Sin categorizar

Boost Productivity with Simultaneous PAH and PCB GC-MS Analysis

19 Jan 2026

feature EVFA5232
  • Increase laboratory efficiency and profitability by consolidating PAH and PCB GC-MS methods.
  • Next-generation TriMax column deactivation provides an exceptionally inert sample flow path, ensuring sharp peaks and consistent separations.
  • Easily adaptable—add compounds, explore other column dimensions, and optimize in seconds with Restek’s free EZGC chromatogram modeler.
helpful person on the phone
RMX Graphic Ad Contact us HTML Block SALES

Polycyclic aromatic hydrocarbons (PAHs) and polychlorinated biphenyls (PCBs) are persistent pollutants that are closely monitored in the environment as potential ecological and human health hazards. Routine testing is often performed by GC-MS, allowing labs to improve productivity and decrease cost per sample by consolidating GC-MS screening methods for PAHs and PCBs. Simultaneous analysis of these two contaminant groups often utilizes “XLB-type” columns for separation of PCB 28/31, but this comes at a cost as some PAH separations tend to suffer. A “5-type” column provides alternate selectivity, making it a suitable phase for method consolidation. The quick GC-MS method for PAHs and PCBs presented here can help labs operate more efficiently, providing excellent chromatography for concurrent analysis of both pollutant classes.

This method was developed on an RMX-5Sil MS column because it combines standard “5sil” selectivity with a next-generation TriMax deactivation, creating a highly inert sample flow path. Figure 1 demonstrates that across a fast, <20-minute run, the target PAHs and PCBs exhibited excellent peak shapes and separations. For example, while benz[a]anthracene; chrysene; and PCB 180 normally elute as a triplet, baseline USP resolution (≥1.5) was achieved on the RMX-5Sil MS column. To assess how consistent column performance was for key parameters, 100 injections at 1 ng on-column were made, and the results were extremely stable with tailing %RSD ranging 3.9-6.6% and retention time %RSD ranging 0.01-0.02%.  While this analysis is not a comprehensive view of all PAHs and PCBs of interest, it can be customized for user-specific compound lists—even on different column dimensions—using Restek’s free Pro EZGC chromatogram modeling software.

ad for RMX columns and Topaz liners
Figure 1: Excellent peak shapes and separations were achieved for simultaneous PAH and PCB analysis on an RMX-5Sil MS column.
Combined Analysis of PAHs and PCBs on RMX-5Sil MS by GC-MS

GC_GN1252

Peaks

PeakstR (min)Conc.
(µg/mL)
1.Naphthalene4.8541
2.2-Methylnapthalene5.4081
3.Acenaphthylene6.1341
4.Acenaphthene6.2771
5.Fluorene6.7001
6.Phenanthrene7.4811
7.Anthracene7.5241
8.PCB 287.7641
9.PCB 527.9931
10.Fluoranthene8.4611
11.PCB 1018.5641
12.Pyrene8.6441
PeakstR (min)Conc.
(µg/mL)
13.PCB 1188.9761
14.PCB 1539.1131
15.PCB 1389.2901
16.Benz[a]anthracene9.6541
17.Chrysene9.6851
18.PCB 1809.7251
19.Benzo[b]fluoranthene10.5481
20.Benzo[k]fluoranthene10.5591
21.Benzo[a]pyrene10.8061
22.Dibenz[a,h]anthracene11.6581
23.Indeno[1,2,3-cd]pyrene11.6581
24.Benzo[g,h,i]perylene11.8461

Conditions

ColumnRMX-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 17323)
Standard/Sample
2-Methylnaphthalene (cat.# 31285)
SV calibration mix #5 (cat.# 31011)
PCB congener mix #2 (cat.# 32294)
Diluent:Methylene chloride
Conc.:1 µg/mL
Injection
Inj. Vol.:1 µL splitless (hold 1.26 min)
Liner:Topaz 4.0 mm ID single taper inlet liner w/wool (cat.# 23303)
Inj. Temp.:250 °C
Purge Flow:3 mL/min
Oven
Oven Temp.:40 °C (hold 1 min) to 120 °C at 30 °C/min to 350 °C at 20 °C/min (hold 2 min)
Carrier GasHe, constant flow
Flow Rate:1.2 mL/min
Linear Velocity:39.7 cm/sec
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
1350-5505
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Source Temp.:330 °C
Quad Temp.:180 °C
Solvent Delay Time:3 min
Tune Type:PFTBA
Ionization Mode:EI
InstrumentAgilent 7890B GC & 5977B MSD
EZGC software ad
RMX benefits graphic

Products Mentioned


RMX-5Sil MS GC Capillary Column, 30 m, 0.25 mm ID, 0.25 µm
Topaz, Single Taper Inlet Liner, 4.0 mm x 6.5 x 78.5, for Agilent GCs, w/Quartz Wool, Premium Deactivation, 5-pk.
Restek Electronic Leak Detector
2-Methylnaphthalene Standard, 1000 µg/mL, Methylene Chloride, 1 mL/ampul
SV Calibration Mix #5, 2000 µg/mL, Methylene Chloride, 1 mL/ampul
PCB Congener Mix #2, 10 µg/mL, Isooctane, 1 mL/ampul

Autores

EVFA5232