>.kb-image76393_a29653-42:not(.kb-image-is-ratio-size) .kb-img, .kb-image76393_a29653-42.kb-image-is-ratio-size{padding-top:var(--global-kb-spacing-xs, 1rem);padding-bottom:var(--global-kb-spacing-xs, 1rem);}.kb-image76393_a29653-42 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<figure class=\"wp-block-kadence-image kb-image76393_a29653-42 size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"316\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-adding-THC-O-acetate-to-your-potency-method-01-1024x316.jpg\" alt=\"\" class=\"kb-img wp-image-76961\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-adding-THC-O-acetate-to-your-potency-method-01-1024x316.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-adding-THC-O-acetate-to-your-potency-method-01-300x93.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-adding-THC-O-acetate-to-your-potency-method-01-768x237.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-adding-THC-O-acetate-to-your-potency-method-01-1536x474.jpg 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-adding-THC-O-acetate-to-your-potency-method-01.jpg 1800w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Figure 2<\/strong>. Method conditions and extracted chromatogram for the analysis of 13 cannabinoids by LC-UV using acetonitrile as an organic modifier without the use of buffer.<\/p>\n\n\n<style>.kb-image76393_090c70-7b:not(.kb-image-is-ratio-size) .kb-img, .kb-image76393_090c70-7b.kb-image-is-ratio-size{padding-top:var(--global-kb-spacing-xs, 1rem);padding-bottom:var(--global-kb-spacing-xs, 1rem);}.kb-image76393_090c70-7b .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<figure class=\"wp-block-kadence-image kb-image76393_090c70-7b size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"319\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-adding-THC-O-acetate-to-your-potency-method-02-1024x319.jpg\" alt=\"\" class=\"kb-img wp-image-76955\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-adding-THC-O-acetate-to-your-potency-method-02-1024x319.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-adding-THC-O-acetate-to-your-potency-method-02-300x93.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-adding-THC-O-acetate-to-your-potency-method-02-768x239.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-adding-THC-O-acetate-to-your-potency-method-02-1536x478.jpg 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-adding-THC-O-acetate-to-your-potency-method-02.jpg 1800w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>Figure 3<\/strong>. Method conditions and extracted chromatogram for the analysis of 13 cannabinoids by LC-UV using methanol as an organic modifier and buffered aqueous mobile phase.<\/p>\n\n\n<style>.kb-image76393_18d7ab-96:not(.kb-image-is-ratio-size) .kb-img, .kb-image76393_18d7ab-96.kb-image-is-ratio-size{padding-top:var(--global-kb-spacing-xs, 1rem);padding-bottom:var(--global-kb-spacing-xs, 1rem);}.kb-image76393_18d7ab-96 .kb-image-has-overlay:after{opacity:0.3;}<\/style>\n<figure class=\"wp-block-kadence-image kb-image76393_18d7ab-96 size-large\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"316\" src=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-adding-THC-O-acetate-to-your-potency-method-01-1024x316.jpg\" alt=\"\" class=\"kb-img wp-image-76961\" title=\"-\" srcset=\"https:\/\/discover.restek.com\/wp-content\/uploads\/blog-adding-THC-O-acetate-to-your-potency-method-01-1024x316.jpg 1024w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-adding-THC-O-acetate-to-your-potency-method-01-300x93.jpg 300w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-adding-THC-O-acetate-to-your-potency-method-01-768x237.jpg 768w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-adding-THC-O-acetate-to-your-potency-method-01-1536x474.jpg 1536w, https:\/\/discover.restek.com\/wp-content\/uploads\/blog-adding-THC-O-acetate-to-your-potency-method-01.jpg 1800w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">By using one of these methods, resolution can be achieved for all compounds analyzed within a 12 minute total analysis time.&nbsp;<\/p>\n\n\n\n<div style=\"height:100px\" aria-hidden=\"true\" class=\"wp-block-spacer\"><\/div>\n","protected":false},"excerpt":{"rendered":"<p>Recently there has been a lot of interest in tetrahydrocannabinol acetate, or THC-O acetate. THC-O acetate is a psychoactive synthetic compound produced by acetylating THC. As far as legality, there is arguably some grey area and varies by state. There are claims of THC-O acetate being stronger than delta-9-THC and many products containing THC-O acetate are becoming available on the market. As such, labs are interested in including this compound into their cannabinoid potency methods. Herein, three methods have been developed to monitor 12 cannabinoids, including delta-10-THC and THC-O acetate by LC-UV.