Artikel

Analysis of Impurities in Ethylene by ASTM D6159-97

29 Jan 2025

feature PCAN1174

Ethylene is one of the highest volume chemicals produced in the world with global production exceeding 100 million metric tons annually. Ethylene is primarily used in the manufacture of polyethylene, ethylene oxide, and ethylene dichloride as well as many other lower volume products. Most of these production processes use various catalysts to improve product quality and process yield. Impurities in ethylene can damage the catalysts, resulting in significant replacement costs, reduced product quality, process downtime, and decreased yield.

Ethylene is typically manufactured through the use of steam cracking. In this process, gaseous or light liquid hydrocarbons are combined with steam and heated to 750–950 °C in a pyrolysis furnace. Numerous free radical reactions are initiated, and larger hydrocarbons are converted (cracked) into smaller hydrocarbons. The high temperatures used in steam cracking promote the formation of unsaturated or olefinic compounds like ethylene. Ethylene feedstocks must be tested to ensure that only high-purity ethylene is delivered for subsequent chemical processing.

Testing typically follows ASTM D6159-97, a GC/FID method that employs a two-column configuration consisting of an alumina PLOT column with KCl deactivation (50 m x 0.53 mm ID) coupled to a methyl silicone column (30 m x 0.53 mm ID x 5.0 μm df).

Figure 1: Methane and ethane are well resolved in high-purity ethylene samples.
Ethylene Purity Analysis using ASTM D6159-97

GC_PC01109

Peaks

Peaks
1.Methane
2.Ethane
3.Ethylene

Conditions

ColumnSee notes.
Standard/SampleEthylene gas
Injection
Inj. Vol.:1.0 µL split
Liner:2mm Splitless
Inj. Temp.:200 °C
Split Vent Flow Rate:60 mL/min.
Oven
Oven Temp.:35 °C (hold 2 min) to 190 °C at 4 °C/min (hold 15 min)
Carrier GasHe, constant pressure (8.0 psi, 55.2 kPa)
Linear Velocity:25.4 cm/sec @ 35 °C
DetectorFID @ 200 °C
Make-up Gas Type:N2
Data Rate:20 Hz
InstrumentHP5890 GC
NotesGC liner cat.# 20712 was used to produce this chromatogram, but it has since been discontinued. We recommend GC liner cat.# 20713 as an alternative.

Rt-Alumina BOND/KCl, 50 m, 0.53 mm ID, 10.0 µm (cat.# 19760) in series with Rtx-1, 30 m, 0.53 mm ID, 5.0 µm (cat.# 10179), connected with a Universal Press-Tight Connector (cat.# 20401)

Samples of high-purity ethylene typically contain only two minor impurities, methane and ethane, which can be detected in low ppmv concentrations (Figure 1). However, steam cracking can also produce higher molecular weight hydrocarbons, especially when propane, butane, or light liquid hydrocarbons are used as starting materials. Although fractionation is used in the final production stages to produce a high-purity ethylene product, it is still important to be able to identify and quantify any other hydrocarbons present in an ethylene sample. Achieving sufficient resolution of all of these compounds can be challenging due to their similarities in boiling point and chemical structure. ASTM D6159-97 addresses this issue by combining the separation power of two different types of capillary columns.

The Rt-Alumina BOND/KCl PLOT column has excellent separation capabilities for low molecular weight hydrocarbons ranging from C1 through C12, but complete resolution of all compounds is not always possible depending on the conditions that are employed. Figure 2 shows the analysis of an ethylene sample that has been spiked with the typical hydrocarbons that may be present after ethylene production. When using the temperature conditions supplied in the method, there are coelutions between three different peak pairs. Acetylene and isobutane (peaks 7 and 8) elute at the same retention time; propadiene and n-butane (peaks 9 and 10) are only partially resolved; and there is a complete coelution between methyl acetylene and n-pentane (peaks 16 and 17).

