Artikel

Avoid Mycotoxin Quantitation Errors When Using Stable Isotope Dilution Assay (SIDA)

21 Oct 2020

feature FFAR3036

There are significant benefits to using the stable isotope dilution assay (SIDA) calibration approach for mycotoxin analyses [1], and one of the biggest is being able to use a single calibration curve for multiple matrix types. Compared to making individual matrix-matched calibration standards for each sample type, the time savings offered by the SIDA technique can make up for the relatively high price of isotopically labeled internal standards (IS), especially for high-throughput food safety laboratories that frequently deal with a wide variety of samples. However, labeled standards are not always commercially available for every mycotoxin of interest. Due to these cost and availability issues, labs may be interested in using a practice that is common among many other types of methods that use internal standards: using the same internal standard to calibrate a group of analytes. That approach may seem logical, but in the case of mycotoxin SIDA methods, it can lead to significant errors.

Table I compares the observed concentrations of four different mycotoxins present in two maize flour reference materials to the concentrations reported by the reference material supplier. For the three mycotoxins that were quantified using their corresponding isotopically labeled internal standards, the agreement between the two values is excellent. However, in the case of the mycotoxin zearalenone, we did not have its analogous labeled internal standard, so we quantified it using 13C17-aflatoxin G1 because it elutes nearby. As the data show, the agreement between the observed results and those reported by the reference material supplier is very poor, which illustrates that similar chromatographic retention alone is not enough to predict the effects of sample preparation and/or matrix-related changes in ionization efficiency. A more detailed description of Restek’s research on the LC-MS/MS analysis of mycotoxins in various foods comparing SIDA to matrix-matched calibration is published in a peer-reviewed study [2].

In light of these results, it is strongly recommended that only matching isotopically labeled internal standards be used for quantification in mycotoxin SIDA methods. If matching internal standards are not available, other calibration approaches, such as matrix-matched calibration, should be used.

Table I: Mycotoxin SIDA calibration should only be used for matching analyte and IS pairs.
Reference
Material
AnalyteISMeasured
Concentration
(ng/g), n=3
Assigned
Concentration
(ng/g)
Percent
Accuracy
(RSD, %)
TET017RMDeoxynivalenol13C15-Deoxynivalenol1867.9 ± 37.361971 ± 19594.8 (2)
TET017RMAflatoxin B113C17-Aflatoxin B18.68 ± 0.4349.49 ± 0.8591.4 (5)
TET017RMOchratoxin A13C20-Ochratoxin A4.48 ± 0.1344.81 ± 0.7593.2 (3)
TET017RMZearalenone13C17-Aflatoxin G131.26 ± 2.19a231 ± 2513.5 (7)a
T04301QDeoxynivalenol13C15-Deoxynivalenol639.7 ± 19.19649 ± 22298.6 (3)
T04301QAflatoxin B113C17-Aflatoxin B18.69 ± 0.3489.21 ± 4.0594.4 (4)
T04301QOchratoxin A13C20-Ochratoxin A2.81 ± 0.1973.03 ± 1.3392.6 (7)
T04301QZearalenone13C17-Aflatoxin G116.2 ± 0.810a138.5 ± 59.611.7 (5)a

a Results quantified using a nonmatched labeled IS.

Learn more at www.restek.com/inert

teaser product lc columns inert lc 02

References

  1. K. Zhang, K., M.R. Schaab, G. Southwood, E.R. Tor, L.S. Aston, W. Song, B. Eitzer, S. Majumdar, T. Lapainis, H. Mai, K. Tran, A. El-Demerdash, V. Vega, Y. Cai, J.W. Wong, A.J. Krynitsky, T.H. Begley, A collaborative study: determination of mycotoxins in corn, peanut butter, and wheat flour using stable isotope dilution assay (SIDA) and liquid chromatography-tandem mass spectrometry (LC-MS/MS), J. of Agric. Food Chem. 65 (33) (2017) 7138–7152. DOI: 10.1021/acs.jafc.6b04872 https://pubs.acs.org/doi/10.1021/acs.jafc.6b04872
  2. D. Li, J.A. Steimling, J.D. Konschnik, S. Grossman, T. Kahler, Quantitation of mycotoxins in four food matrices comparing stable isotope dilution assay (SIDA) with matrix-matched calibration methods by LC–MS/MS, J. AOAC Int. (2019) DOI: 10.5740/jaoacint.19-0028 https://academic.oup.com/jaoac/article-abstract/102/6/1673/5658363

Author

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    View all posts
FFAR3036
Version: B