Artikel

Fast Analysis of MCPDs by LPGC-MS

Featured Application: MCPDs on LPGC Rxi-17Sil MS

03 Oct 2025

feature GNFA4239
  • 2x faster than conventional GC-MS analysis.
  • Reduces helium usage by 69% compared to conventional GC-MS methods. 
  • Rxi-17Sil MS kits are factory connected with a proven leak-free connector making LPGC as simple as a column change.

MCPDs are impurities found in food products that result from heat processes in manufacturing. They have been largely detected in refined oils, including vegetable oil and palm oil, along with processed foods that use these oils in their formulations, including baby formula. MCPDs are typically analyzed by GC-MS using a 30 m x 0.25 mm ID x 0.25 μm capillary column with a 5% diphenyl, 95% dimethyl polysiloxane or a 50% diphenyl, 50% dimethyl polysiloxane stationary phase. Run times using conventional GC-MS are typically over 12 minutes. With the large volumes of samples laboratories need to process, there is a motivation to cut down on run times, both to increase sample throughput and save on helium costs. However, techniques to shorten analysis times often compromise critical isomer separations.

Low-pressure gas chromatography (LPGC) offers a unique solution: by lowering the pressure inside the GC column, analysis times and helium usage are significantly reduced, allowing laboratories to analyze more samples at a lower production cost. Utilizing our LPGC Rxi-17Sil MS kit (cat.# 11805), analysis time was cut in half, and helium consumption was reduced by 69% compared to conventional GC-MS methods. Additionally, good resolution was achieved for the critical separation of 2-MCPD and 3-MCPD, and these analytes were both resolved well before the derivatization agent. LPGC is as easy as a column change and offers a significant benefit by reducing analysis time, allowing for increased sample throughput while also conserving helium. Each factory-coupled LPGC kit is individually tested to ensure a leak-free connection. To learn more about the LPGC technique, visit www.restek.com/lpgc

Figure 1: MCPDs on LPGC Rxi-17Sil MS Compared to Conventional GC-MS.
Comparison of Conventional and LPGC Analysis of 2-MCPD and 3-MCPD

GC_FS0595

Peaks

PeakstR  (30 m)tR  (LPGC)Conc. (µg/mL)
1.3-MCPD7.7672.92130
2.2-MCPD8.1033.07550

Conditions

ColumnSee notes
Standard/Sample
3-MCPD
2-MCPD
Diluent:Isooctane
Conc.:30-50 ppm (See the peak table.)
Injection
Inj. Vol.:1 µL split (split ratio 10:1)
Liner:Topaz 4.0 mm ID single taper inlet liner w/ wool (cat.# 23303)
Inj. Temp.:280 °C
Carrier GasHe
DetectorMS
Mode:Scan
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Source Temp.:330 °C
Quad Temp.:180 °C
Electron Energy:70 eV
Tune Type:PFTBA
Ionization Mode:EI
InstrumentAgilent 7890B GC & 5977A MSD
Sample PreparationStandards were derivatized with 20 µL phenylboronic acid (PBA), saturated solution in diethyl ether, dried, and then reconstituted in 1 mL isooctane in a 2 mL, short-cap, screw-thread vial (cat.# 21143) and capped with a short-cap, screw-vial closure (cat.# 24495). Final concentrations are given in the peak table.
NotesConventional (30 m) Analysis:
Column: Rxi-17Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 14123)
Temp. program: 100 °C (hold 0.5 min) to 180 °C at 12 °C/min to 320 °C at 25 °C/min (hold 4 min)
Flow: 1.4 mL/min
Scan start time: 3 min
Scan range: 50-350 amu
Scan rate: 8.5 scans/sec

LPGC-MS Analysis:
Column: LPGC Rxi-17Sil MS column kit, includes 10 m x 0.32 mm ID x 0.25 μm Rxi-17Sil MS analytical column and 5 m x 0.15 mm ID Rxi restrictor factory connected via SilTite connector (cat.# 11805)
Temp. program: 100 °C (hold 0.5 min) to 320 °C at 30 °C/min (hold 5 min)
Flow: 0.9 mL/min
Scan start time: 1.5 min
Scan range: 50-350 amu
Scan rate: 8.5 scans/sec

Products Mentioned


LPGC Rxi-17Sil MS Säulenkit
Topaz, Single Taper Inlet Liner, 4.0 mm x 6.5 x 78.5, für Agilent GCs, mit Quarzwolle, Premium-Deaktivierung, 5er Pack
GC Accelerator-Ofeneinsatz-Kit, für Agilent 5890, 6890, 7890 und 8890 GCs
Short-Cap-Vial mit graduiertem Markierungsfeld, 9-425 Screw-Thread, 2.0 mL, 9 mm, 12 x 32 (nur Vial), braun, 1000er Pack
Short Screw Kappen, Polypropylen, Screw-Thread, PTFE/Silikon/PTFE Septen, schwarz, vormontiert, 2.0 mL, 9 mm, 100er Pack

Author

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    View all posts
GNFA4239