Blogs

Is Inert LC Column Hardware Beneficial for Pesticide or Mycotoxin Analysis?

19 Aug 2024

For some years now, the development departments of high-pressure liquid chromatography companies have been dealing with the question of what influence the surfaces of the instrument and component hardware can have on peak shape and sensitivity in the detection of polar, chelating molecules. Biomolecules in particular, with their complex acid/base properties, but also compounds such as glyphosate and AMPA, with their organophosphate groups, show that the surface properties of the components in the sample path have a decisive influence on peak shape and sensitivity.

Recently, Restek developed a inert LC column line to address this issue more efficiently and effectively than short-term solutions, such as priming or medronic acid deactivation. Reports from several independent laboratories indicate that these new columns can offer substantial benefits for pesticides and mycotoxins analysis.

In June 2024, Maria Antonietta Carrera from the Department of Desertification and Geo-ecology, Experimental Station of Arid Zones in Almeria, under the supervision of Amadeo Fernandez-Alba from the European Union Reference Laboratory for Pesticide Residues in Fruit & Vegetables (EURL-FV), published a remarkable paper on Simultaneous analysis of pesticides and mycotoxins in primary processed foods: The case of bee pollen”, in which she compared a column commonly used for routine analysis of the analytes under consideration with a Raptor Inert ARC-18 column from Restek.

In her publication she describes “As reported in Fig. 3, no substantial differences were revealed by this test, since almost the same number of compounds could be detected at a concentration of 1 μg/kg with both columns. However, by looking at peaks obtained for about 15 % of analytes at 1 μg/kg, it was possible to notice a visible improvement in the peak shape (Fig. 4) when using the Raptor Inert column. This particular feature can be sometimes decisive for the identification of the analytes in real samples.”

Please find the full publication here: https://www.cell.com/heliyon/fulltext/S2405-8440(24)09543-4 

In addition to influential labs, such as EURL, that assess the application of new analytical developments within their specific fields, method developers in instrument companies must ensure that the performance statement about their devices is not countered by poor chromatography, for which they bear no responsibility.

LCTech GmbH has made a name for itself in automated sample preparation for the analysis of mycotoxins, PFAS, and other multicomponent analyte panels. Frederik Wuppermann, Product Manager Biotechnology at LCTech, describes his experience with the inert columns in mycotoxin analysis as follows.

“The separation of aflatoxin B1, G1, B2, G2, OTA, zearalenone, deoxynivalenol, T2, H-T2, fumonisin B1 and B2 after CrossTOX cleanup in high-throughput mycotoxin analysis was very good. The shorter conditioning times were also very beneficial. I can certainly also imagine the Inert Biphenyl column as an alternative for other applications.”

In addition to these independent reports, Restek has assessed the impact of the new inert LC columns on pesticides and mycotoxins. We selected these groups of analytes because they include a variety of diverse compound chemistries and are often analyzed together in multi-class screening analyses. In such screening analyses, it is very important to ensure reliable identification and quantification of the individual components at trace levels (low LOQs), and an inert LC column can help with this by providing proper peak shape and good sensitivity. For both groups, we selected an appropriate stationary phase, analyzed the target compounds on both conventional LC columns and the new inert LC columns, and compared the chromatographic results. Here are our findings.

Pesticides

For this experiment, we compared a Raptor Inert ARC-18 column to a standard Raptor ARC-18 column. Pesticides analysis can be challenging because these panels typically contain a wide variety of compound chemistries. Phosphorylated, acidic, polar compounds, and/or metal chelating species, such as organophosphate pesticides, can react to the metal surfaces inside standard, stainless-steel analytical columns, which can degrade chromatographic performance. In contrast, the deactivated hardware in Restek’s inert LC columns significantly improved chromatographic results (Figures 1-2, Table I).

Figure 1: Analysis of pesticides on a Raptor Inert ARC-18 column without any preconditioning.
Pesticides on Raptor Inert ARC-18

LC_EV0596

Peaks

PeakstR (min)Precursor IonProduct Ion 1Product Ion 2Peak AreaPeak Height
1.Methamidophos1.33142.094.0125.1428941105189
2.Acephate1.55184.0143.048.9300642104729
3.Omethoate1.72214.0125.0182.9892008337690
4.Monocrotophos2.21224.1127.0193.121581078425
5.Dicrotophos2.35238.1112.172.0404916159292
6.Dimethoate2.52230.0125.0199.0807805342939
7.Trichlorfon2.53257.0108.9220.817394263266
8.Vamidothion2.54288.0146.0118.01333829547308
9.Mevinphos isomer 12.55241.9126.9192.9311274129961
10.Mevinphos isomer 22.76241.9126.9192.97403029802
11.Carbaryl3.18202.1145.0127.03967111924
12.Isocarbophos3.52291.1231.1121.13329411941
13.Dimethomorph isomer 13.96388.2300.9165.1511766172977
14.Dimethomorph isomer 24.13388.2300.9165.1877031328826
15.Temephos5.70467.1124.9418.916431064751