Figure 2: Analyzing ethylene on an alumina column alone results in coelutions that prevent quantification of several impurities.
Ethylene Purity Analysis per ASTM D6159-97 (except without Rtx-1)

GC_PC01138

Peaks

Peaks
1.Methane
2.Ethane
3.Ethylene
4.Propane
5.Cyclopropane
6.Propylene
7.Acetylene
8.Isobutane
9.Propadiene
Peaks
10.n-Butane
11.<a class="cmpd_link" title="View compound information for trans-2-Butene” title=”View compound information for trans-2-Butene” href=”https://ez.restek.com/compound/view/en/624-64-6/trans-2-Butene”>trans-2-Butene
12.1-Butene
13.Isobutylene
14.<a class="cmpd_link" title="View compound information for cis-2-Butene” title=”View compound information for cis-2-Butene” href=”https://ez.restek.com/compound/view/en/590-18-1/cis-2-Butene”>cis-2-Butene
15.Isopentane
16.Methylacetylene
17.<a class="cmpd_link" title="View compound information for n-Pentane” title=”View compound information for n-Pentane” href=”https://ez.restek.com/compound/view/en/109-66-0/n-Pentane”>n-Pentane
18.1,3-Butadiene

Conditions

ColumnRt®-Alumina BOND/KCl, 50 m, 0.53 mm ID, 10 µm (cat.# 19760)
Standard/Sampleethylene gas plus C1-C5 hydrocarbons
Injection
Inj. Vol.:1 µL split
Liner:2mm Splitless
Inj. Temp.:200 °C
Split Vent Flow Rate:60 mL/min.
Oven
Oven Temp.:35 °C (hold 2 min) to 190 °C at 4 °C/min (hold 15 min)
Carrier GasHe, constant pressure (5.0 psi, 34.5 kPa)
Linear Velocity:25 cm/sec @ 35 °C
DetectorFID @ 200 °C
Make-up Gas Type:N2
Data Rate:20 Hz
InstrumentHP5890 GC
NotesGC liner cat.# 20712 was used to produce this chromatogram, but it has since been discontinued. We recommend GC liner cat.# 20713 as an alternative.

By combining an Rt-Alumina BOND/KCl column with an Rtx-1 column, complete resolution for all of the compounds of interest can be achieved. The Rtx-1 column supplements the separation achieved on the Rt-Alumina BOND/KCl column by contributing additional selective retention of less polar compounds like isobutane, n-butane, and n-pentane. The extra retention of these compounds allows for the complete separation of the slightly more polar compounds like acetylene, propadiene, and methyl acetylene. Figure 3 shows the analysis of the same ethylene sample spiked with hydrocarbons. All of the compounds that are identified in the method can now be resolved for accurate identification and quantitation.

Figure 3: All impurities are fully resolved and easily quantifiable when using an Rt-Alumina BOND/KCl column coupled to an Rtx-1 column.
Ethylene and C1-C5 Hydrocarbons by ASTM D6159-97 on Rt-Alumina BOND/KCl, Rtx-1

GC_PC01110

Conditions

ColumnRt-Alumina BOND/KCl *, 50 m, 0.53 mm ID, 10 µm (cat.# 19760)
Standard/SampleEthylene gas plus C1 through C5 hydrocarbons
Injection
Inj. Vol.:1 µL split
Liner:2 mm splitless
Inj. Temp.:200 °C
Split Vent Flow Rate:60 ml/min.
Oven
Oven Temp.:35 °C (hold 2 min) to 190 °C at 4 °C/min (hold 15 min)
Carrier GasHe, constant pressure (8.0 psi, 55.2 kPa)
Linear Velocity:25.4 cm/sec @ 35 °C
DetectorFID @ 200 °C
Make-up Gas Type:N2
Data Rate:20 Hz
InstrumentHP5890 GC
NotesGC liner cat.# 20712 was used to produce this chromatogram, but it has since been discontinued. We recommend GC liner cat.# 20713 as an alternative.

* Rt-Alumina BOND/KCl, 50 m, 0.53 mm ID, 10.0 μm (cat.# 19760) in series with an Rtx-1, 30 m, 0.53 mm ID, 5.0 μm (cat.# 10179) connected using a universal Press-Tight connector (cat. # 20401).

(conditions as per ASTM D6159-97)

When testing for impurities in ethylene using ASTM D6159-97, the combination of an Rt-Alumina BOND/KCl column coupled to an Rtx-1 column provides the best resolution of the most common hydrocarbon contaminants. Restek PLOT columns are manufactured using a new technology that significantly reduces particle release, extending column lifetime and giving highly reproducible retention times. Restek columns provide reliable results that can be used to protect expensive catalysts, make faster process adjustments, and improve product yield.

Author

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    View all posts
PCAN1174
Version: A