Artikel

Improve Your Linear Alpha Olefins (LAO) Impurity Analysis with Restek’s New Rxi-LAO GC Columns

04 Oct 2022

feature PCSS3860

Specifically applicated for linear alpha olefins (LAO) impurity analysis, Restek’s new Rxi-LAO GC columns provide accurate analysis of LAO compounds including 1-butene, 1-hexene, and 1-octene. Combining a stationary phase with a unique selectivity and an optimal, one-column method, these new columns help labs improve their LAO analysis by increasing system uptime and sample throughput.

Rxi-LAO columns are defined by four key features:

  • Unique selectivity enables high resolution of impurities from peaks of interest for excellent data quality.
  • One-column method reduces instrument setup and analysis time, resulting in fast, accurate, and complete analysis.
  • Application-specific column dimensions increase sample throughput.
  • Pro EZGC chromatogram modeler libraries simplify analysis optimization.

A GC Column Applicated for Linear Alpha Olefins Analysis

Restek’s Rxi-LAO columns are the first GC columns specifically applicated for the impurity analysis of linear alpha olefins. Our unique selectivity achieves high resolution of impurities from peaks of interest, ensuring these new columns provide labs with the separation power needed for LAO impurity analysis (Figures 1 and 2).

Figure 1: The unique selectivity of Rxi-LAO columns provides high resolution of impurities from peaks of interest in the analysis of 1-Hexene.
1-Hexene on Rxi-LAO (60 m x 0.25 mm x 1.4 µm)

GC_PC1378

Peaks

PeakstR (min)
1.3-Methyl-1-pentene14.06
2.3-Methylpentane16.31
3.1-Hexene18.04
4.Hexane18.39
5.2-Ethyl-1-butene19.08
6.cis-3-Hexene19.61
PeakstR (min)
7.trans-2-Hexene19.61
8.cis-3-Methyl-2-pentene20.41
9.cis-2-Hexene20.81
10.trans-3-Methyl-2-pentene21.19
11.Methyl-cyclopentene22.87
12.Cyclohexene23.24

Conditions

ColumnRxi-LAO, 60 m, 0.25 mm ID, 1.4 µm (cat.# 13876)
Standard/Sample1-Hexene
Conc.: Neat solvent
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID low pressure drop Precision inlet liner w/wool (cat.# 23309)
Inj. Temp.:250 °C
Split Vent Flow Rate:125 mL/min
Oven
Oven Temp.:35 °C (hold 20 min) to 160 °C at 30 °C/min (hold 20 min)
Carrier GasHe, constant flow
Linear Velocity:23 cm/sec @ 35 °C
DetectorFID @ 300 °C
Make-up Gas Flow Rate:45 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:450 mL/min
Data Rate:20 Hz
InstrumentAgilent 7890B GC
Sample PreparationThe sample was pipetted into a 2 mL vial (cat.# 21142) and capped with a short screw cap (cat.# 24498).
Notes• Compounds were tentatively identified using a mass spectrometer and method translation.
• A 208V instrument was used.
• Benzene is not present in the sample, but if it were, it would elute at 23.73 minutes under these conditions.
Figure 2: Rxi-LAO columns reliably separate 1-Octene from impurities.
1-Octene on Rxi-LAO (60 m x 0.25 mm x 1.4 µm)

GC_PC1381

Conditions

ColumnRxi-LAO, 60 m, 0.25 mm ID, 1.4 µm (cat.# 13876)
Standard/Sample1-Octene
Conc.: Neat solvent
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID low pressure drop Precision liner w/wool (cat.# 23309)
Inj. Temp.:250 °C
Split Vent Flow Rate:125 mL/min
Oven
Oven Temp.:35 °C (hold 20 min) to 160 °C at 30 °C/min (hold 20 min)
Carrier GasHe, constant flow
Linear Velocity:23 cm/sec @ 35 °C
DetectorFID @ 300 °C
Make-up Gas Flow Rate:45 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:450 mL/min
Data Rate:20 Hz
InstrumentAgilent 7890B GC
Sample PreparationThe sample was pipetted into a 2 mL vial (cat.# 21142) and capped with a short screw cap (cat.# 24498).
NotesPeaks are tentatively identified.

