Sin categorizar

Maximize Confidence in Seized Drug Identification by GC-MS

26 Jan 2026

drug powder in bags
  • Next-generation TriMax column deactivation creates a robust neutral surface and exceptionally inert sample flow path.
  • Maximum inertness improves peak symmetry for a wide range of challenging drug classes.
  • Stable, symmetrical peaks mean more accurate identification for even the most problematic drugs.
helpful person on the phone
RMX Graphic Ad Contact us HTML Block SALES

Seized drug labs frequently analyze samples via GC-MS to determine if an unknown substance is an illicit drug, and GC column choice plays an integral role in data quality and method reliability. Since a wide variety of chemical compounds need to be identified, midpolarity columns, such as “5 type” columns, are a common choice for drug analysis. Silarylene polymers are structurally modified “5 type” phases that provide enhanced thermal stability, making a “5sil” column the perfect choice for extended lifetime and consistent performance with a mass spectrometer. When analyzing a wide range of drugs, some compound classes exhibit worse behavior than others, which can make drug identification more complex due to tailing peaks and drifting retention times. Acidic compounds (e.g., barbiturates) are prone to uneven phase partitioning due to strong interactions between weakly acidic N-H groups and silanols on the surface of the GC column. Other drugs may have basic functionalities (e.g., amines) and form hydrogen bonds with the surface, again delaying partitioning out of the column phase, or sometimes being completely adsorbed. By preventing surface silanol interactions, seized drug labs can see improved performance for the most problematic compounds across a wide range of functionalities.

To minimize the impact of silanols, Restek has developed a next-generation TriMax deactivation that is applied to all columns in the RMX family. This revolutionary treatment creates a robust, highly inert surface that improves peak shape for a wide range of drug classes. Further, analysts can be confident in column performance because each column is individually QC tested with acidic, basic, and neutral compounds to ensure deactivation effectiveness and process control. This makes RMX-5Sil MS columns—which combine plug-in 5sil polarity, high thermal stability, and maximum surface inertness—ideal for improving seized drug identification (Figure 1). Further, as shown in Figure 2, the RMX-5Sil MS column outperforms a competitor’s premium column when comparing peak shape for several difficult basic analytes (doxylamine, cyclobenzaprine, codeine) as well as a weakly acidic analyte (alprazolam). Under the same instrument conditions, we can see an improvement in peak shape that can only be attributed to the maximum inertness of the RMX-5Sil MS column surface, which minimizes secondary analyte interactions with silanols that would otherwise cause peak tailing and poor symmetry.

ad for RMX columns and Topaz liners
Figure 1: RMX-5Sil MS columns provide excellent chromatographic results for 34 drugs of abuse analyzed by GC-MS in less than 21 minutes.
Acid, Base, and Neutral Drugs and Metabolites on RMX-5Sil MS

GC_CF1185

Peaks

PeakstR (min)Conc.
(µg/mL)
1.Propofol2.5325
2.Benzocaine4.7325
3.Butabarbital6.2525
4.Butalbital6.3325
5.Phenacetin6.5225
6.Cotinine6.6125
7.Amobarbital7.1725
8.Pentobarbital7.625
9.Secobarbital8.5425
10.Norketamine9.0625
11.Prilocaine9.3325
PeakstR (min)Conc.
(µg/mL)
12.Ketamine9.7125
13.Diphenhydramine10.0225
14.Phencyclidine (PCP)10.5425
15.Doxylamine10.9725
16.Tramadol11.725
17.Phenobarbital12.1225
18.Chlorpheniramine12.9225
19.Procaine13.3925
20.Venlafaxine14.6525
21.Brompheniramine15.0225
22.Methadone15.6425
PeakstR (min)Conc.
(µg/mL)
23.Dextromethorphan15.6925
24.Amitriptyline16.7025
25.Cocaine16.8225
26.Imipramine17.0325
27.Tetracaine17.1725
28.Cyclobenzaprine17.2425
29.Sertraline18.1625
30.Codeine18.2925
31.6-Acetylmorphine18.9625
32.Heroin19.3825
33.Zolpidem19.9725
34.Alprazolam20.5325