<\/p>\n","protected":false},"author":47,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"_kad_blocks_custom_css":"","_kad_blocks_head_custom_js":"","_kad_blocks_body_custom_js":"","_kad_blocks_footer_custom_js":"","_kadence_starter_templates_imported_post":false,"_kad_post_transparent":"","_kad_post_title":"","_kad_post_layout":"","_kad_post_sidebar_id":"","_kad_post_content_style":"","_kad_post_vertical_padding":"","_kad_post_feature":"","_kad_post_feature_position":"","_kad_post_header":false,"_kad_post_footer":false,"footnotes":"","_ppma_block_editor_authors":""},"categories":[9],"tags":[],"industries-application":[],"post-badge":[],"resource-type":[],"product-library":[],"resource-technique":[],"ppma_author":[579],"class_list":["post-76393","post","type-post","status-publish","format-standard","hentry","category-blogs"],"acf":[],"taxonomy_info":{"category":[{"value":9,"label":"Blogs"}]},"featured_image_src_large":false,"author_info":{"display_name":"Jamie L. York, PhD","author_link":"https:\/\/discover.restek.com\/zh-hans\/author\/jamie-l-york-phd\/"},"comment_info":0,"category_info":[{"term_id":9,"name":"Blogs","slug":"blogs","term_group":0,"term_taxonomy_id":9,"taxonomy":"category","description":"","parent":0,"count":432,"filter":"raw","cat_ID":9,"category_count":432,"category_description":"","cat_name":"Blogs","category_nicename":"blogs","category_parent":0}],"tag_info":false,"authors":[{"term_id":579,"user_id":47,"is_guest":0,"slug":"jamie-l-york-phd","display_name":"Jamie L. York, PhD","avatar_url":{"url":"https:\/\/discover.restek.com\/wp-content\/uploads\/york-jamie-1000x1000-1-e1765483005988.jpg","url2x":"https:\/\/discover.restek.com\/wp-content\/uploads\/york-jamie-1000x1000-1-e1765483005988.jpg"},"author_category":"1","first_name":"Jamie","last_name":"York","user_url":"https:\/\/www.linkedin.com\/in\/jamielyork\/","job_title":"Principal Scientist, LC Solutions","description":"Jamie York is a principal scientist in the Applications Lab at Restek Corporation. She leads the development of innovative analytical methods tailored to the food, clinical, environmental, and cannabis markets. Jamie earned her PhD in chemistry from The University of Texas at Arlington, where she gained extensive expertise in a range of analytical techniques, including gas chromatography\u2013vacuum ultraviolet (GC\u2013VUV); gas chromatography\u2013mass spectrometry (GC\u2013MS); matrix-assisted laser desorption\/ionization (MALDI); and liquid chromatography\u2013 mass spectrometry (LC\u2013MS\/MS); with a research emphasis on food and environmental analysis. Today, her work focuses on complex method development and advanced sample preparation strategies to support the evolving needs of the scientific community."}],"_links":{"self":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts\/76393","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/users\/47"}],"replies":[{"embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/comments?post=76393"}],"version-history":[{"count":2,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts\/76393\/revisions"}],"predecessor-version":[{"id":76971,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/posts\/76393\/revisions\/76971"}],"wp:attachment":[{"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/media?parent=76393"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/categories?post=76393"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/tags?post=76393"},{"taxonomy":"industries-application","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/industries-application?post=76393"},{"taxonomy":"post-badge","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/post-badge?post=76393"},{"taxonomy":"resource-type","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/resource-type?post=76393"},{"taxonomy":"product-library","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/product-library?post=76393"},{"taxonomy":"resource-technique","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/resource-technique?post=76393"},{"taxonomy":"author","embeddable":true,"href":"https:\/\/discover.restek.com\/zh-hans\/wp-json\/wp\/v2\/ppma_author?post=76393"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}