Conditions

ColumnRaptor Inert ARC-18 (cat.# 9314A12-T)
Dimensions:100 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Temp.:50 °C
Standard/SampleLC multiresidue pesticide standard #1 (cat.# 31972)
Diluent:Water, 0.1% formic acid
Conc.:1 ng/mL
Inj. Vol.:5 µL
Mobile Phase
A:Water, 2 mM ammonium formate, 0.1% formic acid
B:Methanol, 2 mM ammonium formate, 0.1% formic acid
Time (min)Flow (mL/min)%A%B
0.000.4955
2.000.44060
4.000.42575
6.000.40100
7.500.40100
7.510.4955
9.000.4955
DetectorShimadzu LCMS-8060
Ion Mode:ESI+
Mode:MRM
InstrumentShimadzu Nexera X2
Figure 2: Compared to standard LC columns, Raptor Inert LC columns provide increased peak area and height for pesticides.
Comparison of Pesticides on Raptor Inert ARC-18 and Raptor ARC-18

LC_EV0591

Comparison of Pesticides on Raptor Inert ARC-18 and Raptor ARC-18 - Overlay View

LC_EV0593

Table I: Pesticide peak areas and heights were up to 2X higher on Raptor Inert LC columns compared to standard columns.

CompoundPeak AreaPeak Height
Stainless SteelInertAreas Ratio (Inert/Stainless Steel)Stainless SteelInertHeight Ratio (Inert/Stainless Steel)
Methamidophos2549694289411.68525531051892.00
Acephate1687763006421.78584181047291.79
Omethoate5795028920081.542161573376901.56
Monocrotophos1400952158101.5451402784251.53
Dicrotophos3409784049161.191353801592921.18
Dimethoate4611568078051.751887463429391.82
Trichlorfon842331739422.0734793632661.82
Vamidothion91326413338291.463543115473081.54
Mevinphos isomer 12136323112741.46821051299611.58
Mevinphos isomer 25609374030 1.3229070298021.03 
Carbaryl43590396710.9114563119240.82
Isocarbophos21587332941.549062119411.32
Dimethomorph isomer 14624255117661.111669901729771.04
Dimethomorph isomer 28961098770310.98 3116573288261.06 
Temephos987931643101.6635383647511.83

Mycotoxins

In addition to pesticides analysis, we also wanted to test whether using an inert LC column could improve results for mycotoxins analysis. Because mycotoxins are reactive compounds that can have acidic, polar, or metal-chelating groups, extensive column conditioning and equilibration is often required to obtain adequate chromatography. For these experiments, we used columns with a Biphenyl stationary phase for optimal selectivity. The results shown in Figures 3-4 and Table II clearly demonstrate that peak shape and response for mycotoxins can also be improved by using an inert column.

Figure 3: Raptor Inert Biphenyl columns provide excellent peak shape for mycotoxins without additional acid passivation or mobile phase additives.
Multi-Mycotoxin Analysis on Raptor Inert Biphenyl

LC_FS0552

Peaks

PeakstR (min)Conc.
(ng/mL)
Precursor IonProduct Ion Peak AreaPeak Height
1.Nivalenol 0.8810295.1137.1418264495
2.Deoxynivalenol1.2510297.2231.017346281906
3.Fusarenon-X1.9210355.1137.17668121790
4.15-Acetyldeoxynivalenol3.0810339.2137.131369517570
5.3-Acetyldeoxynivalenol3.1410339.2213.122613296396
6.Tenuazonic acid4.1110198.1125.047828197658
7.Altenuene4.6010293.2257.11138502059699
8.Alternariol5.2710259.0185.1732721302192
9.Ergosine5.2810548.4208.14866209366601
10.Citrinin5.4610251.2233.110078809828889
11.Ergosinine5.4610548.4208.14967348740527
12.Fumonisin B15.4810722.5352.31228782415567
13.Diacetoxyscirpenol5.6210384.2247.1681391208825
14.Ergotamine5.7110582.4223.24930039274155
15.Ergocornine5.8510562.4268.23870257732744
16.Ergotaminine5.9610582.4223.24621199237991
17.HT-26.1310447.2345.115221323765
18.Ergocryptine6.1910576.4268.252220411360838
19.Fumonisin B36.2310706.4336.21433023444421
20.Ergocristine6.4410610.4223.21955624450058
21.Fumonisin B26.5910706.4336.21517193869822
22.Tentoxin6.6210415.2312.2951752131906
23.α-Zearalenol6.9110303.1285.130224702420
24.Ergocorninine6.9310562.4268.270402914389283
25.Aflatoxin G26.9710331.2189.02628245274353
26.T-27.0910489.2387.1565351394735
27.Ergocryptinine7.1810576.4268.277897216765348
28.Ergocristinine7.4010610.4223.2158305332975663
29.Aflatoxin G17.4510329.1199.73043896102959
30.Zearalenone7.5910319.2283.137162927455
31.Alternariol monomethylether7.6210273.0199.131024640689
32.Aflatoxin B27.6310315.1287.02956485724754
33.Aflatoxin B18.0210313.2241.12235204425821
34.Ochratoxin A8.2510404.1239.01900604411953