Conditions

ColumnRMX-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 17323)
Standard/Sample
Diluent:Ethyl acetate
Conc.:25 ppm
Injection
Inj. Vol.:1 µL split (split ratio 10:1)
Liner:Topaz 4 mm single taper inlet liner with wool (cat.# 23303)
Inj. Temp.:250 °C
Split Vent Flow Rate:25 mL/min
Oven
Oven Temp.:150 °C (hold 1.0 min) to 210 °C at 4 °C/min to 320 °C at 30 °C/min (hold 2.0 min)
Carrier GasHe, constant flow
Flow Rate:2.0 mL/min
Linear Velocity:54 cm/sec @ 150 °C
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
11.040-5001562
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Source Temp.:230 °C
Electron Energy:70 eV
Tune Type:PFTBA
Ionization Mode:EI
InstrumentAgilent 7890A GC & 5975C MSD
Sample PreparationIndividual standards (100 ppm) were combined into a final 25 ppm solution in ethyl acetate.
Figure 2: Highly inert RMX-5Sil MS columns improve peak shape for challenging acidic and basic drugs, allowing for more confidence in identification.
Comparison of Seized Drugs on RMX-5Sil MS and a Premium Competitor Column (Zooms)

GC_CF1186

Peaks

PeaksConc.
(µg/mL)
tR (min) for RMX-5Sil MStR (min) for Premium Competitor Column
1.Doxylamine2510.9711.54
2.Cyclobenzaprine2517.2417.54
3.Codeine2518.2918.53
4.Alprazolam2520.5320.83

Conditions

ColumnSee notes for names, 30 m, 0.25 mm ID, 0.25 µm
Standard/Sample
Diluent:Ethyl acetate
Conc.:25 ppm
Injection
Inj. Vol.:1 µL split (split ratio 10:1)
Liner:Topaz 4 mm single taper inlet liner with wool (cat.# 23303)
Inj. Temp.:250 °C
Split Vent Flow Rate:25 mL/min
Oven
Oven Temp.:150 °C (hold 1.0 min) to 210 °C at 4 °C/min to 320 °C at 30 °C/min (hold 2.0 min)
Carrier GasHe, constant flow
Flow Rate:2.0 mL/min
Linear Velocity:54 cm/sec @ 150 °C
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
11.040-5001562
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Source Temp.:230 °C
Electron Energy:70 eV
Tune Type:PFTBA
Ionization Mode:EI
InstrumentAgilent 7890A GC & 5975C MSD
Sample PreparationIndividual standards (100 ppm) were combined into a final 25 ppm in ethyl acetate.
NotesColumns tested: RMX-5Sil MS (cat.# 17323) and a premium competitor column
EZGC software ad
RMX benefits graphic

Products Mentioned


RMX-5Sil MS GC Capillary Column, 30 m, 0.25 mm ID, 0.25 µm
Topaz, Single Taper Inlet Liner, 4.0 mm x 6.5 x 78.5, for Agilent GCs, w/Quartz Wool, Premium Deactivation, 5-pk.
Restek Electronic Leak Detector

Autores

  • Corby Hilliard

    Corby Hilliard started his Restek career in the Quality Assurance department where he spent his first seven years as a QA analyst and worked his way up to a senior QA analyst. He then moved on to the Innovations department in 2009 as a GC Solutions Advanced Scientist. His experience is predominately in gas chromatography (GC) using various detectors and mass spectrometry as well as troubleshooting and method development. His primary work for Restek is generating product application data, using new and existing products in food safety, petrochemical, environmental, and the GC accessories line. He also is involved with quality and R&D as well as new product development.

    View Articles
  • Erica Pack, PhD
  • Ramkumar Dhandapani
  • Colton Myers
CFFA4918
Version: A