Artikel

How to Read a Trace Analysis Data Packet

03 Mar 2026

feature GNAR5426A

Trace analysis data packets, which are part of our expanded data packs for Restek’s single and multicomponent PFAS reference standards, include detailed information to help you understand the composition of your standards. This article explains how to read your packet, using our br-PFHxS standard (cat.# 30803) as an example.

You can access the trace analysis data packet for your standard by using our documentation search tool and entering your standard’s cat.# and lot#. You can also access your standard’s certificate of analysis (CofA) at the same location.

Restek’s trace analysis data packets include data from two methods, each for a different purpose. The first section of the data pack uses LC-MS/MS for trace analysis of samples. The second section, which is explained later in this article, uses NMR to identify individual isomers.

LC-MS/MS Trace Analysis of Samples

Analytical Method Summary Table

The first table (figure 1) within the trace analysis data packet includes a summary of the analytical method used for trace analysis of samples.

Figure 1: LC MS/MS Data

LC MS/MS Data
Instrument – Waters ACQUITY Premier
Autosampler: Acquity Binary
Pump: Acquity SM
Detector: Xevo TQ-Abs
ColumnForce C18 (Cat#: 9634252)
Mobile Phase A0.5mM ammonium acetate in water
Mobile Phase BMethanol
Injection Volume3uL
Method Details
Gradient: Initial 80% A, decrease to 5% A by 6mins, hold at 5% A until 8.5mins, change to 80% A then hold at 80% A until 9.5min.
Sample Preparation: Samples were prepared in a mixture of approximately 200ppb, 2ppb, and 0.02ppb in 80:20 methanol:5mM ammonium acetate in water for analysis.
MDL: Estimated for each analytical batch by averaging the response of a minimum of 5 blank samples run throughout the analytical batch.
Summary of Results
PFHxS-99.44, PFPeS-0.3, PFOA-0.1, PFBS-0.07, 6:2 FTS-0.05

Method detection limit (MDL) is determined in each analytical batch to ensure there is no background contamination within the instrument. This is used to determine the final LOQ.

The Summary of Results for each analyte is measured >LOQ in % wt.

MRM Transitions Table

The second table (figure 2) contains an analyte list and transitions measured by the trace analysis.

Figure 2: MRM Transitions

CompoundParent Ion 1Daughter Ion 1Parent Ion 2Daughter Ion 2CompoundParent Ion 1Daughter Ion 1Parent Ion 2Daughter Ion 2
PFBA213169PFNA463219463419
PFMPA22985PFOS4998049999
3:3FTCA241117241177FOSA49878498478
PFPeA26369263219NMeFOSA512169512219
PFMBA27985PFDA513219513469
HFPO-DA285169285185NEtFOSA526169526219
NFDHA295852952018:2 FTS52781527507
PFBS29980299999Cl-PF3ONS531351533353
PFHxA313119313269PFNS5498054999
PFEESA31583315135PFUnA563269563519
4:2 FTS32781327307NMeFOSAA570419570483
5:3FTCA341217341237NEtFOSAA584419584526
PFPeS3498034999PFDS5998059999
PFHpA363169363319PFDOA613319613569
ADONA37785377251NMeFOSE61659
PFHxS3998039999NEtFOSE63059
PFOA41316941336911Cl-PF3OUdS631451633453
6:2 FTS42781427407PFTrDA663169663619
7:3 FTCA441317441337PFDoS6998069999
PFHpS4498044999PFTeDA713169713669

Quantification is based on response from transition corresponding to Parent 1 > Daughter 1. Secondary transition is used for qualitative identification.

Trace Analysis Summary

The third table (figure 3) is summary of the trace analysis. We have color coded the columns to help explain the table. The columns will be uncolored in your data packet.

