Sin categorizar

Reduce Downtime and Cost of Materials with Rugged Rxi-5Sil MS GC Columns

26 Oct 2020

  • Save costs with long column lifetime.
  • Reduce downtime from column trimming and replacement.
  • Improve peak shape for active compounds.

When performing GC/MS analysis of drugs, many chemists choose to derivatize samples prior to analysis. Derivatization not only increases the volatility of some drug compounds, but it also reduces activity, resulting in improved peak shape and more accurate quantification. An additional advantage is that derivatized compounds have a higher molecular weight, thus producing more reliable mass spectra than underivatized compounds. Despite these benefits, derivatization reagents are often harsh and can damage analytical columns, leading to high bleed, significant reduction in retention times, and increased tailing for active compounds. Often, this damage is concentrated near the head of the column, so trimming a short length can improve results. However, trimming is a finite solution as repeated clipping ultimately results in decreased efficiency and shorter column lifetimes. Choosing a more rugged column, such as the Rxi-5Sil MS column, is a better alternative. The Rxi-5Sil MS column is extremely stable and holds up to harsh treatment, including repeated exposure to derivatization reagents.

The analysis of amphetamine illustrates the ruggedness of the arylene-based Rxi-5Sil MS polymer. Amphetamine is typically derivatized, because the underivatized form is an active basic compound that produces only a few low molecular weight ions for monitoring. In contrast, upon derivatization, activity decreases, resulting in dramatically improved peak shape and more accurate quantitation. Additionally, several higher molecular weight ions are produced, which can be monitored for definitive identification (Figure 1).

Figure 1: Derivatizing amphetamine results in more definitive identification by creating higher molecular weight ions.
figure article CFAR3234 01

Phase Stability Extends Column Lifetime

In order to demonstrate the ruggedness of the Rxi-5Sil MS column, 400 injections of heptafluorobutyric acid anhydride (HFBA) in butyl chloride were performed. HFBA is a very harsh derivatization reagent, and the concentration of reagent in the solvent was equivalent to that of a derivatized sample. Throughout the course of 400 injections, bleed, retention, and peak shape for active compounds was monitored by periodically injecting a column test mix containing active compounds (1,6-hexanediol, 4-chlorophenol, and dicyclohexylamine). Chromatographic results were remarkably consistent, even after 400 injections (Figure 2). Column bleed was monitored over the course of the experiment and remained below 5pA (Figure 3). The consistency of retention time data and low bleed levels demonstrate phase stability, which results in longer column lifetimes and reduced maintenance and replacement costs.

Figure 2: Rugged Rxi-5Sil MS columns produce consistent retention times, even after 400 injections of derivatization reagent.
Column Test Mix on Rxi-5Sil MS

GC_CF01131

Conditions

ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
Standard/SampleRxi-5Sil MS/XLB column test mix (cat.# 35226)
Injection
Inj. Vol.:1.0 µL split (split ratio 60:1)
Liner:4 mm recessed single taper
Inj. Temp.:250 °C
Oven
Oven Temp.:125 °C
Carrier GasHe, constant pressure
Linear Velocity:36 cm/sec @ 125 °C
DetectorFID @ 320 °C
InstrumentAgilent/HP6890 GC
NotesLiner cat.# 20983 was used to produce this chromatogram, but it has since been discontinued. We recommend cat.# 20984 as an alternative. For assistance choosing a replacement for this application, contact Restek Technical Service or your local Restek representative.
Figure 3: Low column bleed results in long column lifetimes, saving labs replacement costs.
figure article CFAR3234 02

Column bleed over 400 injections of HFBA derivatization reagent.

Symmetric Peaks for More Accurate Results

Peak shape was also monitored to ensure column inertness was stable over time—an important factor in maintaining accuracy. Peaks for the active test probes were symmetric even after 400 injections, allowing easy identification and consistent integration (Figure 4). In a second experiment to complement the test probe results, underivatized amphetamine was injected onto a new Rxi-5Sil MS column, an Rxi-5Sil MS column after 400 injections of derivatization reagent, and a new competitor column of equivalent phase chemistry. Even though underivatized amphetamine is highly active, peak symmetry on the Rxi-5Sil MS column was consistent and unaffected by exposure of the column to derivatization reagent (Figure 5). Additionally, peak shape on both the exposed and unexposed Rxi-5Sil MS column was better than that on the new competitor column (Figure 5).

Figure 4: Active probes show consistent, symmetric peak shape, demonstrating the inertness needed for accurate quantification.
figure article CFAR3234 03

Test probe symmetry over 400 injections of HFBA derivatization reagent.

Symmetry values <1 indicate peak tailing and values >1 indicate fronting.

Figure 5: Peak symmetry for underivatized amphetamine is significantly better on an Rxi-5Sil MS than on a competitor column, even after 400 injections of HFBA derivatization reagent.
figure article CFAR3234 04

Symmetry values <1 indicate peak tailing and values >1 indicate fronting.

The rugged arylene phase of the Rxi-5Sil MS column results in highly stable performance, even under the most demanding of analytical conditions, and its exceptional inertness ensures good peak shape for reproducible quantitation. The stability of the Rxi-5Sil MS column results in longer column lifetimes, reducing both downtime and replacement costs.

Products Mentioned


Rxi-5Sil MS GC Capillary Column, 15 m, 0.25 mm ID, 0.10 µm
XIL-350 Column Test Mix, 350 µg/mL ea. in Methylene Chloride, 1 mL/ampul

Autores

  • Amanda Rigdon is an analytical chemist with more than 21 years of experience in chromatography and mass spectrometry. Throughout her career, she has worked in a variety of analytical fields, including clinical forensics, food safety, environmental analysis, and cannabis testing. Her experience spans method development, validation, troubleshooting, and laboratory operations using both GC and LC methodologies. Amanda also served as Chief Scientific Officer of an ISO/IEC 17025-accredited cannabis testing laboratory, where she gained extensive experience in quality systems, regulatory compliance, and laboratory management. She is currently a Principal Scientist in the GC Applications Group at Restek, where she enjoys helping customers solve analytical challenges and sharing practical chromatography knowledge through technical support, presentations, and educational content.

    View all posts
  • Gary Stidsen
CFAR3234