Conditions

ColumnRaptor Inert Biphenyl (cat.# 9309A12-T)
Dimensions:100 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Temp.:60 °C
Standard/Sample
Aflatoxins standard (cat.# 34121)
Ochratoxin A standard (cat.# 34122)
Diluent:50:50 Water:methanol
Conc.:10 ng/mL
Inj. Vol.:5 µL
Mobile Phase
A:Water, 0.05% formic acid
B:Methanol, 0.05% formic acid
Time (min)Flow (mL/min)%A%B
0.000.47525
5.000.45050
9.000.40100
9.010.47525
11.00.47525
Max Pressure:440 bar
DetectorWaters Xevo TQ-S
Ion Mode:ESI+
Mode:MRM
InstrumentWaters ACQUITY UPLC I-Class
Notes

Figure 4: Substantial increases in peak area and height were seen for fumonisins when using a Raptor Inert Biphenyl column.
Comparison of Fumonisin B Compounds on Raptor Inert Biphenyl and Raptor Biphenyl

LC_FS0554

Comparison of Fumonisin B Compounds on Raptor Inert Biphenyl and Raptor Biphenyl—Overlay View

LC_FS0556

Table II: For mycotoxins, peak height increased up to 10X and peak area increased up to 6X when using a Raptor Inert Biphenyl column compared to a standard column.

CompoundPeak AreaPeak Height
Stainless SteelInertAreas Ratio (Inert/Stainless Steel)Stainless SteelInertHeight Ratio (Inert/Stainless Steel)
Fumonisin B1325781228783.7739954424155676.05
Fumonisin B2234271517196.48383130386982210.10
Fumonisin B3298641433024.8047227934444217.29
Ergocristine1711971955621.14386589844500581.15
Ergocristinine139311615830531.1429212317329756631.13
Ergotamine4336354930031.14814951892741561.14
Ergotaminine3973704621191.16788540392379911.17
Ergocryptine4464815222041.179671753113608391.17
Ergocryptinine6587887789721.1813680420167653481.23
Ergocornine3705093870251.04724898177327441.07
Ergocorninine5901677040291.1912052359143892831.19
Ergosine4452434866201.09863093293666021.09
Ergosinine4390264967341.13782078587405271.12
T-243286565351.31104623313947351.33
HT-210183152211.492167033237651.49
Tentoxin70973951751.34157716421319071.35
Ochratoxin1736861900601.09403968244119531.09
Diacetoxyscirpenol47850681391.4284640312088261.43
Fusarenone X386576681.98604091217902.02
15-acetyl-DON17055313691.842698625175701.92
3-acetyldeoxyvinalenol13353226131.691792042963961.65
Aflatoxin G21715972628241.53342950152743541.54
Aflatoxin G12240583043891.36460795961029591.32
ZON25617371621.456569159274551.41
Aflatoxin B21593892956481.85346248957247541.65
Aflatoxin B12659352235200.84533557644258210.83
Alpha-zearalenol16202302241.873820927024201.84
Deoxynivalenol6935173462.501179272819062.39
Nivalenol179041822.3425276644952.55
Altenuene632241138501.80118795820597001.73
Alternariol monomethyl ether19537310241.594289226406891.49
Alternariol48204732721.5283741013021921.56
Citrinin49990010078802.02503118298288901.95
Tenuazonic acid21503478282.22892931976582.21

While our pesticides and mycotoxins experiments showed that columns made with inert hardware can improve chromatographic performance, the independent results from Carrera and Wuppermann are even more compelling. Based on the experiences of these well-regarded scientists, in addition to our own testing, it seems clear that Restek’s Inert LC columns have many potential benefits to laboratories analyzing pesticides and mycotoxins. It will be interesting to see what other analyses may be improved as well!

Products Mentioned


Raptor Inert Biphenyl, 2.7 µm, 100 x 2.1 mm HPLC Säule
Ochratoxin-Standard, 10 µg/mL, Acetonitril, 1 mL/Ampulle
Aflatoxine (B1, B2, G1, G2) Standard 10 µg/mL, Acetonitril, 1 mL/Ampulle
LC Standard Nr. 1 für Multipestizidrückstände, Organophosphorverbindungen, 100 μg/mL, Acetonitril, 1 mL/Ampulle
Raptor Inert ARC-18, 2.7 µm, 100 x 2.1 mm HPLC Säule
GNBL4720