Figure 3: Trace Analysis Summary

 Reporting
Limit (%wt)
LOQ
(ng/g)
Measured Conc (ng/mL)Dilution factor%
Composition
PFHxS0.050.012.4010099.45
PFPeS0.050.010.7310.30
PFOA0.100.010.2510.10
PFBS0.050.010.1710.07
6:2 FTS0.050.050.1410.06
PFBA0.100.010.041<0.1
PFMPA0.100.05N.D.1<0.1
3:3FTCA0.100.03N.D.1<0.1
PFPeA0.100.06N.D.1<0.1
PFMBA0.100.03N.D.1<0.1
HFPO-DA0.100.03N.D.1<0.1
NFDHA0.100.05N.D.1<0.1
PFHxA0.100.20N.D.1<0.1
PFEESA0.100.01N.D.1<0.1
4:2 FTS0.100.02N.D.1<0.1
5:3FTCA0.050.05N.D.1<0.05
PFHpA0.100.31N.D.1<0.1
ADONA0.100.01N.D.1<0.1
7:3 FTCA0.100.05<LOQ1<0.1
PFHpS0.100.31N.D.1<0.1
PFNA0.050.01N.D.1<0.05
PFOS0.050.06N.D.1<0.05
FOSA0.130.02N.D.1<0.125
NMeFOSA0.130.01N.D.1<0.125
PFDA0.100.01N.D.1<0.1
NEtFOSA0.100.01N.D.1<0.1
8:2 FTS0.050.05N.D.1<0.05
9Cl-PF3ONS0.050.05<LOQ1<0.05
PFNS0.050.01<LOQ1<0.05
PFUnA0.100.01N.D.1<0.1
NMeFOSAA0.130.01N.D.1<0.125
NEtFOSAA0.130.01N.D.1<0.125
PFDS0.050.01N.D.1<0.05
PFDOA0.100.03N.D.1<0.1
NMeFOSE0.130.09N.D.1<0.125
NEtFOSE0.130.08<LOQ1<0.125
11Cl-PF3OUdS0.050.05N.D.1<0.05
PFTrDA0.100.38N.D.1<0.1
PFDoS0.050.01N.D.1<0.05
PFTeDA0.100.025N.D.1<0.1

Reporting Limit (red column): Typically, 0.05% for sulfonates, 0.10% carboxylic acids, and 0.125% for sulfonamides.

LOQ (blue column): Limit of Quantitation (LOQ) of the instrument. Value is equal to the lowest calibrator or the batch specific MDL, whichever is greater.

Measured Conc (green column): Measured concentration on instrument. <LOQ means the peak was detected but below the analyte’s LOQ. Not Detected (N.D.) analyte was below the detection limit of the instrument.

Dilution factor (orange column): The dilution factor each analyte was quantified from. Target analyte typically had to be diluted to ensure the measured concentration was within the linear range of the analyte.

% Composition (purple column): Results in % wt for each target analyte. Values are calculated from Measured Concentration * Dilution Factor. Final values are normalized to 100%. Reporting limits are applied after normalization.

Overlay MRM Chromatogram

The next data (figure 4) is an overlay MRM chromatogram of the trace analysis. The full scan data can be found in the CofA. The data is typically from the most concentrated value measured.

Figure 4: Overlay MRM Chromatogram

figure article GNAR5426 04 1

NMR Analysis to Identify Individual Isomers

NMR Table

Figure 5: NMR Table

The fourth table (figure 5) includes NMR data, which lists sample preparation information and summary of results.

NMR
Isomeric ratios determined by NMR will differ from ratios determined chromographically due to unresolved separation. Identification of individual isomers is based on published work. (doi:10.1016/j.chemosphere.2007.06.096)
Samples were prepared at approximately 10mM with an approximate MDL of 1%wt.
Summary of Results
The following isomers were identified: 1-m, 2-m, 3-m, 4-m, and 3,3-dm.

In the Summary of Results, the number denotes distance from end-functional group. m stands for “methyl” and dm for “dimethyl,” indicating the isomer contains either a single-branching CF3 group or two CF3 groups respectively.

Figure 6: Example Numbering Convention for PFAS Isomers

Isomer structure

Figure 7: Example Isomer Naming Convention

examples of isomer nomenclature

NMR Data

The final data (figure 8) included in the trace analysis data packet is additional NMR data.

Figure 8: NMR Data-Annotated

figure article GNAR5426 08A

Figure annotations:

  1. Acquisition parameters for NMR data.
  2. Hexafluorobenzene (HFB) added as an internal standard to determine relative chemical shifts.

Full Support for Your Restek PFAS Standards

For more information about our expanded data packs and PFAS reference standards, review our full reference standard FAQ. Additionally, if you have questions about your specific standard, contact our Technical Service team.

Find our full selection of reference standards here.

Need Help with Reference Standards?

From method selection to product recommendations, our reference standards experts are here to help you navigate complex workflows and achieve accurate, reliable results.

Tags

Autoren

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    View Articles
GNAR5426