FAQ

How do I install a GC column into the new Flip Seal adaptor nut?

31 Jul 2025

Insert the GC column through the 1/16” adaptor sealing nut and a standard 1/16” graphite or graphite-Vespel capillary ferrule. Make a clean cut across the GC column. Install the column into the adaptor 30-32 mm from the bottom of the 1/16” nut, which places the column end a few millimeters above the bottom of the inlet seal, the perfect installation distance! Don’t forget to leak check with a Restek Electronic Leak Detector (cat.# 28500) after tightening the adaptor nut to seal the GC column to the inlet and re-pressurizing the inlet.

Column sealing nut shown 30-32 mm from the end of the column.

Seat the column sealing nut so that
the bottom is 30-32 mm from
the end of the column.

Column being inserted into the adaptor fitting.

Insert the column into the adaptor fitting.

Sealing nut being tightened onto the adaptor fitting.

Tighten the sealing nut onto the adaptor fitting.

Flip Seal inlet seal in adaptor fitting.

Perfect installation distance for column in adaptor fitting.

Autoren

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    Alle Posts ansehen
GNOT4962