An LAO Workhorse: Increased Uptime and Increased Sample Throughput with a Single Column

As worldwide demand for polyethylene increases, so does the pressure on labs performing LAO impurity analysis. These labs need their GC columns to do more—accuracy isn’t enough: their columns need to be dependable and fast.

Rxi-LAO columns achieve high selectivity for liner alpha olefins impurities while requiring only a one-column method, resulting in substantial benefits compared to two-column or even current one-column methods. To help meet the needs of different labs, Rxi-LAO columns are available in two dimensions (Table I).

For labs using a two-column/two-GC method, we recommend our Rxi-LAO 60 m, 0.25 mm ID, 1.4 µm column (cat.# 13876). Compared to a two-column method, this single-column option halves the number of columns needed, resulting in reduced setup time, column costs, and analysis time while maintaining data quality (Figures 1 and 2).

For labs using a single-column method, we recommend our shorter, narrow-bore Rxi-LAO 40 m, 0.18 mm ID, 1.0 µm column (cat.# 40815). The smaller dimensions of this column enables fast analysis times while maintaining the separation power needed for LAO impurity analysis (Figures 3 and 4, Table II).

Table I: Switching to a one-column method using Rxi-LAO columns can offer labs substantial benefits over current methods.
Current MethodRxi-LAO Column Dimension for One-Column MethodRxi-LAO Column AdvantageBenefits vs. Current Method
2 GC/2 Column60 m, 0.25 mm ID, 1.4 µm (cat.# 13876)One-column method reduces setup time. Optimal stationary phase selectivity reduces analysis time and yields good resolution of impurities.Increased instrument uptime. Increased sample throughput. Accurate, complete results.
1 GC/1 Column40 m, 0.18 mm ID, 1.0 µm (cat.# 40815)Shorter, small ID column produces faster analysis times while maintaining impurity resolutions.Increased sample throughput. Good data quality.
Figure 3: The optimal column dimensions of our 40 m, 0.18 mm ID, 1.0 µm Rxi-LAO column (cat.# 40815) allows even faster run times from a single column analysis while maintaining resolution of impurities from 1-Hexene.
1-Hexene on Rxi-LAO (40 m x 0.18 mm x 1.0 µm)

GC_PC1380

Peaks

PeakstR (min)
1.3-Methyl-1-pentene5.25
2.3-Methylpentane6.10
3.1-Hexene6.87
4.Hexane6.94
5.2-Ethyl-1-butene7.16
6.cis-3-Hexene7.34
7.trans-2-Hexene7.35
8.cis-3-Methyl-2-pentene7.66
9.cis-2-Hexene7.91
10.trans-3-Methyl-2-pentene8.23

Conditions

ColumnRxi-LAO, 40 m, 0.18 mm ID, 1.0 µm (cat.# 40815)
Standard/Sample1-Hexene
Conc.: Neat solvent
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID low pressure drop Precision inlet liner w/wool (cat.# 23309)
Inj. Temp.:250 °C
Split Vent Flow Rate:178 mL/min
Oven
Oven Temp.:35 °C (hold 11.85 min) to 160 °C at 50 °C/min (hold 12 min)
Carrier GasHe, constant flow
Linear Velocity:39.19 cm/sec @ 35 °C
DetectorFID @ 300 °C
Make-up Gas Flow Rate:45 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:400 mL/min
Data Rate:20 Hz
InstrumentAgilent 7890B GC
Sample PreparationThe sample was pipetted into a 2 mL vial (cat.# 21142) and capped with a short screw cap (cat.# 24498).
Notes• Compounds were tentatively identified using a mass spectrometer and method translation.
• A 208V instrument was used.
• Benzene is not present in the sample, but if it were, it would elute at 12.36 minutes under these conditions.
Figure 4: The narrow-bore Rxi-LAO column (cat.# 40815) accurately resolves 1-Octene within a fast, 18-minute run time.
1-Octene on Rxi-LAO (40 m x 0.18 mm x 1.0 µm)

GC_PC1382

Conditions

ColumnRxi-LAO, 40 m, 0.18 mm ID, 1.0 µm (cat.# 40815)
Standard/Sample1-Octene
Conc.: Neat solvent
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID low pressure drop Precision inlet liner w/wool (cat.# 23309)
Inj. Temp.:250 °C
Split Vent Flow Rate:178 mL/min
Oven
Oven Temp.:35 °C (hold 11.85 min) to 160 °C at 50 °C/min (hold 12 min)
Carrier GasHe, constant flow
Linear Velocity:39.19 cm/sec @ 35 °C
DetectorFID @ 300 °C
Make-up Gas Flow Rate:45 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:400 mL/min
Data Rate:20 Hz
InstrumentAgilent 7890B GC
Sample PreparationThe sample was pipetted into a 2 mL vial (cat.# 21142) and capped with a short screw cap (cat.# 24498).
Notes• Peaks are tentatively identified.
• A 208V instrument was used.

