FAQ

Is conditioning necessary for SPE cartridges?

30 Jul 2025

Yes. Detailed instructions for conditioning SPE cartridges can be found below. All SPE cartridges should be conditioned to wet and settle the bed, activate the packing materials, and remove any residual process materials (e.g., fines).

As a general guide:

  • Use one to two column volumes of the conditioning solution(s) recommended in the table below.
  • Use a high flow rate.
  • Do not allow the packing bed to go dry. The packing bed should be wet before adding sample.

Characteristics of Reversed- and Normal Phase SPE

PackingPolaritySample MatrixConditioning Solution(s)Elution Solution
C18, carbonpacking and analytes are nonpolarpolar, often aqueous(9a) methanol, then water or buffer (same as sample), or
(9b) final extraction solvent, then water-miscible solvent, then water or buffer
nonpolar solvent or mixed solvent solution
silica, Florisil, carbonpacking and analytes are polarnonpolar, often an organic solventfresh solvent (same as sample)fresh solvent

Autoren

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    View Articles
GNOT5144