Articoli

PFAS Analysis by U.S. EPA Method 533 for Drinking Water

03 Oct 2025

feature EVAR4285
  • Sample Matrix: drinking water
  • Sample Preparation: isotope dilution anion exchange solid-phase extraction
  • Detection Technique: LC-MS/MS

Determination of Per- and Polyfluorinated Alkyl Substances in Drinking Water by Isotope Dilution Anion Exchange Solid Phase Extraction and Liquid Chromatography/Tandem Mass Spectrometry

U.S. EPA Method 533 is used for the analysis of 25 PFAS in drinking water. It combines solid-phase extraction sample preparation using weak anion exchange (WAX) cartridges with LC-MS/MS analysis on a C18 column. The method allows flexibility in some aspects, such as sample volume, LC column, LC conditions, and detector conditions, but changes to sample preservation, quality control requirements, and the extraction procedure are prohibited. In addition, the ratio of sorbent mass to aqueous sample volume cannot be reduced. The full analyte list and recommended products for PFAS analysis in drinking water by Method 533 are provided below. In addition, you can register and directly compare analyte lists and matrices across PFAS analysis methods here or visit www.restek.com/PFAS to explore our technical resources for PFAS analysis.

Analyte List for U.S. EPA Method 533 for PFAS Analysis

PFAS acronyms, compound names, and CAS numbers vary across methods and guidelines, and they also depend on compound form (e.g., neutral vs. salt). The following list uses standardized terminology to allow easy comparison across the methods and guidelines in our comprehensive guide. Please refer to U.S. EPA Method 533 for method-specific nomenclature.

AcronymCompound NameCAS#
PFBSPerfluoro-1-butanesulfonic acid375-73-5
PFDAPerfluorodecanoic acid335-76-2
PFDoAPerfluorododecanoic acid307-55-1
PFHpAPerfluoroheptanoic acid375-85-9
PFHxSPerfluoro-1-hexanesulfonic acid355-46-4
PFHxAPerfluorohexanoic acid307-24-4
PFNAPerfluorononanoic acid375-95-1
PFOSPerfluorooctanesulfonic acid1763-23-1
PFOAPerfluorooctanoic acid335-67-1
PFUnAPerfluoroundecanoic acid2058-94-8
HFPO-DA, GenXHexafluoropropylene oxide dimer acid13252-13-6
PFBAPerfluorobutanoic acid375-22-4
PFPeAPerfluoropentanoic acid2706-90-3
PFHpSPerfluoro-1-heptanesulfonic acid375-92-8
PFPeSPerfluoro-1-pentanesulfonic acid2706-91-4
4:2 FTS1H,1H,2H,2H-Perfluorohexane sulfonic acid757124-72-4
6:2 FTS1H,1H,2H,2H-Perfluorooctane sulfonic acid27619-97-2
8:2 FTS1H,1H,2H,2H-Perfluorodecane sulfonic acid39108-34-4
9Cl-PF3ONS9-Chlorohexadecafluoro-3-oxanonane-1-sulfonic acid (F-53B Major)756426-58-1
11Cl-PF3OUdS11-Chloroeicosafluoro-3-oxaundecane-1-sulfonic acid (F-53B Minor)763051-92-9
ADONA4,8-Dioxa-3H-perfluorononanoic acid919005-14-4
PFEESAPerfluoro(2-ethoxyethane)sulfonic acid113507-82-7
NFDHANonafluoro-3,6-dioxaheptanoic acid151772-58-6
PFMPAPerfluoro-3-methoxypropanoic acid377-73-1
PFMBAPerfluoro-4-methoxybutanoic acid863090-89-5

Recommended Products for U.S. EPA Method 533 for PFAS Analysis

Analytical LC Columns:


Colonna HPLC Raptor C18, 2,7 µm, 50 x 2,1 mm
Colonna LC Force C18, 1,8 µm, 50 x 2,1 mm

Delay Columns:


PFAS Delay Column, 5 µm, 50 x 2,1 mm, colonna HPLC

PFAS delay column: 5 µm, 50 x 3.0 mm (cat.# 580959 [custom product])

Guard Columns:


Filtro precolonna UHPLC UltraShield, frit 0,2 µm, cad.
Holder EXP a connessione diretta per cartucce per precolonne EXP, comprende raccordo e ferrule
Cartuccia per precolonna Raptor C18 EXP, 2,7 µm, 5 x 2,1 mm, 3 pz.

Standards:


Standard Primario di Diluizione (PDS) dell' Analita secondo Metodo PFAS EPA 533, 2 µg/mL, Metanolo (1 mM NaOH)/2-Propanolo (98:2), 1 mL/ampolla
Standard di calibrazione PFAS 24, 1 μg/mL, metanolo (1 mM KOH), 1 mL/ampolla
Standard di calibrazione PFAS 28, 1 μg/mL, metanolo (1 mM KOH), 1 mL/ampolla

Sample Prep:


Cartuccia SPE Resprep Polimerica, WAX, 6ml/500mg, 30μm, 30-conf.
Cartuccia SPE Resprep Polimerica, WAX, 6ml/200mg, 30μm, 30-pk.

Sample Reservoirs:


Provette vuote SPE, serbatoio per campioni, 75 mL, polipropilene, 12 pz.
Connettore SPE, 1, 3, 6, 10 o 15 mL, polipropilene, 15 pz.

Vacuum Manifolds:


Vacuum Manifold Resprep QR-24
Vacuum Manifold Resprep QR-12

Vacuum Pumps:


Pompa da vuoto in PTFE Chemker 300, 20 l/min, CA 110 V, 60 Hz
Pompa da vuoto in PTFE Chemker 300, 18 l/min, CA 220 V, 50 Hz

General Lab Supplies:


Vial a volume ridotto in polipropilene con filettatura, 9 mm, 700 µL, 12 x 32 mm, 100 pz.
Vial a volume ridotto in polipropilene con filettatura, 9 mm, 700 µL, 12 x 32 mm, 1000 pz.
Tappi in polietilene da 2,0 mL con top solido di 9 mm, filettatura, membrana spessore 10 mm, trasparente, 100 pz.
Tappi in polietilene da 2,0 mL con top solido di 9 mm, filettatura, membrana spessore 10 mm, trasparente, 1000 pz.
Provette da centrifuga vuote da 50 ml, Polipropilene con tappo, 50 provette, 50-conf.
Provette da centrifuga vuote da 50 ml, Polipropilene con tappo, 50 provette, 500-conf.
Siringa di Plastica Norm-Ject, Luer Lock Tip, graduazione fino a 12 mL, 10 mL, Conf. da 100
Siringa di Plastica Norm-Ject, Luer Lock Tip, graduazione fino a 6 mL, 5 mL, Conf. da 100

Discover More Solutions for PFAS Analysis!

Start Saving
Time Today

Develop PFAS Methods in Minutes with our Pro EZLC Chromatogram Modeler

Start Saving Time Today

Develop PFAS Methods in Minutes with our Pro EZLC Chromatogram Modeler

Tags

Author

  • Restek Corporation

    Restek is a leading provider of chromatography columns, accessories, and certified reference materials. Trust Restek for reliable, high-quality analytical solutions.

    View all posts
EVAR4285
Version: B