Table II: The faster analysis times produced by the 40 m Rxi-LAO column (cat.# 40815) make it ideal for labs seeking a high throughput solution.
Rxi-LAO Column DimensionRun time (min)
1-Hexene Analysis1-Octene Analysis
60 m, 0.25 mm ID, 1.4 µm (cat.# 13876)2432
40 m, 0.18 mm ID, 1.0 µm (cat.# 40815)918

Renowned Rxi Column Technology as Standard

As part of the Rxi GC column family, Rxi-LAO columns benefit from the industry-leading GC column technology that is standard across our entire Rxi column line.

High Reproducibility

Our Rxi-LAO columns feature unmatched manufacturing precision and stringent quality control to deliver high column-to-column reproducibility. This ensures your Rxi-LAO column performs to the same high standard with every analysis, even after changing the column (Figure 5).

Figure 5: Rxi-LAO columns are engineered and QC tested to ensure column-to-column and lot-to-lot reproducibility.
Lot-to-Lot Comparison: 1-Hexene on Rxi-LAO

GC_PC1379

Peaks

PeakstR (min)
1.1-Hexene18.04
2.Hexane18.39
3.2-Ethyl-1-butene19.08
4.cis-3-Hexene19.60
5.trans-2-Hexene19.61
6.cis-3-Methyl-2-pentene20.41
7.cis-2-Hexene20.81
8.trans-3-Methyl-2-pentene21.19

Conditions

ColumnRxi-LAO, 60 m, 0.25 mm ID, 1.4 µm (cat.# 13876)
Standard/Sample1-Hexene
Conc.: Neat solvent
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID low pressure drop Precision inlet liner w/wool (cat.# 23309)
Inj. Temp.:250 °C
Split Vent Flow Rate:125 mL/min
Oven
Oven Temp.:35 °C (hold 20 min) to 160 °C at 30 °C/min (hold 20 min)
Carrier GasHe, constant flow
Linear Velocity:23 cm/sec @ 35 °C
DetectorFID @ 300 °C
Make-up Gas Flow Rate:45 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:450 mL/min
Data Rate:20 Hz
InstrumentAgilent 7890B GC
Sample PreparationThe sample was pipetted into a 2 mL vial (cat.# 21142) and capped with a short screw cap (cat.# 24498).

Low Bleed

The high stability of our Rxi columns means they generate less bleed and reduce background to further improve signal-to-noise ratios, enhance sensitivity, and lower detection limits. Together, these two advanced technologies help ensure your LAO data is highly accurate and that you get the right results fast.

Learn more about the technology that goes into our Rxi GC columns at www.restek.com/rxi

Simplify Your Analysis Optimization with Pro EZGC Software

Rxi-LAO columns are fully supported by our Pro EZGC chromatogram modeler, allowing labs to simplify their analysis optimization. This free, easy-to-use software helps you perform GC method optimization and method development in minutes—without needing to use an instrument.

A powerful feature of Pro EZGC software is its ability to model the performance of different column dimensions. If your LAO impurity application currently uses a single-column method, and you would like to see how changing to the shorter, 40 m Rxi-LAO column can speed up your analysis, Pro EZGC software can be used to generate simulated results with exceptional accuracy.

Figure 6: Rxi-LAO GC columns are fully supported by Restek’s Pro EZGC chromatogram modeler, helping labs quickly and easily optimize their LAO analysis.
figure article PCSS3860 01

Erwähnte Produkte


Rxi-LAO GC Kapillarsäule, 60 m x 0.25 mm ID x 1.4 µm
Rxi-LAO GC Kapillarsäule, 40 m x 0.18 mm ID x 1.0 µm

Autoren

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    Alle Posts ansehen
PCSS3860