FAQ

How much sample can I inject on my LC column and still avoid band broadening?

31 Jul 2025

Injection volumes, as well as optimal flow rates, are limited by the size of the column. Ideally, if your sample is in the same solvent as your mobile phase, approximate volumes should range as follows:

Column IDVolume (µL)
2.1 mm (30 -100 mm length)1-3
3.0-3.2 mm (50-150 mm length)2-12
4.6 mm (50-250 mm length)8-40

Autoren

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    View Articles
GNOT5051